Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Spectrometers double-beam recording

Infrared Spectrometer, double-beam or single-beam, suitable for recording accurate measurements between 690 cm (14.49 pm) and 440 cm" (22.73 pm). Refer to Practice E275. [Pg.625]

In double-beam spectrometers, the correction is made before the spectrum is recorded by dividing ... [Pg.54]

Spectra were recorded on a Grubb-Parsons double beam infrared spectrometer, type S4, equipped with sodium chloride prism. The positions of sharp bands were reproducible to 15 cm-1 in the 3000 cm"1 region, and to d 2 cm-1 in the 1700 cm-1 region. Calibration was against water vapour, and the sharp NH band of A-methylaniline at 3430 cm-1 [12, 13]. [Pg.481]

To obtain a sample spectrum equivalent to the one obtained on a double beam spectrometer, two spectra of transmitted intensities are recorded the first without sample (absorption background), and the second with sample. The conventional %Tspectrum can be obtained from the two preceding measurements (Fig. 10.12). [Pg.171]

Figure 10.12—Sequence of events necessary to obtain a pseudo-double beam spectrum with a Fourier transform IR spectrometer. The instrument records and stores in its memory two spectra representing the variation of lu (blank) and / (sample) as a function of wavenumber (emission spectra 1 and 2 above). Then, it calculates the conventional spectrum, which is identical to that obtained on a double beam instrument, by calculating the ratio T — /// — f(A) for each wavenumber. Atmospheric absorption (CO2 and H20) is thus eliminated. The figure illustrates the spectrum of a polystyrene film. Figure 10.12—Sequence of events necessary to obtain a pseudo-double beam spectrum with a Fourier transform IR spectrometer. The instrument records and stores in its memory two spectra representing the variation of lu (blank) and / (sample) as a function of wavenumber (emission spectra 1 and 2 above). Then, it calculates the conventional spectrum, which is identical to that obtained on a double beam instrument, by calculating the ratio T — /// — f(A) for each wavenumber. Atmospheric absorption (CO2 and H20) is thus eliminated. The figure illustrates the spectrum of a polystyrene film.
Figure 11.12—Schematic ant optica path of a single beam spectrophotometer equipped with electronic regulation (Hitachi U-1000). Measurements in solution are often carried out at a fixed wavelength after a calibration curve has been plotted. The use of higher performance double beam UV/Visible spectrometers is not necessary for these measurements in which the spectrum is not recorded. On the other hand, quantitative measurements from mixtures represent a different type of analysis. Figure 11.12—Schematic ant optica path of a single beam spectrophotometer equipped with electronic regulation (Hitachi U-1000). Measurements in solution are often carried out at a fixed wavelength after a calibration curve has been plotted. The use of higher performance double beam UV/Visible spectrometers is not necessary for these measurements in which the spectrum is not recorded. On the other hand, quantitative measurements from mixtures represent a different type of analysis.
N, adsorption for determination of pore size distribution(BJH method) using a Micrometries ASAP 2000 automatic analyzer. FTIR spectra were recorded on a PerkinElmer 221 spectrometer and UV-vis spectroscopic measurments were carried out using Varian CARY 3E double beam spectrometer in the range of 190 - 820 nm. [Pg.783]

All UV/vis spectra were recorded on a Varian Cary 50 Scan spectrophotometer and a PERKIN ELMER UV/vis Spectrometer Lambda 2 (double beam) in solution. Absorption maxima Xmax are given in nm. 0.4 cm quartz cuvettes were used for all measurements. [Pg.71]

