Temperature measurement resistance thermometers

Check whether the instrument supplied is equipped for automatic temperature compensation, and, if so, that the temperature probe (resistance thermometer) is available. If it is not so equipped, then the temperature of the solutions to be used must be measured, and the appropriate setting made on the manual temperature control of the instrument. [Pg.570]

Resistance Thermometer A temperature-measuring device consisting of an encapsulated, fine coil of platinum wire whose resistance increases substantially and nearly linearly with rising temperature. See Thermistor. The change in resistance is sensed and converted to a temperature reading. Resistance thermometers have found considerable use in plastics processing. [Pg.625]

Measurement of the hotness or coldness of a body or fluid is commonplace in the process industries. Temperature-measuring devices utilize systems with properties that vaiy with temperature in a simple, reproducible manner and thus can be cahbrated against known references (sometimes called secondaiy thermometers). The three dominant measurement devices used in automatic control are thermocouples, resistance thermometers, and pyrometers and are applicable over different temperature regimes. [Pg.759]

As normally used in the process industries, the sensitivity and percentage of span accuracy of these thermometers are generally the equal of those of other temperature-measuring instruments. Sensitivity and absolute accuracy are not the equal of those of short-span electrical instruments used in connection with resistance-thermometer bulbs. Also, the maximum temperature is somewhat limited. [Pg.760]

Temperature The level of the temperature measurement (4 K, 20 K, 77 K, or higher) is the first issue to be considered. The second issue is the range needed (e.g., a few degrees around 90 K or 1 to 400 K). If the temperature level is that of air separation or liquefact-ing of natural gas (LNG), then the favorite choice is the platinum resistance thermometer (PRT). Platinum, as with all pure metals, has an electrical resistance that goes to zero as the absolute temperature decreases to zero. Accordingly, the lower useful limit of platinum is about 20 K, or liquid hydrogen temperatures. Below 20 K, semiconductor thermometers (germanium-, carbon-, or silicon-based) are preferred. Semiconductors have just the opposite resistance-temperature dependence of metals—their resistance increases as the temperature is lowered, as fewer valence electrons can be promoted into the conduction band at lower temperatures. Thus, semiconductors are usually chosen for temperatures from about 1 to 20 K. [Pg.1136]

The temperature gradient in the direction of flow can be measured directly with Pt-resistance thermometers, but it is difficult and expensive. When this is small, it is better to calculate from the material balance and thermochemical properties. [Pg.75]

Temperature gradient normal to flow. In exothermic reactions, the heat generation rate is q=(-AHr)r. This must be removed to maintain steady-state. For endothermic reactions this much heat must be added. Here the equations deal with exothermic reactions as examples. A criterion can be derived for the temperature difference needed for heat transfer from the catalyst particles to the reacting, flowing fluid. For this, inside heat balance can be measured (Berty 1974) directly, with Pt resistance thermometers. Since this is expensive and complicated, here again the heat generation rate is calculated from the rate of reaction that is derived from the outside material balance, and multiplied by the heat of reaction. [Pg.77]

Electrical resistance thermometers, the most widely used of which is Callendar s platinum resistance thermometer. This is probably the most convenient and accurate apparatus for measuring temperatures between the boiling-point of liquid air (—190° C.) and the melting-point of platinum (1,500° C.). Lead has recently been applied at very low temperatures. [Pg.3]

These fixed points are used to calibrate a different kind of thermometer that is easier to use than a gas thermometer. Over the temperature range from 13.8033 to 1234.93 °A (or K), which is the temperature interval most commonly encountered, the thermometer used for ITS-90 is a platinum resistance thermometer. In this thermometer, the resistance of a specially wound coil of platinum wire is measured and related to temperature. More specifically, temperatures are expressed in terms of W(T9o), the ratio of the resistance R(Ttriple point of water R (273.16 K), as given in equation (1.11)... [Pg.13]

Reliable micro-scale measurement and control of the temperature are required in developing thermal micro-devices. Available measurement techniques can be largely classified into contact and non-contact groups. While the resistance thermometer, thermocouples, thermodiodes, and thermotransistors measure temperature at specific points in contact with them, infrared thermography, thermochromic liquid crystals (TLC), and temperature-sensitive fluorescent dyes cover the whole temperature field (Yoo 2006). [Pg.27]

