Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Thermogravimetric analysis instrumentation

Figure 7-18. Thermogravimetric analysis instrument. (Courtesy of Mattler-Toledo, Inc.)... Figure 7-18. Thermogravimetric analysis instrument. (Courtesy of Mattler-Toledo, Inc.)...
Mixtures can be identified with the help of computer software that subtracts the spectra of pure compounds from that of the sample. For complex mixtures, fractionation may be needed as part of the analysis. Commercial instmments are available that combine ftir, as a detector, with a separation technique such as gas chromatography (gc), high performance Hquid chromatography (hplc), or supercritical fluid chromatography (96,97). Instmments such as gc/ftir are often termed hyphenated instmments (98). Pyrolyzer (99) and thermogravimetric analysis (tga) instmmentation can also be combined with ftir for monitoring pyrolysis and oxidation processes (100) (see Analytical methods, hyphenated instruments). [Pg.315]

Nitrogen adsorption was performed at -196 °C in a Micromeritics ASAP 2010 volumetric instrument. The samples were outgassed at 80 °C prior to the adsorption measurement until a 3.10 3 Torr static vacuum was reached. The surface area was calculated by the Brunauer-Emmett-Teller (BET) method. Micropore volume and external surface area were evaluated by the alpha-S method using a standard isotherm measured on Aerosil 200 fumed silica [8]. Powder X-ray diffraction (XRD) patterns of samples dried at 80 °C were collected at room temperature on a Broker AXS D-8 diffractometer with Cu Ka radiation. Thermogravimetric analysis was carried out in air flow with heating rate 10 °C min"1 up to 900 °C in a Netzsch TG 209 C thermal balance. SEM micrographs were recorded on a Hitachi S4500 microscope. [Pg.390]

TA instruments has developed automated thermogravimetric analysis and related kinetic programs that enable a rapid determination of decomposition rates to be made. The following excerpt from a TA application brief [57] explains the method ... [Pg.106]

The essential requirements for an instrument (Figure 11.1) meant for thermogravimetric analysis are, namely ... [Pg.194]

Discuss, the fundamental theory of thermogravimetric analysis , and its instrumentation aspects in an elaborated manner. [Pg.203]

Major instrumentation involved with the generation of thermal property behavior of materials includes thermogravimetric analysis (TG, TGA), DSC, differential thermal analysis (DTA), torsional braid analysis (TBA), thermomechanical analysis (TMA), thermogravimetric-mass spectrometry (TG-MS) analysis, and pyrolysis gas chromatography (PGQ. Most of these analysis techniques measure the polymer response as a function of time, atmosphere, and temperature. [Pg.437]

The section Analysis starts with elemental composition of the compound. Thus the composition of any compound can be determined from its elemental analysis, particularly the metal content. For practically all metal salts, atomic absorption and emission spectrophotometric methods are favored in this text for measuring metal content. Also, some other instrumental techniques such as x-ray fluorescence, x-ray diffraction, and neutron activation analyses are suggested. Many refractory substances and also a number of salts can be characterized nondestructively by x-ray methods. Anions can be measured in aqueous solutions by ion chromatography, ion-selective electrodes, titration, and colorimetric reactions. Water of crystallization can be measured by simple gravimetry or thermogravimetric analysis. [Pg.1092]

The fresh and spent catalysts were characterized with the physisorption/chemisorption instrument Sorptometer 1900 (Carlo Erba instruments) in order to detect loss of surface area and pore volume. The specific surface area was calculated based on Dubinin-Radushkevich equation. Furthermore thermogravimetric analysis (TGA) of the fresh and used catalysts were performed with a Mettler Toledo TGA/SDTA 851e instrument in synthetic air. The mean particle size and the metal dispersion was measured with a Malvern 2600 particle size analyzer and Autochem 2910 apparatus (by a CO chemisorption technique), respectively. [Pg.417]

Thermal Analysis. The water contents of CPA and zeolite were measured by thermogravimetric analysis (TGA). Differential thermal analyses (DTA) were done for comparison. The instrument used was a Du Pont 900 with a TGA attachment. [Pg.303]

Scanning electron micrographs were taken with an ISIDS-130 microscope. Thermogravimetric analysis (TGA) was performed with an SDT Q600 (TA Instruments). A Seiko DMS 210 dynamic-mechanical analyzer was used for DMA measurements. For water content determination samples were allowed to equilibrate over saturated salt solutions in sealed jars at controlled relative... [Pg.232]