Figure 4.6-12 Schematic representation of the experiment for recording the optical rotation of the sample S P vector of polarizer, i.c. direction of the electric vector of transmitted radiation, A vector of analyzer the sample rotates the electric vector of the radiation through an angle g leading (as indicated by the broken lines) to different intensities in the two experimental configurations to be realized readily with a double-beam spectrometer or sequentially with a single-beam in.strument. Figure 4.6-12 Schematic representation of the experiment for recording the optical rotation of the sample S P vector of polarizer, i.c. direction of the electric vector of transmitted radiation, A vector of analyzer the sample rotates the electric vector of the radiation through an angle g leading (as indicated by the broken lines) to different intensities in the two experimental configurations to be realized readily with a double-beam spectrometer or sequentially with a single-beam in.strument.
Double-beam or dispersive instruments in which the IR radiation from a single source is split into two identical beams. One beam passes through the sample and the other is used as a reference and passes through air or the pure solvent used to dissolve the sample. The difference in intensity of the two beams is detected and recorded as a peak the principal components of this type of instrument are shown in Fig. 28.2. The important controls on the spectrometer are ... [Pg.182]

Zhurkov, Novak and Vettegran have recently reported studies where IR was used to determine the formation of CH and C C bonds during deformation of polyethylene and polypropylene at room temperature and atmospheric conditions (11). In these studies, specimens of equal thickness, one unstrained, the other fractured, were Interposed In the balanced light beams of double-beam spectrometer. In this mode the spectrometer reportedly records the difference in absorption AD, of the undeformed and fractured specimen. Strong absorption bonds were noted at 910, 965, 1379 and 1735, cm. These were attributed respectively to (RCft CH ), (RCH=CHR ), (R-CH ), and (RCHO) groups. [Pg.205]

Bearing in mind a resolution factor imposed by the method of calculation, the classical representation of the spectrum is obtained, 7 = f (A) or 7 = f (i ). According to Nyquist s theory, at least two points per period are required in order to find, by calculation, a given wavelength of the spectrum. To obtain a sample spectrum equivalent to the one obtained with a double beam spectrometer, two spectra of transmitted intensities are recorded the first, without sample (absorption background) and the second with sample. The conventional spectrum, in percentage r, is obtained from these two spectra (Figure 10.9). [Pg.218]

Figure 10.9 Sequence for obtaining a pseudo-double beam spectrum with a Fourier transform infrared spectrometer. The apparatus records and memorizes two spectra, which represent the variations of 7q (the blank) and I (sample) as a function of the wavenumber (these are emission spectra 1 and 2). Next the conventional spectrum is calculated, identical to that of an instrument of the double beam type, by calculating the ratio T = 1/4 = fW for each wavenumber. Strong atmospheric absorption (COj and HjO), which is present along the optical path is eliminated in this way. The illustrations correspond to the spectrum of a polystyrene film. Figure 10.9 Sequence for obtaining a pseudo-double beam spectrum with a Fourier transform infrared spectrometer. The apparatus records and memorizes two spectra, which represent the variations of 7q (the blank) and I (sample) as a function of the wavenumber (these are emission spectra 1 and 2). Next the conventional spectrum is calculated, identical to that of an instrument of the double beam type, by calculating the ratio T = 1/4 = fW for each wavenumber. Strong atmospheric absorption (COj and HjO), which is present along the optical path is eliminated in this way. The illustrations correspond to the spectrum of a polystyrene film.
Figure 11.9 Schematic of a Shimadzu F-4500 spectrofluorometer. A fraction of the incident beam, reflected by a semi-transparent mbror, reaches a reference PMT. A comparison of the signals from the two PMTs leads to the elimination of any drifting of the source. This procedure, for single beam instruments, gives approximately the same stability as with a double beam spectrometer. However, the spectrum of a given solution will often present minor differences when recorded upon different instruments (reproduced courtesy of Shimadzu). Figure 11.9 Schematic of a Shimadzu F-4500 spectrofluorometer. A fraction of the incident beam, reflected by a semi-transparent mbror, reaches a reference PMT. A comparison of the signals from the two PMTs leads to the elimination of any drifting of the source. This procedure, for single beam instruments, gives approximately the same stability as with a double beam spectrometer. However, the spectrum of a given solution will often present minor differences when recorded upon different instruments (reproduced courtesy of Shimadzu).
Spectrometers that use phototubes or photomultiplier tubes (or diode arrays) as detectors are generally called spectrophotometers, and the corresponding measurement is called spectrophotometry. More strictly speaking, the journal Analytical Chemistry defines a spectrophotometer as a spectrometer that measures the ratio of the radiant power of two beams, that is, PIPq, and so it can record absorbance. The two beams may be measured simultaneously or separately, as in a double-beam or a single-beam instrument—see below. Phototube and photomultiplier instruments in practice are almost always used in this maimer. An exception is when the radiation source is replaced by a radiating sample whose spectrum and intensity are to be measured, as in fluorescence spectrometry—see below. If the prism or grating monochromator in a spectrophotometer is replaced by an optical filter that passes a narrow band of wavelengths, the instrument may be called a photometer. [Pg.492]

