Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Blank variability

A variability risk table (as shown in Figure 2.28 for the eover support leg analysis above) is a more elfieient and traeeable way of presenting the results of the first part of the analysis. A blank variability risks results table is provided in Appendix VII. It eatalogues all the important design information, sueh as the toleranee plaeed on the eharaeteristie, the eharaeteristie dimension itself, surfaee roughness value, and then allows the praetitioner to input the results determined from the variability risks analysis, both manufaeturing and assembly. [Pg.81]

The detection limit for measuring a species by filter collection and subsequent analysis is determined either by the uncertainty in the filter blank, that is, the variability in the amount of a species in an unexposed filter, or by the limit of detection imposed by the analytical method. Most frequently, the limit of detection is determined by blank variability, thus care in preparation and handling of filters is important when using such systems for aircraft sampling, particularly if sampling is conducted in remote areas where concentrations are low. In this regard, it is crucial to establish the blank... [Pg.126]

Note that the differences lie in the choice of a and 3 values. The Important thing is not how these definitions differ, but rather that which they have in common. All are based on a consideration of the variability of the blank and an assumption that sample variability is the same as blank variability at low concentrations. [Pg.85]

The Plackett-Burmaa plan, which is based on statistics and combinatorial analysis, allows N— 1 effects of variables to be determined simultaneously in N tests. In this highly simplified test plan, only the main effects can be determined numerically at the same time, error estimation is performed by means of a blank variable. Interactions between the variables can not be determined. A test matrix for seven variables is shown in Table 13-11. [Pg.375]

The blank variable G also exhibited an effect and therefore could not be used for estimating the standard deviation. The reason for this is probably that the highly simplified test plan did not take any interaction effects into account. Hence the standard deviation of s = 0.9 known from other test series was used for tiie significance test (7-test) ... [Pg.375]

Line 3 There must be a blank line between the list of atoms and the list of variables. [Pg.73]

Line 8 The second blank line sets aside variables that are not to be optimized in the geometry optimization. [Pg.74]

Molecule Specification blank line Variables Section blank line... [Pg.285]

This file introduces the concept of variables within the molecule specification. Here, variables are simply named constants variable names are substituted for literal values within the Z-matrix, and their values are defined in a separate section following it. The two sections are separated by a blank line, or a line with a blank in the first column and the label Variables placed elsewhere on it (this is one exception Gaussian makes in its requirement for completely blank lines). ... [Pg.291]

Variables represent labels for the internal degrees of freedom being specified for the structure. As we note, the values of the variables ate defined in a separate section below the Z-matrix. It is also possible to specify a third section for constants, which immediately follows the variables section its separator line is either blank or it contains the label Constants preceded by one or more spaces. In traditonal Opt=Z-AAalrix geometry optimiaations, the values of variables are optimized while the values of constants remain fixed throughout. [Pg.291]

In our application of the transformation given in Equation 1 to these data, we restrict t to positive values. This restriction is based on models of the various sources of processing contamination. Possible sources of contamination include the chemical reagents which might add a constant level to the blank and air borne particles which might add a variable level with positive skewness. There does not seem to be any reason to include a constant level that is negative. Therefore, we have adopted this restriction. [Pg.126]

E. Procedure. The procedure section should unambiguously describe the stepwise preparation of samples, standards, and blanks. Instrumental variables should be described. Weight and volume measurements should include the acceptable range. The procedure should also include methods for any calculations. Procedures should include, but are not limited to, the following recommended elements ... [Pg.87]

Standards and blanks are the usual controls used in analytical HPLC. Standards are usually interspersed with samples to demonstrate system performance over the course of a batch run. The successful run of standards before beginning analysis demonstrates that the system is suitable to use. In this way, no samples are run until the system is working well. Typically, standards are used to calculate column plate heights, capacity factors, and relative response factors. If day-to-day variability has been established by validation, the chromatographic system can be demonstrated to be within established control limits. One characteristic of good science is that samples... [Pg.44]

Operationally, such a plot can be obtained by subtracting a titration curve containing no sample ( blank titration left curve in Fig. 3.1a) from a titration curve with sample (right curve in Fig. 3.1a) at fixed values of pH. The resultant difference plot is shown in Fig. 3.1b. The plot is then rotated (Fig. 3. Id), to emphasize that % is the dependent variable and pH is the independent variable [163], The volume differences can be converted to proton counts as described in the preceding paragraph, to obtain the final form, shown in Fig. 3.Id. [Pg.27]

Carpenter and Smith [9] and Wong [8] pointed out that iodate is also a natural constituent of seawater. Thus it may cause a variable positive blank... [Pg.123]

It is slow and tedious for an experiment in which a molecular absorption spectrum is measured. It is slow and tedious because the wavelengths are manually scanned in small increments, and each time the wavelength is changed, the calibration step with the blank needs to be performed due to the variability of light intensity from the light source at the different wavelengths. [Pg.521]

CONTINUE INTRODUCING EQUATION NUMBERS AND VARIABLE NUMBERS UNTIL ALL THE NONZERO ELEMENTS HAVE BEEN ACCOUNTED FOR. EACH CARD MUST BE FILLED COMPLETELY, EXCEPT THE LAST CARD WHICH MUST HAVE TWO BLANK FIELDS, OR THE READING WILL STOP. [Pg.237]

AS ITERATES. IN THE FIRST FIELD ENTER A VARIABLE NUMBER WHICH SHOULD NOT BE CHOSEN AS AN ITERATE. IN THE SECOND FIELD ENTER ANOTHER VARIABLE NUMBER WHICH SHOULD NOT BE AN ITERATE. AT LEAST ONE OF THESE CARDS IS REQUIRED. EVEN IF NO VARIABLES ARE TO BE EXCLUDED AS ITERATES. THIS MAY NECESSITATE ADDING A BLANK CARD AS DATA. [Pg.237]

Fenvalerate Detection Limits. To the extent that detection limits require knowledge of the calibration curve and random error (for x) as a function of concentration, all of the foregoing discussion is relevant — both for detection and estimation. However, curve shape and errors where x x, are relatively unimportant at the detection limit, in contrast to direct observations of the initial slope and the blank and its variability. (It will be seen that the initial observation in the current data set exceeded the ultimate detection limit by more than an order of magnitude )... [Pg.63]

Note The purchaser may consider the use of variable speed drive capability and/or the use of blank stages (to add impellers in the future) for multistage pumps to meet this requirement. [Pg.19]


See other pages where Blank variability is mentioned: [Pg.338]    [Pg.375]    [Pg.376]    [Pg.401]    [Pg.401]    [Pg.338]    [Pg.375]    [Pg.376]    [Pg.401]    [Pg.401]    [Pg.28]    [Pg.314]    [Pg.200]    [Pg.619]    [Pg.619]    [Pg.648]    [Pg.693]    [Pg.764]    [Pg.153]    [Pg.359]    [Pg.247]    [Pg.195]    [Pg.231]    [Pg.480]    [Pg.336]    [Pg.315]    [Pg.49]    [Pg.58]    [Pg.72]    [Pg.128]    [Pg.446]    [Pg.154]   
See also in sourсe #XX -- [ Pg.22 , Pg.188 ]




SEARCH



Blank

Blank responses, variability

Blank, blanking

Blanking

Variable blank

Variable blank

© 2024 chempedia.info