Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration wave number

Compound 6 contains seven iron-based units [ 12], of which the six peripheral ones are chemically and topologically equivalent, whereas that constituting the core (Fe(Cp)(C6Me6)+) has a different chemical nature. Accordingly, two redox processes are observed, i.e., oxidation of the peripheral ferrocene moieties and reduction of the core, whose cyclic voltammetric waves have current intensities in the 6 1 ratio. Clearly, the one-electron process of the core is a convenient internal standard to calibrate the number of electron exchanged in the multi-electron process. In the absence of an internal standard, the number of exchanged electrons has to be obtained by coulometry measurements, or by comparison with the intensity of the wave of an external standard after correction for the different diffusion coefficients [15]. [Pg.208]

The wavelength (or wave number) scale calibration of infrared spectrophotometers is usually carried out with the aid of a strip of polystyrene film fixed on a frame. It consists of several sharp absorption bands, the wavelengths of which are known accurately and precisely. Basically, all IR-spectrophotometers need to be calibrated periodically as per the specific instructions so as to ascertain their accuracy and precision. [Pg.330]

Wavelength or wave-number calibration of infrared absorption spectra is normally accomplished by simultaneously recording the spectrum along... [Pg.171]

Owing to the very high rate of decomposition, in-situ measurement of concentration by means of Raman spectroscopy was applied. The peroxide used was f-butylperoxy pivalate (see Chapter 5.1, Table 5.1-2) dissolved in n-heptane at a concentration of 1 wt.%. In order to observe the change in intensity of absorption of the 0-0 bond at 861 cm 1, the spectrometer was adjusted to this wave number. The change of intensity is an indication of the reduction in the peroxide concentration, and was recorded as a function of time. The apparatus was calibrated before measuring the intensity of peroxide solutions of different concentrations [22],... [Pg.87]

Despite the advantages of employing wave numbers, intensity vs. A plots will probably continue to be common, both because of habit and because of the design of commercial, visible and ultraviolet spectrometers many of which employ scales calibrated in wavelength. [Pg.15]

The present position is that some quantum yields are known with an accuracy of 10-20%, but that further work is required before the reliability of published data can be properly assessed. A relatively new development, however, is capable of exploitation. This rests on the determination of the true shapes on a quantum intensity-wave number plot of the fluorescence bands of certain solutions selected as standards by means of carefully calibrated monochromator-photomultiplier combinations, and on agreement on their yields (14,37,44,45,49). It is then possible to use these solutions to calibrate other monochromator-photomultiplier instruments and so to measure true band shapes for other solutions. Comparison of band areas for any solution and for a standard under conditions of equal amounts of monochromatic exciting light absorbed will then give a value of the quantum yield. [Pg.31]

Different references give variations to the values for the different bands. These vary depending on the calibration of the instrument used, the exact chemistry of the system, and the method of preparation and testing of the sample. Samples tested in a solvent may indicate the carbonyl band at a wave number of 1745 cm-1, whereas the solid material will be at approximately 1730 cm-1. [Pg.33]

Spectrometers can be calibrated for wave-numbers using a polystyrene film supplied by the NIST (National Institute of Standards and Technology). Table 12.2 shows the confidence inter als of some wave-numbers for a resolution of 0.5 cm" under vacuum. [Pg.229]

Furthennore, at wave numbers between 2850 and 3100 cm", the Si-CHs stretch bending could be observed, which will be used for the internal calibration of the Sih-H peak of the measured system. This peak remains constant in intensity during the hydrosilylation reaction (see Fig. 2). [Pg.434]

Fig. 2. Analytical bands and base lines for some common termonomer units in C2-C3-diene terpolymers. A Dicyclopentadiene at 3.284 p. The spectrum is scanned on a linear wave number instrument (Perkin-Elmer Model 125). CC14 solution, concentration 0.032 g cm-3, cell thickness 0.10 cm. The thinner spectrum shows the v(C-H) bands, mostly arising from the C2—C3 portion of the terpolymer. B 1.4-hexadiene at 10.35 p base line is drawn accounting for superposition of a C3 band and out-of-plane deformation band of the hydrogens on the tram double bond of the third monomer. Film thickness 0.010 nm. C Ethylidene norbomene at 5.93 p, Absorption at 5.76 p due to antioxidant Film thickness 0.024 cm. D Methyl tetrahydroindene at 12.56 p. Band strongly overlapped by the broad CH2 rocking band of C3 with maximum absorption at ca. 12.3 p. Film thickness 0.014 cm. A calibration straight line through tlx origin is obtained, for terpolymers with, 4C-labelled methyl tetrahydroindene, only by drawing base line from 12.35... Fig. 2. Analytical bands and base lines for some common termonomer units in C2-C3-diene terpolymers. A Dicyclopentadiene at 3.284 p. The spectrum is scanned on a linear wave number instrument (Perkin-Elmer Model 125). CC14 solution, concentration 0.032 g cm-3, cell thickness 0.10 cm. The thinner spectrum shows the v(C-H) bands, mostly arising from the C2—C3 portion of the terpolymer. B 1.4-hexadiene at 10.35 p base line is drawn accounting for superposition of a C3 band and out-of-plane deformation band of the hydrogens on the tram double bond of the third monomer. Film thickness 0.010 nm. C Ethylidene norbomene at 5.93 p, Absorption at 5.76 p due to antioxidant Film thickness 0.024 cm. D Methyl tetrahydroindene at 12.56 p. Band strongly overlapped by the broad CH2 rocking band of C3 with maximum absorption at ca. 12.3 p. Film thickness 0.014 cm. A calibration straight line through tlx origin is obtained, for terpolymers with, 4C-labelled methyl tetrahydroindene, only by drawing base line from 12.35...
We start with the observations of Kirchoff and Bunsen in Germany in 1859. They observed the bright line spectra for many alkali and alkaline-earth metal-based salts and are credited with the discovery of spectro-chemical analysis. The so-called principal atomic emission series for the common alkali metals is shown in Fig. 4.66. Note that these older atomic spectra are calibrated in terms of wave number, denoted by v, of the emitted radiation whose units are in reciprocal centimeters (denoted by cm ). [Pg.416]

