Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Transmissive mode operation

Operating Modes of Liquid Crystal Displays 2.3.1. Transmissive Mode Operation... [Pg.103]

Then a hostaphan foil coated with A1 for reflective mode operation or In203 for transmissive mode operation is rolled on. Finally the display is glued liquid-tight around the circumference. Experimental displays of this type have functioned well for some time, and they demonstrate the basic feasibility of a large-area liquid crystal picture screen using foils. However, the following problems became apparent ... [Pg.111]

In many ways the nanocrystal characterization problem is an ideal one for transmission electron microscopy (TEM). Here, an electron beam is used to image a thin sample in transmission mode [119]. The resolution is a sensitive fimction of the beam voltage and electron optics a low-resolution microscope operating at 100 kV might... [Pg.2903]

The requirement of single-mode operation is that for each component an each tjq)e of interaction (reflection, beam splitting, transmission), one has... [Pg.318]

Fig. 3.19 Schematic illustration of the measurement geometry for Mossbauer spectrometers. In transmission geometry, the absorber (sample) is between the nuclear source of 14.4 keV y-rays (normally Co/Rh) and the detector. The peaks are negative features and the absorber should be thin with respect to absorption of the y-rays to minimize nonlinear effects. In emission (backscatter) Mossbauer spectroscopy, the radiation source and detector are on the same side of the sample. The peaks are positive features, corresponding to recoilless emission of 14.4 keV y-rays and conversion X-rays and electrons. For both measurement geometries Mossbauer spectra are counts per channel as a function of the Doppler velocity (normally in units of mm s relative to the mid-point of the spectrum of a-Fe in the case of Fe Mossbauer spectroscopy). MIMOS II operates in backscattering geometry circle), but the internal reference channel works in transmission mode... Fig. 3.19 Schematic illustration of the measurement geometry for Mossbauer spectrometers. In transmission geometry, the absorber (sample) is between the nuclear source of 14.4 keV y-rays (normally Co/Rh) and the detector. The peaks are negative features and the absorber should be thin with respect to absorption of the y-rays to minimize nonlinear effects. In emission (backscatter) Mossbauer spectroscopy, the radiation source and detector are on the same side of the sample. The peaks are positive features, corresponding to recoilless emission of 14.4 keV y-rays and conversion X-rays and electrons. For both measurement geometries Mossbauer spectra are counts per channel as a function of the Doppler velocity (normally in units of mm s relative to the mid-point of the spectrum of a-Fe in the case of Fe Mossbauer spectroscopy). MIMOS II operates in backscattering geometry circle), but the internal reference channel works in transmission mode...
Figure 3.9 Conceptual view of tandem mass spectrometry with a tandem-inspace triple quadrupole mass analyzer." The first mass analyzer (Ql) selects the precursor ion of interest by allowing only it to pass, while discriminating against all others. The precursor ion is then fragmented, usually by energetic collisions, in the second quadrupole (q2) that is operated in transmissive mode allowing all fragment ions to be collimated and passed into the third quadrupole (Q3). Q3 performs mass analysis on the product ions that compose the tandem mass spectra and are rationalized to a structure. Figure 3.9 Conceptual view of tandem mass spectrometry with a tandem-inspace triple quadrupole mass analyzer." The first mass analyzer (Ql) selects the precursor ion of interest by allowing only it to pass, while discriminating against all others. The precursor ion is then fragmented, usually by energetic collisions, in the second quadrupole (q2) that is operated in transmissive mode allowing all fragment ions to be collimated and passed into the third quadrupole (Q3). Q3 performs mass analysis on the product ions that compose the tandem mass spectra and are rationalized to a structure.
CW) output of microwaves capable of sweeping 10 MHz to 47 GHz. Depending upon the physical dimensions of the resonator and their coupling with the source, this spectrum analyzer can operate to detect the microwave absorption profile of the resonator either in the reflection or in the transmission mode. In this work the frequency range... [Pg.356]

X-ray Absorption Spectra (XAS). X-ray absorption measurements were performed at station 9.2 of the SRS at Daresbury (UK) with an electron beam energy of 2 GeV and a stored current varying between 290 and 160 mA. The wiggler was operational at 5.0 Tesla. Data were collected in the transmission mode from 11.37 keV to 13.43 keV (Pt Li -edge 11.564 keV, Pt Vedge 13.273 keV) with a Si (220) monochromator detuned to 50 % of the maximum intensity... [Pg.300]

