Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration optimal time

Christian, N. P, Arnold, R. J., and Reily, J. P, Improved Calibration of Time-of-Flight Mass Spectra by Simplex Optimization of Electrostatic Ion Calculations, Anal. Chem., 11, 3317, 2000. [Pg.516]

Christian, N.P., Arnold, R.J., and Reilly, J.P. (2000) Improved calibration of time-of-flight mass spectra by simplex optimization of electrostatic ion calculations. Anal. Chem., 71,... [Pg.98]

To achieve a steady state, simulation time was set to 60 days which is six times of the SRT. The settling was assumed to ideal settling. A stepwise calibration nethdology was applied to the SBR process and some key parameters in ASM2d were optimized (results not shown). [Pg.402]

Often, it is not quite feasible to control the calibration variables at will. When the process under study is complex, e.g. a sewage system, it is impossible to produce realistic samples that are representative of the process and at the same time optimally designed for calibration. Often, one may at best collect representative samples from the population of interest and measure both the dependent properties Y and the predictor variables X. In that case, both Y and X are random, and one may just as well model the concentrations X, given the observed Y. This case of natural calibration (also known as random calibration) is compatible with the linear regression model... [Pg.352]

Optimizing the GC instrument is crucial for the quantitation of sulfentrazone and its metabolites. Before actual analysis, the temperatures, gas flow rates, and the glass insert liner should be optimized. The injection standards must have a low relative standard deviation (<15%) and the calibration standards must have a correlation coefficient of at least 0.99. Before injection of the analysis set, the column should be conditioned with a sample matrix. This can be done by injecting a matrix sample extract several times before the standard, repeating this conditioning until the injection standard gives a reproducible response and provides adequate sensitivity. [Pg.576]

Sample preparation, injection, calibration, and data collection, must be automated for process analysis. Methods used for flow injection analysis (FLA) are also useful for reliable sampling for process LC systems.1 Dynamic dilution is a technique that is used extensively in FIA.13 In this technique, sample from a loop or slot of a valve is diluted as it is transferred to a HPLC injection valve for analysis. As the diluted sample plug passes through the HPLC valve it is switched and the sample is injected onto the HPLC column for separation. The sample transfer time typically is determined with a refractive index detector and valve switching, which can be controlled by an integrator or computer. The transfer time is very reproducible. Calibration is typically done by external standardization using normalization by response factor. Internal standardization has also been used. To detect upsets or for process optimization, absolute numbers are not always needed. An alternative to... [Pg.76]

CV or bootstrap is used to split the data set into different calibration sets and test sets. A calibration set is used as described above to create an optimized model and this is applied to the corresponding test set. All objects are principally used in training set, validation set, and test set however, an object is never simultaneously used for model creation and for test. This strategy (double CV or double bootstrap or a combination of CV and bootstrap) is applicable to a relatively small number of objects furthermore, the process can be repeated many times with different random splits resulting in a high number of test-set-predicted values (Section 4.2.5). [Pg.123]

Different selections of calibration and test data set may lead to different answers for the errors. In the following, we present results from one random split however, in the final overall comparison (Section 5.8.1.8) the evaluation scheme is repeated 100 times to get an idea of the distribution of the test error for the optimal parameter choice. [Pg.250]

On the other hand, it is possible to measure even in non-linear regions of the calibration curve if the optimal measuring conditions are carefully maintained, the relative error of measurement usually does not exceed about 20%. At very low concentrations, semiquantitative procedures can be employed for example, the sample is compared with standards and the direction of the drift of the unstabilized potential indicates whether the sample concentration is higher or lower than that in the standard [147, 162). It is necessary to bear in mind that the ISE response at very low concentrations is generally slow and the potential is unstable, so that potential values read after a certain, fixed time interval must often be used instead of stabilized values. [Pg.103]

The crucial intuition driving these calibration results is the incremental nature of most addictive behavior. At each point in time, people choose whether to indulge now, and the cumulative effect of these decisions determines whether people get and remain addicted. With self-control problems, a sequence of incremental decisions can lead to behavior very different from how people would behave if committing up front to a lifetime path of behavior. In a rational choice model, in contrast, the incremental nature of addiction is irrelevant. If people know exactly what the future holds, and have no self-control problems, then people become addicted only if that is the optimal lifetime path of behavior. [Pg.199]

What does optimization mean in an analytical chemical laboratory The analyst can optimize responses such as the result of analysis of a standard against its certified value, precision, detection limit, throughput of the analysis, consumption of reagents, time spent by personnel, and overall cost. The factors that influence these potential responses are not always easy to define, and all these factors might not be amenable to the statistical methods described here. However, for precision, the sensitivity of the calibration relation, for example (slope of the calibration curve), would be an obvious candidate, as would the number of replicate measurements needed to achieve a target confidence interval. More examples of factors that have been optimized are given later in this chapter. [Pg.69]

An easy calibration strategy is possible in ICP-MS (in analogy to optical emission spectroscopy with an inductively coupled plasma source, ICP-OES) because aqueous standard solutions with well known analyte concentrations can be measured in a short time with good precision. Normally, internal standardization is applied in this calibration procedure, where an internal standard element of the same concentration is added to the standard solutions, the samples and the blank solution. The analytical procedure can then be optimized using the internal standard element. The internal standard element is commonly applied in ICP-MS and LA-ICP-MS to account for plasma instabilities, changes in sample transport, short and long term drifts of separation fields of the mass analyzer and other aspects which would lead to errors during mass spectrometric measurements. [Pg.193]

Fig. 3. Temperature-time curve for standardizing silver-enhancement Silver-enhancement time in IGSS increases with lower operating temperature The use of a calibration curve assists in optimizing the enhancement procedure... Fig. 3. Temperature-time curve for standardizing silver-enhancement Silver-enhancement time in IGSS increases with lower operating temperature The use of a calibration curve assists in optimizing the enhancement procedure...

See other pages where Calibration optimal time is mentioned: [Pg.680]    [Pg.341]    [Pg.138]    [Pg.255]    [Pg.165]    [Pg.371]    [Pg.54]    [Pg.1152]    [Pg.357]    [Pg.139]    [Pg.566]    [Pg.179]    [Pg.145]    [Pg.188]    [Pg.251]    [Pg.215]    [Pg.445]    [Pg.157]    [Pg.420]    [Pg.286]    [Pg.131]    [Pg.441]    [Pg.453]    [Pg.516]    [Pg.40]    [Pg.4]    [Pg.153]    [Pg.34]    [Pg.658]    [Pg.201]    [Pg.14]    [Pg.109]    [Pg.658]    [Pg.272]    [Pg.15]    [Pg.376]    [Pg.18]    [Pg.154]   
See also in sourсe #XX -- [ Pg.15 , Pg.18 , Pg.134 , Pg.135 , Pg.150 , Pg.229 , Pg.261 ]




SEARCH



Calibration optimizing

Time calibration

© 2024 chempedia.info