An instrument for measuring temperatures, in the rubber industry the term is usually applied to an instrument for determining the surface temperature of mill and calender rolls, moulds, etc. The instrument is usually based on thermocouples or, where higher accuracy is required, platinum resistance thermometers. Infrared (IR) techniques are now used which have the advantage of non contact but require careful calibration for the emissivity of the surface. [Pg.51]

The most interesting liquids for low-temperature thermometry are 3He and 4He, especially for the calibration of resistance thermometers in the range from 0.5 to 4.2 K. Vapour pressure of H2 is also interesting to realize vapour pressure-fixed points included in ITS-90. The measure of He vapour pressure has been carried out with great accuracy [42,43] to establish the ITS-90 (see Section 8.3). There are several experimental precautions to be observed in order to obtain reliable measurements [2],... [Pg.213]

The contact resistance may change when the thermometer is moved from a position to another. Hence the accuracy of resistance temperature measurements below about 25 mK... [Pg.224]

Capacitance measurements are quite simple. A typical drawback is the need of coaxial cables that introduce a thermal load which is not negligible in low-power refrigerators. On the other hand, capacitance bridges null the cable capacitance. Multiplexing is more difficult than for resistance thermometers. In principle, capacitors have low loss due to Joule heating. This is not always true losses can be important, especially at very low temperatures. Dielectric constant thermometers have a high sensitivity capacitance differences of the order of 10-19F can be measured. [Pg.227]

During the reaction of the hot catalyst surface with a flammable gas the temperature of the device increases. The Platinum coil itself serves at the same time as a resistance thermometer. The resistance increase of the coil then is a direct measure for the amount of combusted gas. Usually the amount of heat that develops during combustion is small and amounts to 800 kj/mol for methane, for example [8], Therefore the sensor is connected in a bridge circuit to a second resistor which shows the same setup as the pellistor but is catalytically inactive. The bridge voltage is then controlled by the temperature difference of the two sensors (see Fig. 5.34). [Pg.144]

A plant having a shelf area of = 30 m2 has been loaded with 300 kg of water in trays and frozen on the shelves. Water vapor transport and condenser temperatures have been measured in this case between 0.4 and 0.6 mbar, which is approx, two to three times higher than the normally expected operation pressure of the plant (to get an measurable quantity of ice sublimed in a reasonable test time). The data of the test are shown in Fig. 2.19. Three Pt 100 (resistance thermometers) have been frozen in the ice. One CA each have been... [Pg.146]

SAQ 1.1 A temperature is measured with the same platinum-resistance thermometer used in Worked Example 1.1, and a resistance R = 11.4 x 10 4 2 determined. What is the temperature ... [Pg.13]

Melting point instruments consist of a resistance heater to increase temperature, a sample holder and a temperature-measuring device. In its simplest form, an oil bath can be heated with a Bunsen burner, while a capillary tube with a few milligrams of sample is attached to a thermometer with the sample next to the mercury bulb. [Pg.54]

The temperature coefficient of conductance is approximately 1-2 % per °C in aqueous 2> as well as nonaqueous solutions 27). This is due mainly to thetemper-ature coefficient of change in the solvent viscosity. Therefore temperature variations must be held well within 0.005 °C for precise data. In addition, the absolute temperature of the bath should be known to better than 0.01 °C by measurement with an accurate thermometer such as a calibrated platinum resistance thermometer. The thermostat bath medium should consist of a low dielectric constant material such as light paraffin oil. It has been shown 4) that errors of up to 0.5 % can be caused by use of water as a bath medium, probably because of capacitative leakage of current. [Pg.10]


See other pages where Temperature measurement resistance thermometers is mentioned: [Pg.759]    [Pg.56]    [Pg.33]    [Pg.56]    [Pg.583]    [Pg.931]    [Pg.936]    [Pg.763]    [Pg.106]    [Pg.278]    [Pg.180]    [Pg.384]    [Pg.20]    [Pg.174]    [Pg.397]    [Pg.759]    [Pg.262]    [Pg.77]    [Pg.168]    [Pg.301]    [Pg.243]    [Pg.521]    [Pg.73]    [Pg.402]    [Pg.223]    [Pg.68]    [Pg.305]    [Pg.12]    [Pg.436]    [Pg.235]    [Pg.238]   
See also in sourсe #XX -- [ Pg.2 , Pg.5 ]