Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Zero Detector

Distance to resolution conversions can be calculated using simple trigonometry by using the expressions (for zero detector swing) ... [Pg.83]

At first, only one pair (a, b) of operands is considered. Their upper halves are called ah and bh, the lower halfs ai andbi. The ZERO detector (see Fig. 1) decides whether the upper halves entirely consist of zeroes. To detect the number of leading zeroes, we can start a binary search from the most significant bit. This work can be done prior to the computation, e.g. in a pipelined manner. In case the upper halves don t contain even a single 1-bit, the respective adder block can be used to add a further pair of operands -which must fit into the remaining n/2 adder bits, of course (to be checked by the same method). Alternative 2 ... [Pg.184]

The probe receives a signal when either the driver or detector coil passes a flaw or other feature in the tube A signal is produced over the full length of the flaw. It is affected by geometry and permeability changes which cause the instrument zero to wander. [Pg.322]

Time-of-flight mass spectrometers have been used as detectors in a wider variety of experiments tlian any other mass spectrometer. This is especially true of spectroscopic applications, many of which are discussed in this encyclopedia. Unlike the other instruments described in this chapter, the TOP mass spectrometer is usually used for one purpose, to acquire the mass spectrum of a compound. They caimot generally be used for the kinds of ion-molecule chemistry discussed in this chapter, or structural characterization experiments such as collision-induced dissociation. Plowever, they are easily used as detectors for spectroscopic applications such as multi-photoionization (for the spectroscopy of molecular excited states) [38], zero kinetic energy electron spectroscopy [39] (ZEKE, for the precise measurement of ionization energies) and comcidence measurements (such as photoelectron-photoion coincidence spectroscopy [40] for the measurement of ion fragmentation breakdown diagrams). [Pg.1354]

In time-of-flight (TOF) mass spectrometers, a pulse of ions is accelerated electrically at zero time. Having attained a maximum velocity, the ions drift along the flight tube of the analyzer. The times of arrival of ions at a detector are noted. [Pg.410]

Fig. 25-2. Double-beam, double-pass transmissometer for measuring smoke density in stacks. A[, chopper wheel A, beam gating wheel A3, aperture D, detector Fj, spectral filter F2, solenoid-activated neutral density filter L, lamp M, half-mirror/beam splitter Rj, solenoid-activated zero calibration reflector R2, retroreflector (alignment bullseye not shown). Design patented. Source Drawing courtesy of Lear Siegler, Inc. Fig. 25-2. Double-beam, double-pass transmissometer for measuring smoke density in stacks. A[, chopper wheel A, beam gating wheel A3, aperture D, detector Fj, spectral filter F2, solenoid-activated neutral density filter L, lamp M, half-mirror/beam splitter Rj, solenoid-activated zero calibration reflector R2, retroreflector (alignment bullseye not shown). Design patented. Source Drawing courtesy of Lear Siegler, Inc.
Calibrate the detector tube pump for proper volume measurement at least quarterly. Simply connect the pump directly to the bubble meter with a detector mbe in-line. Use a detector mbe and pump from the same manufacturer. Wet the inside of the 100 cc bubble meter with soap solution. For volume calibration, experiment to get the soap bubble even with the zero ml mark of the buret. For piston-type pumps, pull the pump handle all the way out (full pump stroke) and note where the soap bubble stops for bellows-type pumps, compress the bellows fully for automatic pumps, program the pump to take a full pump stroke. [Pg.249]

The contribution of the equipment between injection unit and detector cell should be negligable in relation to the column for a sufficient column characterization short connections with narrow capillaries and zero dead volume unions are the precondition for reliable plate numbers. Every end fitting of a column causes additional band broadening. In the past a column type was offered that could be directly combined without any capillary links unfortunately, it has disappeared from the market. [Pg.434]

Two-position detectors such as thermostats can be fitted with an anticipatory bias to reduce the amount of overshoot. In such instruments, a small bias heater accelerates the action of the control. An alternative method to reduce overshoot is to introduce a timing device so that it acts intermittently. Where the two-position sensor is also the controller it provides only two plant outputs, maximum or zero. [Pg.325]

Potential of zero charge, 20, 23, 25, 66 Potential scanning detector, 92 Potential step, 7, 42, 60 Potential window, 107, 108 Potentiometry, 2, 140 Potentiometric stripping analysis, 79 Potentiostat, 104, 105 Preconcentrating surfaces, 121 Preconcentration step, 121 Pretreatment, 110, 116 Pulsed amperometric detection, 92 Pulse voltammetry, 67... [Pg.208]

In real curvature sensors, a vibrating membrane mirror is placed at the telescope focus, followed by a collimating lens, and a lens array. At the extremes of the membrane throw, the lens array is conjugate to the required planes. The defocus distance can be chosen by adjusting the vibration amplitude. The advantage of the collimated beam is that the beam size does not depend on the defocus distance. Optical fibers are attached to the individual lenses of the lens array, and each fiber leads to an avalanche photodiode (APD). These detectors are employed because they have zero readout noise. This wavefront sensor is practically insensitive to errors in the wavefront amplitude (by virtue of normahzing the intensity difference). [Pg.190]


See other pages where Zero Detector is mentioned: [Pg.247]    [Pg.72]    [Pg.277]    [Pg.277]    [Pg.8]    [Pg.69]    [Pg.8]    [Pg.314]    [Pg.194]    [Pg.331]    [Pg.247]    [Pg.72]    [Pg.277]    [Pg.277]    [Pg.8]    [Pg.69]    [Pg.8]    [Pg.314]    [Pg.194]    [Pg.331]    [Pg.1165]    [Pg.1308]    [Pg.1574]    [Pg.2955]    [Pg.236]    [Pg.500]    [Pg.55]    [Pg.369]    [Pg.394]    [Pg.68]    [Pg.296]    [Pg.256]    [Pg.267]    [Pg.275]    [Pg.310]    [Pg.273]    [Pg.1132]    [Pg.33]    [Pg.297]    [Pg.417]    [Pg.75]    [Pg.85]    [Pg.311]    [Pg.153]    [Pg.171]    [Pg.384]    [Pg.34]    [Pg.190]   


SEARCH



© 2024 chempedia.info