Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Toluene, external standard

The determination of the relationship between detector response and the sample concentration is termed the calibration of the method. There are two types of methods in use for the quantitative analysis of a sample, i.e., the external standard and the internal standard method. An external standard method is a direct comparison of the detector response of a pure compound (standard) to a sample.2 The calibration of the method is performed by preparing standards of varying concentration and analyzing them by a developed method. Method 1 (below) was developed for toluene, and standards of varying concentration were prepared and analyzed. The results obtained are summarized in Table 2 see Figure 3. [Pg.156]

For the example of toluene given above, the external standard method can be converted into an internal standard method by adding anisole (an appropriate internal standard) to both standard and sample. The retention time of anisole is 4.5 minutes if analyzed by the method above. To calibrate the internal standard method for toluene, toluene standards of concentration 0.3 to 1.5 mg/ml containing 0.5 mg/ml anisole were prepared. The detector response as a function of the amount of sample injected is shown in Figure 4B. [Pg.160]

For general purpose tracer work, however, and particularly in polymer chemistry, the liquid scintillation counter surpasses all other instruments in its sensitivity and adaptability. There is no question on the author s mind that at the present time such an instrument would be the first choice, particularly where tritium, carbon-14 or sulphur-35 were involved. Samples for assay are dissolved in a phosphor whose major solvent usually consists of toluene, toluene-alcohol, or dioxan. Many polymers and low molecular weight compounds are readily soluble in these solvents. Prospective users should not be deterred by alleged complications due to "variable quench effects" as these effects are readily corrected for via internal or external standards or the channels ratio method (7, 46, 91). Dilution quench corrections, though valid, are tedious and unnecessary. Where samples are insoluble in phosphor they may be suspended (e.g. as gels or as paper cut from chromatograms, etc.) or they can be burnt and the combustion products absorbed in a suitable phosphor solution. A modification of the Schoniger flask combustion technique is particularly suitable for this purpose (43—45). [Pg.134]

The final analytical method for the simultaneous determination of A9-THC and its metabolites consists of the following sequence the cannabinoids are extracted from plasma with toluene they are then back extracted from toluene into Claisen s alkali the Claisen s alkali is diluted with water, tetrahexyl ammonium hydroxide is added and the alkaline solution is extracted with methylene chloride containing ethyl iodide. The overall recoveries were 45% for A9-THC and 83% for 11-hydroxy-A9-THC. External standards (l-0-ethyl-A9-THC and l-0-ethyl-ll-hydroxy-A9-THC) were added to the methylene chloride phase followed by a small amount of Florosil, which absorbed the tetra-hexylammonium hydroxide and tetrahexylammonium iodide. The methylene chloride was decanted and evaporated. [Pg.90]

The samples were counted in a liquid scintillator spectrometer. Quenching was monitored by employing automatic external standard ratio. The counting efficiency for carbon-14 was measured by using toluene- C as a reference standard. The efficiency factor was used to convert the collected data in counts per minute (cpm) to disintegration per minute (dpm). [Pg.481]

Compare the chromatogram obtained with external standards. In a standard solution, which may conveniently be prepared with toluene or hexane, the following active substances (in picograms per pi) give approximately the same peak heights lindane and other HCH isomers 30, heptachlor 30, HCB 40, aldrin 40, heptachlorepoxide 80, dieldrin 80, O -endosulfan 100, DEE 120, B-endosulfan 200, DDD 250, DDT 500, methoxychlor 500. The peak area is used as the basis for evaluation. [Pg.592]

This test method covers the determination of the total nonaromatic hydrocarbons and trace monocyclic aromatic hydrocarbons in toluene and mixed xylenes by gas chromatography. The purity of toluene or mixed xylenes can also be calculated. Calibration of the gas chromatographic system is done by the external standard calibration technique. A similar test method, using the internal standard calibration technique, is Test Method D 2360. [Pg.1005]

Fig. 6-7. The external magnetic field dependence of AR(B) observed for the singlet-sensitized decomposition of DBF in toluene at room temperature ( ) for PB, (o) for o-MB, and (x) a mixture of p-MB and DPE. The standard deviation of each AR(B) is 3 %. (Reproduced from Ref. [21a] by permission from Elsevier Science B. V.)... Fig. 6-7. The external magnetic field dependence of AR(B) observed for the singlet-sensitized decomposition of DBF in toluene at room temperature ( ) for PB, (o) for o-MB, and (x) a mixture of p-MB and DPE. The standard deviation of each AR(B) is 3 %. (Reproduced from Ref. [21a] by permission from Elsevier Science B. V.)...
Because of thermodynamic and electrochemical conventions, standard potentials are defined in the direction of reduction, independently of the respective chemical stabilities of the molecules involved. Thus for the oxidation of toluene to its cation radical, E° refers to the reduction of the highly unstable cation radical into the highly stable toluene. To overcome such a priori chemical nonsence, E is frequently designated as the standard oxidation potential of toluene for example. However, such a term should not be accepted according to canonical rules because it formally implies that the cell now operates in a driven mode, that is, is connected to an external power supply [19]. Thus in this chapter we prefer to use the denomination standard reduction potentials, rather than the usual temi standard potential, as a reminder of the E° definition, although such as expression is basically a pleonasm. [Pg.9]

Pumps should be cahbrated with a rotameter [27] prior to and after sampling. Analytical instruments must also be calibrated before measurements. For example, GC/MS must be cahbrated for mass and retention times using reference standard materials [70] and comparison made with the fragmentation patterns of known standards, usually a deuturated compound like toluene-dg. Similarly, the method detection limit must be determined by finding the standard deviation of seven replicate analyses and multiplying it by the f-test value for 99% confidence of seven values [30,62]. It is also usual for internal standards to be added to the samples and to evaluate the correlation coefficients of each standard used when multilevel calibration is employed. For automatic thermal desorption tubes, external and internal standardisations are achieved by injecting solutions of standards into the tubes [27,28] for canisters, solutions of standards are injected into the canisters followed by zero air. [Pg.14]

The copolymerizations were carried out under argon using a 1 L Biichi A6 Type I autoclave equipped with an additional external cooling system. For the standard experiments, the reactor was evacuated at 95 C for 1 h and subsequently charged with a solution of norbornene in toluene, 190 mL toluene solvent, 500 mg MAO in 10 mL toluene (from Witco/Crompton), and ethylene at different pressures. Norbornene was dried over triisobutylaluminum and subsequently distilled before use. Polymerizations were initiated by injection of a toluenic metallocene solution into the reaction vessel... [Pg.427]


See other pages where Toluene, external standard is mentioned: [Pg.157]    [Pg.198]    [Pg.127]    [Pg.446]    [Pg.274]    [Pg.185]    [Pg.195]    [Pg.125]    [Pg.246]    [Pg.555]    [Pg.225]    [Pg.1009]    [Pg.1286]    [Pg.151]    [Pg.466]    [Pg.46]    [Pg.263]    [Pg.427]    [Pg.263]    [Pg.46]    [Pg.46]    [Pg.154]    [Pg.90]    [Pg.24]    [Pg.432]   


SEARCH



External standardization

External standards

© 2024 chempedia.info