Thermogravimetric analysis (TGA) of the co-crystals was carried out over the temperature range 25 °C to 200 °C by employing a Mettler Toledo instrument. These curves showed mass loss due to the removal of the aromatic guest molecules. The apo-hosts thus obtained did not dissolve in benzene and other aromatic solvents. The apo-hosts were immersed in the respective aromatic liquid for several hours. The crystals were then taken out, and the TGA repeated. This procedure was repeated more than once to find out whether the inclusion of the guest molecule was reversible and also whether there was any change in the temperature of decomposition or the proportion of the aromatic compound in the co-crystal, with such cycling. [Pg.426]

Thermogravimetric analysis (TGA) was conducted using a TA Instrument (SDT Q600). Samples of weight around 6 and 8 mg were ramped at 10°C/mn from room temperature to 800°C under nitrogen atmosphere with a flow rate of 100 mL/min and a purge time of 20 min. [Pg.367]

ThermoGravimetric Analysis (TGA) was performed on TGAQ50 (TA Instruments) under air up to 800°C. FTIR spectra were obtained on a Spectrum One from Perkin Elmer Instruments. Elemental Analysis was performed by Service Central d Analyse from CNRS located at Vemaison (France). Con-ductimetry back-titration of amine functions (after contact with HCL 0.1 N in excess) was performed using a Tacussel conductimetry probe by a 0.1M NaOH solution. Porosimetry measurements were performed on a SORPTOMATIC 1990 from CE Instruments specific surface areas were calculated using BET model between P/P =0 and P/P =0.4 and pore size repartitions were determined using BJH model. SEM photos were obtained on an Ultra 55 (Zeiss). [Pg.214]

Thermogravimetric analysis (TGA) was performed with a TA Instrument 2050 thermal analyzer in order to determine the oxygen deficiency of the samples. The samples were ramped at 10°/min to 900 °C and remained isothermal in the presence of O2 gas until oxidation was complete. The oxidized powders were analyzed by PXD and the lattice parameters were compared to published values of StjCrNbO [9] to confirm complete oxidation. [Pg.14]

The compositions of the products were determined by inductively coupled plasma (ICP) with a Perkin-Elmer plasma 40 emission spectrometer. Simultaneous differential thermal analysis and thermogravimetric (DTA-TG) curves were carried out by using Perkin-Elmer DTA-7000, TGA-7 PC series thermal analysis instrument in air with a heating rate of 10 °C /min. The infrared (IR) spectra were recorded on an Impact 410 IR spectrometer on samples pelletized with KBr powder. Valence states were determined by X-ray photoelectron spectroscopy (XPS). The XPS for powder samples fixed on double sided tapes was measured on an ESCA-LAB MKII X-ray photoelectron spectrometer. The Cis signal was used to correct the charge effects. [Pg.40]

Y. A. Guskov A. A. Medvedev and E. V. Ratnikov, Thermogravimetric analysis during microwave heating. Instruments and Experimental Techniques, 40, 292-294 (1997). [Pg.590]

The thermal stability of the copolymers 6-16 was studied using a TA 2950 TGA instrument in a nitrogen atmosphere. Figure 4 shows the thermogravimetric analysis of copolymer 9. The copolymer exhibits thermal stability to a temperature of over 400 °C before the major weight loss is observed. A smaller weight loss of 2% has been detected at about 345 °C. The amount of organic residue at 800 °C is about 4.4%. [Pg.672]

The use of thermogravimetric analysis (TGA) apparatus to obtain kinetic data involves a series of trade-offs. Since we chose to employ a unit which is significantly larger than commercially available instruments (in order to obtain accurate chromatographic data), it was difficult to achieve time invariant O2 concentrations for runs with relatively rapid combustion rates. The reactor closely approximated ideal back-mixing conditions and consequently a dynamic mathematical model was used to describe the time-varying O2 concentration, temperature excursions on the shale surface and the simultaneous reaction rate. Kinetic information was extracted from the model by matching the computational predictions to the measured experimental data. [Pg.547]


See other pages where Thermogravimetric analysis instrumentation is mentioned: [Pg.49]    [Pg.49]    [Pg.201]    [Pg.7]    [Pg.85]    [Pg.72]    [Pg.193]    [Pg.268]    [Pg.376]    [Pg.879]    [Pg.502]    [Pg.2]    [Pg.643]    [Pg.13]    [Pg.18]    [Pg.111]    [Pg.571]    [Pg.605]    [Pg.205]    [Pg.204]    [Pg.276]    [Pg.446]    [Pg.688]    [Pg.3405]    [Pg.3726]    [Pg.28]    [Pg.154]    [Pg.131]    [Pg.349]    [Pg.705]    [Pg.2965]   


SEARCH



Instrumental analysis

Thermogravimetric analysis

© 2024 chempedia.info