Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analyst 1.4.1 software

API 4000 triple quadrupole mass spectrometer containing a Thrbolonspray interface and Analyst 1.4.1 software (AB/MSD-Sciex)... [Pg.180]

Figure 5.7 Control chart plot using QI Analyst software (Cu in mg/kg on y-axis). Figure 5.7 Control chart plot using QI Analyst software (Cu in mg/kg on y-axis).
Validation of the migration of electronic records generated using MassChrom software on the Macintosh systems to the new Analyst NT environment, as well as data aequisition on some Macintosh platforms with interpretation using Analyst software... [Pg.498]

The MassChrom mass spectrometer software currently used in the Department is a combination of data acquisition software (three versions of RAD and sample control) and data processing software (two versions) that operates on the Macintosh plus the Analyst software designed for the Windows NT environment. The RAD and MacQuan software rutming on the Macintosh Quadra will be retired under the work described in this chapter. [Pg.498]

Over time, the instrument has been used more widely by chromatographers, and the Analyst software is a response to this as the operation of the application is simpler and uses chromatographic terms more than mass spectrometry ones. This difference in design philosophy is a complicating factor for the data migration as terms have to be mapped between the applications, which we will describe later in this chapter. [Pg.500]

Li et al. [40-42] used a QTrap system to incorporate both the conventional SRM-only acquisition of parent compounds and the SRM-triggered IDA of potential metabolites within the same scan cycle during the same LC-MS/MS run in plasma sample analysis. The fast scanning capability of the LIT allowed for the IDA of metabolite MS/MS spectra <1 s/scan, in addition to the collection of adequate data points for SRM-only channels. A SRM survey scan containing 30-150 SRM channels was established by running a script in Analyst software. If the intensities in these... [Pg.237]

Using the analyst software and the IDA Explorer to display IDA samples feature, select the acquired files using Open data file (see Note 14 and Fig. 1). [Pg.134]

On the last released versions of the Analyst software, dynamic background subtraction can be selected with a single click. An independent script is needed for older versions. [Pg.137]

AB Sdex Analyst Software is used to control analytical equipment during sample analysis. The instrument is allowed to equilibrate for at least 15 min prior to sample analysis. An MP-A injection is always performed prior to injections of samples, standard curve, or QC material. Injections of QC material at the beginning and end of the run (same samples injected twice) are advisable in order to ensure minimal change in detection during a run. [Pg.217]

AB Sciex Analyst Software is used for quantitation. The ratio of the peak area of each analyte in the calibration material to that of the peak area of the respective internal standard is used in order to generate a standard curve across concentrations of the... [Pg.218]

Periodic fault detection is readily done by analysts without extensive software support. Process monitoring such as the examination of the traces discussed above are one exam e. However, the number of measurements in a single set have such complex interactions that it is... [Pg.2576]

Note that great care must be taken when comparing machinery vibration data to industry standards or baseline data. The analyst must make sure the frequency and amplitude are expressed in units and running speeds that are consistent with the standard or baseline data. The use of a microprocessor-based system with software that automatically converts and displays the desired terms offers a solution to this problem. [Pg.692]

To enable qnantitative measurements to be made, the analyst requires the ability to determine the areas or heights of the detector responses of analyte(s) and any internal standard that may be present and then, from these figures, to derive the amount(s) of analyte(s) present in the unknown sample. The software provided with the mass spectrometer allows this to be done with a high degree of automation if the analyst so desires. [Pg.84]

From a quantitative perspective, each peak is defined by two parameters, i.e. the position of its baseline and the retention time boundaries, with those derived by the computer system being shown in Figure 3.27. It is not the intention of this present author to discuss how these have been determined but simply to point out that their positions may have a significant effect on the accuracy and precision of any quantitative measurements, especially, as in Figure 3.27, when the baseline is not horizontal and the signals from each of the components are not fully resolved. It is usual for the software to allow the analyst to override the parameters chosen by the computer to provide what they consider to be more appropriate peak limits and/or baseline positions. [Pg.85]

Humans should use computers to do functional work for them in the most efficient manner possible. However, we must not delude ourselves into thinking that the mere use of a computer to analyze adverse events will magically analyze these events in a systematic, efficient way. Computers do not automatically produce coherent, auditable results that can be subsequently reproduced with ease. Computers must be actively programmed through an iterative process involving tight communication between analysts and software developers until these processes are totally functional. [Pg.651]

In Section 42.2 we have discussed that queuing theory may provide a good qualitative picture of the behaviour of queues in an analytical laboratory. However the analytical process is too complex to obtain good quantitative predictions. As this was also true for queuing problems in other fields, another branch of Operations Research, called Discrete Event Simulation emerged. The basic principle of discrete event simulation is to generate sample arrivals. Each sample is characterized by a number of descriptors, e.g. one of those descriptors is the analysis time. In the jargon of simulation software, a sample is an object, with a number of attributes (e.g. analysis time) and associated values (e.g. 30 min). Other objects are e.g. instruments and analysts. A possible attribute is a list of the analytical... [Pg.618]

Assessing the resources available for method development should also be done before beginning a project. The resources available include not only HPLCs, detectors, and columns, but also tools for sample preparation, data capture and analysis software, trained analysts, and especially samples representative of the ultimate analyte matrix. Also, it should be considered whether a fast, secondary method of analysis can be used to optimize sample preparation steps. Often, a simple colorimetric or fluorimetric assay, without separation, can be used for this purpose. A preliminary estimate of the required assay throughput will help to guide selection of methods. [Pg.28]

Analyst to conduct multipathway human health risk assessments and food-web based ecological risk assessment modeling. BREEZE risk analyst combines databases, GIS functionality, fate, transport, and exposure modeling equations into one software application... [Pg.103]

Analysts like to make chemical measurements to the highest quality they can, and this is possible with modem instrumentation and good analytical practice. Chemical parameters are generally measured more accurately and reproducibly than parameters measured in biological systems, and the results are more reliable than values calculated by software from structure alone (e.g., calculated log P, pKa). Of all the information available to medicinal chemists, measured chemical parameters are probably the most reliable. With these thoughts in mind, it makes sense not to... [Pg.39]


See other pages where Analyst 1.4.1 software is mentioned: [Pg.143]    [Pg.380]    [Pg.504]    [Pg.238]    [Pg.50]    [Pg.219]    [Pg.2897]    [Pg.197]    [Pg.636]    [Pg.2275]    [Pg.2277]    [Pg.2551]    [Pg.300]    [Pg.268]    [Pg.269]    [Pg.608]    [Pg.620]    [Pg.13]    [Pg.279]    [Pg.153]    [Pg.2]    [Pg.141]    [Pg.654]    [Pg.674]    [Pg.716]    [Pg.717]    [Pg.718]    [Pg.294]    [Pg.497]    [Pg.729]    [Pg.229]   
See also in sourсe #XX -- [ Pg.143 ]




SEARCH



Analysts

© 2024 chempedia.info