This work demonstrated a number of new results and opportunities for ultrafast XAS 1) it is possible to work with highly dilute solutions in transmission mode without dramatic loss of signal-to-noise ratio. This is very promising as one can envision the study of samples, for which large concentrations are impossible to reach. Biologically relevant samples are usually investigated in solutions with up to 1 mmol/1 concentration, and we therefore can envision such studies on the ultrafast time scales in the near future. 2) It is possible to scan the time delay between the laser pump pulse and the x-ray probe pulse, and therefore follow the evolution of the system from the start. 3) It also demonstrated the operation of an optical-x-ray cross-correlator (Fig. 6.b). The time resolution is not a limiting factor and the experiments are feasible with sources of shorter x-ray pulses, provided the flux is not too low. [Pg.361]

The quadruple Q1 is set to operate in wideband transmission mode with a low mass cutoff — ... [Pg.182]

The 1.7% Co2+ ZnO DMS-QDs in Figure 34 were also examined by Zeeman spectroscopy in transmission mode. An average band-edge Zeeman shift of 53 cm 1/7 over the range of 0-7 T ( Fig. 35) was measured. The Zeeman data were analyzed in the mean-field approximation using Eq. 11, where x is the dopant mole fraction and (Sz) is the expectation value of the Sz operator of the spin Hamiltonian. 7/oa and iVop quantify the exchange interactions between the dopant and unpaired spins in the conduction (CB) and valence bands (VB), respectively (19, 159-161). [Pg.108]

In the usual transmission mode of operation, surface sensitivity can be achieved by examining very small particles (clOnm) or thin foils (<2.5nm) but an alternative approach involving backscattcrcd electrons or fluorescent X-rays is under development. [31]... [Pg.559]

EXAFS data (Rh K-edge ((23220 eV) or Ir Lm-edge (13419 eV)) were collected in transmission mode on station 9.2 of the Daresbury Synchrotron Radiation Source, operating at 2 GeV with an average current of 150 mA. A water-cooled Si(220) double crystal monochromator was used, with its angle calibrated by running an edge scan of a 5 pm Rh or Ir foil. For each sample 2-10 scans were recorded at room temperature in the... [Pg.174]

A large background power, which is an ineluctable consequence of the transmission mode, can dramatically reduce the sensitivity of diodes operated as detectors. The signal current that flows in the detector is given by (Schneider, 1982)... [Pg.293]

XPS analysis was run by an ESCA LAB-200X spectrometer operating in Fixed Analyzer Transmission Mode. [Pg.457]

The principle involves using X-rays (beam of electrons) to image atomic structures. The very short wavelengths allow much higher resolution than with a standard optical microscope. There are several types of electron microscope which allow different types of image to be formed. They operate in either transmission mode (electrons pass through the sample) or reflection mode (electrons are reflected from the surface)... [Pg.71]

In studies conducted at Norwich by Sene and coworkers [67], a BioRad FTS-40 FT-IR instrument interfaced to a Spectra-Tech IR PLAN microscope was used, with sample areas of 100 x 100 pm. The operation was conducted in transmission mode on Bap2 windows, and the spot size used was compatible with areas of the single cell waU fragments of 0.1-0.2 pm thickness. Pectic polysaccharides extracted... [Pg.253]


See other pages where Transmissive mode operation is mentioned: [Pg.128]    [Pg.156]    [Pg.172]    [Pg.121]    [Pg.164]    [Pg.164]    [Pg.128]    [Pg.156]    [Pg.172]    [Pg.121]    [Pg.164]    [Pg.164]    [Pg.1716]    [Pg.203]    [Pg.29]    [Pg.145]    [Pg.375]    [Pg.428]    [Pg.135]    [Pg.109]    [Pg.160]    [Pg.50]    [Pg.51]    [Pg.126]    [Pg.410]    [Pg.120]    [Pg.650]    [Pg.121]    [Pg.129]    [Pg.393]    [Pg.633]    [Pg.6]    [Pg.28]    [Pg.55]    [Pg.51]    [Pg.6]    [Pg.52]    [Pg.297]    [Pg.819]   
See also in sourсe #XX -- [ Pg.103 , Pg.104 , Pg.156 ]




SEARCH



Operating mode

Operation mode

Operation modes mode

Transmission mode

© 2024 chempedia.info