Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sample acceptance

Equation (22) allows the maximum sample volume that can be used without seriously denigrating the performance of the column to be calculated from the retention volume of the solute and the column efficiency. In any separation, there will be one pair of solutes that are eluted closest together (which, as will be seen in Part 3 of this book, is defined as the critical pair) and it is the retention volume of the first of these that is usually employed in equation (22) to calculate the maximum acceptable sample volume. [Pg.195]

Actually, it would be reasonable for the authorities to replace by 0.1% the individual limit concept for all impurities lower than about 0.1% in the accepted sample, provided that toxicity is not an issue, because otherwise undue effort would have to be directed at the smallest impurities. Various modifications, such as less stringent confidence limits, optimistic estimates (line (A) in Fig. 4.6), etc. somewhat alleviate the situation the plant manager is in, but do not change the basic facts. [Pg.199]

Acceptance sampling involves the application of a predetermined sampling plan to decide whether a batch of goods meets the defined criteria for acceptance. The main aim of any acceptance sampling must be to see that the customer gets the quality required, while remembering that financial resources are limited and that the cost of the article must reflect the cost of inspection, as well as the cost of production. [Pg.34]

Acceptance sampling can be either by attributes or by variables. In sampling by attributes, the item in the batch of product either conforms or not. The number of nonconformities in the batch is counted and if this reaches a predetermined figure, the batch is rejected. In sampling by variables, the characteristic of interest is measured on a continuous scale. If the average meets a predetermined value,... [Pg.34]

In addition, published sampling plans for different sectors indicate the number of samples to be taken from a lot. For example, when carrying out acceptance sampling by attributes (see Section 3.3.2), as described in ISO 2859-1 [6], a number of factors have to be taken into account ... [Pg.37]

Welder operator qualifications with or without filler material (autogenous welding) shall be limited to the mechanized WPS/PQR and ASME BPV Code Section IX, or API Standard 1104. Three consecutive, acceptable samples shall be required to support welder operator qualification testing. [Pg.42]

If TSE levels are verified as higher than acceptable, sampling and analysis is to be terminated with prejndice, as the data generated are nmefiable, fntile and a waste of practical effort. It is an economic waste to sample, monitor or make decisions based on a docnmentable inferior basis. [Pg.79]

Two types of seals for the vials have been employed and both give satisfactory results. Aluminum, one-piece, crimp-on seals used with serum vials and Teflon faced septa are acceptable if the seal is properly made and maintained during shipment. However, several years of experience indicates a success rate significantly less than 100% in making proper seals of this type in the field. Therefore simple screw cap vials used with the Teflon faced septa were evaluated and found to give equivalent results and a very high rate of acceptable samples. Narrow mouth screw cap bottles with Teflon-faced silicone rubber septa cap liners are strongly recommended for sample collection. [Pg.58]

Equations (67) and (68) show that the efficient columns available today tolerate much smaller sample sizes than the columns used a few years ago. Wlicicua lui culuiiiiis With t>p - 25 fiin, h = 5, and /V - 200U tlic maximum permissible sample volume Vs,m is about 60 /xl. tne microparticulate columns can accept samples of only a few microliters without deterioration of the intrinsic column efficiency. [Pg.199]

Collection of particles is based on filtration, gravitational and centrifugal sedimentation, inertial impaction and impingement, diffusion, interception, or electrostatic or thermal precipitation (e.g., see Spurny, 1986, Chapter 3). The choice of method depends on a number of parameters such as the composition and size of the particles, the purpose of the sample, and acceptable sampling rates. Table 11.10 summarizes some of the commonly used methods and the size ranges over which they are effective. [Pg.608]

One popular configuration is the Grimm source, which accepts samples in the form of discs. Such sources usually operate at 500-1000 V, 25-100 mA and 1-5 Torr, wiA detection limits of approximately 0.1 ppm. Another configuration is the hollow-cathode lamp in which the sample can be either machined as a hollow cathode, evaporated to dryness (if a solution) or pressed (if a powder) into a hollow cathode made of pure graphite. Typical operating conditions are 200-500 V, 10-100 mA, and 0.1-1.0 Torr, with detection limits in the range 0.1-10 ppm. [Pg.112]

First article inspection Acceptance sampling According to a quality plan 100% inspection... [Pg.228]

Schilling, E. Acceptance Sampling in Quality Control. 10th ed. New York Marcel Dekker, p. 7 (1982). [Pg.121]

Sa becomes the maximum acceptable sample standard deviation to meet the acceptance criteria. If the sample standard deviation sn is less than scr, then we are guaranteed, with a minimum assurance of 100(1 - a)%, that the upper prediction bound for a future sample of size 10 will not be greater than 6.0% of the target concentration. [Pg.714]

Some databases will not accept samples with the same field ID. A unique identifier (a date or a sequential number) must be added to the monitoring well ID, for example, MW-1-120701, MW-2-120701, MW-3-120701 or MW-1-001, MW-2-002, MW-3-003. [Pg.95]

For trace element analysis, if the preparation blank contains target analytes above the PQL, the lowest concentration of that analyte in a sample must be 10 times the blank concentration for a result to be acceptable. Samples with concentrations below 10 times the blank concentration must be redigested and reanalyzed with a new preparation blank. [Pg.278]

Resample using acceptable sampling plan if original sample is found to be unrepresentative. [Pg.404]

To deposit cells (or tissues) at the repositories a frozen vial or a subconfluent culture in a 25 cm2 flask filled completely with growth medium is required. In the latter case the neck of the flask should be taped and the flask wrapped in pop-film in a polystyrene box or jiffy bag. The package should be kept at room temperature and delivered by the fastest route. Depositories require documentation before they will accept samples and the appropriate paperwork must be completed before the cells are despatched. [Pg.136]

The recommended sample sizes were estimated by dividing the sensitivity by the lower end of the ambient concentration range, and assumes that the sample is diluted to a final volume of 25 ml after dissolution. For ETA devices, which can accept sample injections of different sizes, an injection of lOjul was assumed since this can be accommodated by all of the models. [Pg.134]


See other pages where Sample acceptance is mentioned: [Pg.106]    [Pg.422]    [Pg.252]    [Pg.8]    [Pg.195]    [Pg.197]    [Pg.141]    [Pg.444]    [Pg.432]    [Pg.34]    [Pg.40]    [Pg.267]    [Pg.408]    [Pg.136]    [Pg.51]    [Pg.35]    [Pg.468]    [Pg.192]    [Pg.195]    [Pg.264]    [Pg.258]    [Pg.136]    [Pg.38]    [Pg.246]    [Pg.146]    [Pg.96]    [Pg.26]    [Pg.206]    [Pg.128]    [Pg.66]    [Pg.230]    [Pg.233]   
See also in sourсe #XX -- [ Pg.135 ]




SEARCH



Acceptance-rejection-sampling

Acceptance-rejection-sampling efficiency

Candidate density acceptance-rejection-sampling

Sample acceptance criteria

Sampling acceptable

Sampling acceptable

Sampling acceptable quality level

Sampling acceptance

Sampling acceptance

Sampling acceptance number

Statistics acceptance sampling

© 2024 chempedia.info