Cannes advantage. The intrinsic wavelength scale in an FT-NIR spectrometer provides wavelength repeatability better than one part in a million. The wave number scale of an FT-NIR spectrometer is derived from a HeNe laser that acts as an internal reference for each scan. The wave number of this laser is known very accurately and is very stable. As a result, the wave number calibration of interferometers is much more precise. If a calibration standard such as the NIST standard reference material SRM1920 is used, the wave number calibration is more accurate and has much better long-term stability than the calibration of dispersive instruments. [Pg.84]

Figure 12 Raman-REMPI (resonantly enhanced multiphoton ionization) spectrum of the °Qi(AJ = 0, AK = -2,K- 1) transitions of the Vie e2g) mode of benzene in a molecular beam. An energy-level diagram is shown for the double-resonance experiment. The ultraviolet source was tuned to 36,474 cm and the Raman wave-number calibration is adjusted to match the / = 6 line reported in Ref. 109. The expansion was 13% benzene in argon at 80 kPa and the sampling was done at XfD = 175 (D = 0.20 mm nozzle diameter) using pump and Stokes laser energies of 2 and 0.5 mJ. (From Ref. 117, with permission.)... Figure 12 Raman-REMPI (resonantly enhanced multiphoton ionization) spectrum of the °Qi(AJ = 0, AK = -2,K- 1) transitions of the Vie e2g) mode of benzene in a molecular beam. An energy-level diagram is shown for the double-resonance experiment. The ultraviolet source was tuned to 36,474 cm and the Raman wave-number calibration is adjusted to match the / = 6 line reported in Ref. 109. The expansion was 13% benzene in argon at 80 kPa and the sampling was done at XfD = 175 (D = 0.20 mm nozzle diameter) using pump and Stokes laser energies of 2 and 0.5 mJ. (From Ref. 117, with permission.)...
Databases of spectra of normal and diseased tissue need to be established, which will amply represent the spectral variance that may be encountered in practice. Because these databases will be established over long periods of time (and will most likely be updated on a regular basis), highly repeatable and reproducible instrument calibration (both wave number and intensity axis) is required. This will also facilitate the ability of data transfer from one instrument to another. On-line signal analysis techniques are needed to immediately characterize and/or classify tissue on the basis of its Raman spectrum. Various methods and techniques that can be applied for these purposes have been reviewed recently [8]. Therefore, their discussion will be omitted here. [Pg.561]

With classical least-squares regression we were able to increase the number of absorbances (dependent variable) essentially without limit, but here the absorbances are the independent variable and the number of absorbances cannot exceed the number of calibration spectra. Consequently, the wavenumbers at which the absorbance is measured should be picked with care to ensure that the absorbances at those wave-number positions are reflective of the overall contribution of those components to the spectrum. An examination of Eqs. 9.15 shows that there must be at least one analytical wavenumber for each component. Equations 9.15 represent two components consequently, there are two analytical wavenumbers, vi and V2. [Pg.211]

This full-spectrum method improves the precision over those that use only a few wave numbers. Corrections can be added for Beer s law deviations or fitting spectral baselines. However, all components present must be included in the calibration mixtures. [Pg.200]

In a variation of this type of analysis, the concentrations are expressed as functions of the various absorbances rather that vice-versa as before. This is called the inverse least squares (ILS) (or the Pmatrix method). An advantage of this method is that a quantitative analysis can be performed on some components using calibrated standards, even if some other components with unknown concentrations are present in the standards in amounts bracketing those in the samples. A disadvantage is that it is not a full-spectrum method. In the analysis, there must be at least as many standards for calibration as there are analytical wave numbers used. [Pg.200]

Cyclic voltammetry, square-wave voltammetry, and controlled potential electrolysis were used to study the electrochemical oxidation behavior of niclosamide at a glassy carbon electrode. The number of electrons transferred, the wave characteristics, the diffusion coefficient and reversibility of the reactions were investigated. Following optimization of voltammetric parameters, pH, and reproducibility, a linear calibration curve over the range 1 x 10 6 to 1 x 10 4 mol/dm3 niclosamide was achieved. The detection limit was found to be 8 x 10 7 mol/dm3. This voltammetric method was applied for the determination of niclosamide in tablets [33]. [Pg.83]


See other pages where Calibration wave number is mentioned: [Pg.310]    [Pg.312]    [Pg.310]    [Pg.312]    [Pg.289]    [Pg.172]    [Pg.178]    [Pg.16]    [Pg.326]    [Pg.331]    [Pg.37]    [Pg.168]    [Pg.635]    [Pg.117]    [Pg.155]    [Pg.56]    [Pg.57]    [Pg.179]    [Pg.5]    [Pg.115]    [Pg.166]    [Pg.173]    [Pg.266]    [Pg.267]    [Pg.267]    [Pg.182]    [Pg.176]    [Pg.194]    [Pg.1120]    [Pg.32]    [Pg.420]    [Pg.450]   
See also in sourсe #XX -- [ Pg.5 ]




SEARCH



© 2024 chempedia.info