Double-beam spectrometers can automatically scan the wavelength and record the spectrum. [Pg.497]

Synchroton radiation has been employed as a spectral source for a study of the absorption of HCN and DCN in the wavelength range 80—120nm. A vacuum-u.v. spectrophotometer for absorptions in the region 105—200 nm has been described. Solid-, liquid-, and gas-phase samples could be analysed at temperatures from —200 to 100 °C and at pressures between 0 and 150 atmospheres. The absorption spectrum of tra j-di-imide in the vacuum-u.v. has been measured. First-derivative u.v. spectroscopy has been employed in the analysis of Watts nickel plating solutions for trace amounts of saccharin. Impurity levels of 0.1 p.p.m. have been recorded. A wavelength modulated derivative spectrophotometer with a multi-pass absorption cell has been developed for the automatic analysis of atmospheric pollutants. Traces of SOj, NO, and NO2 were detected with limits of 15, 13, and Sp.p.b., respectively. A double-beam single-detector absorption spectrometer has been constructed. Independence... [Pg.16]

Infrared spectra were recorded with a double beam Perkin Elmer S80B spectrometer. A self-supporting wafer of compressed zeolite powder was suspended in the infrared beam in an home-made vacuum cell, permitting in-situ heating of the sample and adsorption of vapors. [Pg.528]

Record the absorption spectrum of a solution of pure octane. Record the absorption spectrum of a 0.02% v/v toluene in octane solution from 220 to 280 nm. For a double-beam spectrometer, pure octane can be put into the reference cell. Compare with Experiment 5.1. Explain your observations. Change the slit width as in Experiment 5.1 and observe what happens to the resolution. [Pg.378]

Fig. IS. Schematic representation of the optical layout of a double-beam ratio-recording spectrometer. Reprinted from [21], Copyright John Wiley and Sons Ltd. Reproduced with permission... Fig. IS. Schematic representation of the optical layout of a double-beam ratio-recording spectrometer. Reprinted from [21], Copyright John Wiley and Sons Ltd. Reproduced with permission...

See other pages where Spectrometers double-beam recording is mentioned: [Pg.765]    [Pg.1122]    [Pg.68]    [Pg.475]    [Pg.141]    [Pg.27]    [Pg.404]    [Pg.84]    [Pg.384]    [Pg.181]    [Pg.336]    [Pg.156]    [Pg.156]    [Pg.77]    [Pg.146]    [Pg.68]    [Pg.170]    [Pg.3410]    [Pg.45]    [Pg.771]    [Pg.165]    [Pg.99]    [Pg.1122]    [Pg.297]    [Pg.164]    [Pg.250]    [Pg.207]    [Pg.750]    [Pg.230]    [Pg.365]   
See also in sourсe #XX -- [ Pg.357 ]

See also in sourсe #XX -- [ Pg.357 ]




SEARCH



Double beam

Double-beam spectrometer

Spectrometers recording

© 2024 chempedia.info