Argon detector mechanism

The electron capture detector is the result of a series of developments which were initiated in 1951 by D. J. Pompeo and J. W. Otvos (14) of the Shell Company s development laboratory in California. The device they invented was a beta-ray ionization cross-section detector (Section 5.8). Deal et al. (15) at the Shell laboratory in California and Boer (16) in Amsterdam modified the detector, used originally to monitor effluents of a large scale plant process, for applications in GC. From the limited success of the detector. Lovelock (17) produced the beta-ray argon detector in 1958 (Section 5.8). This modification substituted argon as the carrier gas and placed a potential of 1000 V across the electrodes. Argon passing between the electrodes absorbed radiation and formed a metastable species with energy (11.6 eV) sufficient to ionize most substances. Proposed mechanisms for this process are ... [Pg.255]

At the 1960 symposium, research into the mechanism of detection of a number of the more important GC detectors was reported. The FID was carefully examined by Ongkiehong [19] and Desty et al. [20], the different argon detectors were described in detail by Lovelock [21] and the performance of the FID and Argon Detectors were compared by Condon et al. [22]. The integral and differential ionization detectors were also discussed by Matousek [23]. At the end of the 1960 symposium, the majority of the GC detectors that are used today had been described and their function explained and understood. Since that time few, if any, new GC detectors have been introduced and innovations that have been described have been largely extensions and improvements of older concepts. [Pg.94]

Elemental composition C 12.14%, 0 16.17%, Cl 71.69%. Phosgene can be analyzed by GC using FID or a balogen-specific detector or by GC/MS. Ambient air may be collected in a metal container placed in an argon bath or condensed into any other type cryogenically cooled trap. Alternatively, the air may be collected in a Tedlar bag. The sampled air may be sucked by a condensation mechanism into tbe GC column. [Pg.195]

The AED employs a microwave-induced He plasma to dissociate eluted analyte molecules to their component atoms and excite them to emit at characteristic wavelengths. This is very similar to the mechanism in the argon plasma inductively coupled plasma source (cf. Section 7.3.1). A spectrometer with a diode array detector (Figure 7.26b and c) isolates and measures the intensity of sensitive emission lines unique to each element. Depending on the relative sensitivity and proportion of atoms in the molecules, separate element response channels may display peaks in several element-selective chromatograms. These data may be combined with retention... [Pg.904]


See other pages where Argon detector mechanism is mentioned: [Pg.369]    [Pg.545]    [Pg.37]    [Pg.58]    [Pg.140]    [Pg.140]    [Pg.366]    [Pg.780]    [Pg.222]    [Pg.904]    [Pg.239]    [Pg.195]    [Pg.387]    [Pg.170]    [Pg.127]    [Pg.276]   
See also in sourсe #XX -- [ Pg.120 ]