Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Filter Spectrometers

Figure 2 Series of processes a drop of sample undergoes in the ICP to produce ions, some of which are transported through the sampler and skimmer, focused into the mass spectrometer, filtered on the basis of their mass-to-charge ratio, and detected. Figure 2 Series of processes a drop of sample undergoes in the ICP to produce ions, some of which are transported through the sampler and skimmer, focused into the mass spectrometer, filtered on the basis of their mass-to-charge ratio, and detected.
The Raman spectrometer filters out the unwanted laser light and delivers the proper Raman wavelengths to the proper detector(s). Once most of the laser light has been removed, the Raman light can be... [Pg.4213]

In essence, a guided-ion beam is a double mass spectrometer. Figure A3.5.9 shows a schematic diagram of a griided-ion beam apparatus [104]. Ions are created and extracted from an ion source. Many types of source have been used and the choice depends upon the application. Combining a flow tube such as that described in this chapter has proven to be versatile and it ensures the ions are thennalized [105]. After extraction, the ions are mass selected. Many types of mass spectrometer can be used a Wien ExB filter is shown. The ions are then injected into an octopole ion trap. The octopole consists of eight parallel rods arranged on a circle. An RF... [Pg.811]

Due to the rather stringent requirements placed on the monochromator, a double or triple monocln-omator is typically employed. Because the vibrational frequencies are only several hundred to several thousand cm and the linewidths are only tens of cm it is necessary to use a monochromator with reasonably high resolution. In addition to linewidth issues, it is necessary to suppress the very intense Rayleigh scattering. If a high resolution spectrum is not needed, however, then it is possible to use narrow-band interference filters to block the excitation line, and a low resolution monocln-omator to collect the spectrum. In fact, this is the approach taken with Fourier transfonn Raman spectrometers. [Pg.1164]

A single magnetic sector can be used as a mass filter for other apparatus. However, much more infonnation of the simple mass spectrum of a species can be obtained using the tandem mass spectrometer. [Pg.1335]

Another approach to mass analysis is based on stable ion trajectories in quadnipole fields. The two most prominent members of this family of mass spectrometers are the quadnipole mass filter and the quadnipole ion trap. Quadnipole mass filters are one of the most connnon mass spectrometers, being extensively used as detectors in analytical instnunents, especially gas clnomatographs. The quadnipole ion trap (which also goes by the name quadnipole ion store, QUISTOR , Paul trap, or just ion trap) is fairly new to the physical chemistry laboratory. Its early development was due to its use as an inexpensive alternative to tandem magnetic sector and quadnipole filter instnunents for analytical analysis. It has, however, staned to be used more in die chemical physics and physical chemistry domains, and so it will be described in some detail in this section. [Pg.1339]

Aside from the smgle mass filter, the most connnon configuration for quadnipole mass spectrometers is the triple-quadnipole instrument. This is the simplest tandem mass spectrometer using quadnipole mass filters. Typically, the... [Pg.1342]

Another instrument used in physical chemistry research that employs quadnipole mass filters is the guided ion beam mass spectrometer [31]. A schematic diagram of an example of this type of instrument is shown in figure B 1.7.13. A... [Pg.1345]

Figure Bl.23.5. Schematic illustration of tlie TOE-SARS spectrometer system. A = ion gun, B = Wien filter, C = Einzel lens, D = pulsing plates, E = pulsing aperture, E = deflector plates, G = sample, PI = electron multiplier detector with energy prefilter grid and I = electrostatic deflector. Figure Bl.23.5. Schematic illustration of tlie TOE-SARS spectrometer system. A = ion gun, B = Wien filter, C = Einzel lens, D = pulsing plates, E = pulsing aperture, E = deflector plates, G = sample, PI = electron multiplier detector with energy prefilter grid and I = electrostatic deflector.
Quadrupole mass spectrometers (mass filters) allow ions at each m/z value to pass through sequentially for example, ions at m/z 100, 101, 102 will pass one after the other through the quadrupole assembly so that first m/z 100 is transmitted, then m/z 101, then m/z 102 (or vice versa), and so on. Therefore, the ion collector (or detector) at the end of the quadrupole unit needs to cover only one point or focus in space (Figure 29.1a), and a complete mass spectrum is recorded over a period of time. The ions arrive at the collector sequentially, and ions are detected in a time domain, not in a spatial domain. [Pg.205]

In a sector instrument, which acts as a combined mass/velocity filter, this difference in forward velocity is used to effect a separation of normal and metastable mj" ions (see Chapter 24, Ion Optics of Magnetic/Electric-Sector Mass Spectrometers ). However, as discussed above, the velocity difference is of no consequence to the quadmpole instrument, which acts only as a mass filter, so the normal and metastable mj ions formed in the first field-free region (Figure 33.1) are not differentiated. [Pg.233]

An FT-Raman spectrometer is often simply an FTIR spectrometer adapted to accommodate the laser source, filters to remove the laser radiation and a variety of infrared detectors. [Pg.124]

Acousto-optic Filters. The newest type of spectrometer to become commercially available is the acousto-optic tunable filter (AOTF). An AOTF is a sohd-state, electronically tunable bandpass filter based on the diffraction of optical waves by acoustic waves in an optically anisotropic crystal. [Pg.196]

A widely used procedure for determining trace amounts of selenium involves separating selenium from solution by reduction to elemental selenium using tellurium (as a carrier) and hypophosphorous acid as reductant. The precipitated selenium, together with the carrier, are collected by filtration and the filtered soflds examined directly in the wavelength-dispersive x-ray fluorescence spectrometer (70). Numerous spectrophotometric and other methods have been pubHshed for the deterruination of trace amounts of selenium (71—88). [Pg.335]

The characteristic peaks must be separated from the background. In EDS this is usually accomplished by mathematical filtering of the spectrum or by background modeling. In WDS, the peak is sufficiently sharp to permit background measurement by detuning the spectrometer. [Pg.185]

Each type of mass spectrometer has its associated advantages and disadvantages. Quadrupole-based systems offer a fairly simple ion optics design that provides a certain degree of flexibility with respect to instrument configuration. For example, quadrupole mass filters are often found in hybrid systems, that is, coupled with another surface analytical method, such as electron spectroscopy for chemical analysis or scanning Auger spectroscopy. [Pg.552]


See other pages where Filter Spectrometers is mentioned: [Pg.85]    [Pg.144]    [Pg.6516]    [Pg.6515]    [Pg.20]    [Pg.85]    [Pg.144]    [Pg.6516]    [Pg.6515]    [Pg.20]    [Pg.210]    [Pg.812]    [Pg.1234]    [Pg.1248]    [Pg.1631]    [Pg.278]    [Pg.290]    [Pg.290]    [Pg.292]    [Pg.293]    [Pg.418]    [Pg.195]    [Pg.196]    [Pg.198]    [Pg.199]    [Pg.201]    [Pg.202]    [Pg.545]    [Pg.549]    [Pg.315]    [Pg.58]    [Pg.90]    [Pg.134]    [Pg.545]    [Pg.140]    [Pg.551]    [Pg.552]   
See also in sourсe #XX -- [ Pg.56 ]




SEARCH



© 2024 chempedia.info