Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

SOFT SYSTEMS ANALYTIC

Figure 3.2 Soft-systems analytical framework (adapted from [293])... Figure 3.2 Soft-systems analytical framework (adapted from [293])...
Parallels between the soft-systems analytical framework and a constrained relativism are evident when comparing Figure 3.2 with Figure 1.1(b) in Chapter 1 - reproduced as Figure 3.3 below. [Pg.89]

The soft-systems analytical framework facilitated the comparing and contrasting of national approaches, as well as interviewee perspectives of EU decision-making and REACH. In total, the results of these processes formed the basis for proposing a framework for decision-making under REACH. [Pg.90]

Tandem mass spectrometry or ms/ms was first introduced in the 1970s and gained rapid acceptance in the analytical community. The technique has been used for stmcture elucidation of unknowns (26) and has the abiUty to provide sensitive and selective analysis of complex mixtures with minimal sample clean-up (27). Developments in the mid-1980s advancing the popularity of ms/ms included the availabiUty of powerhil data systems capable of controlling the ms/ms experiment and the viabiUty of soft ionisation techniques which essentially yield only molecular ion species. [Pg.405]

On the other hand, when latent variables instead of the original variables are used in inverse calibration then powerful methods of multivariate calibration arise which are frequently used in multispecies analysis and single species analysis in multispecies systems. These so-called soft modeling methods are based, like the P-matrix, on the inverse calibration model by which the analytical values are regressed on the spectral data ... [Pg.186]

Quite often a normal electron ionization mass spectrum appears insufficient for reliable analyte identification. In this case additional mass spectral possibilities may be engaged. For example, the absence of the molecular ion peak in the electron ionization spectrum may require recording another type of mass spectrum of this analyte by means of soft ionization (chemical ionization, field ionization). The problem of impurities interfering with the spectra recorded via a direct inlet system may be resolved using GC/MS techniques. The value of high resolution mass spectrometry is obvious as the information on the elemental composition of the molecular and fragment ions is of primary importance. [Pg.173]

LC-MS uses different types of soft chemical ionization that produces molecular ions and no fragmentation pattern. In MS/MS instruments the molecular ions can be fragmented by collision with a gas for example, He. This fragmentation can be used for identification of a compound. No mass spectral libraries exist for LC-MS hence identification of unknown compounds is more time-consuming than for GC-MS. For known compounds LC-MS is a very sensitive and specific method, using LC-MS/MS systems the analytical performance can be increased even more. LC-MS analysis is especially suitable for non-volatile POMs such as non-ionic surfactants in house dust samples (Clausen et al., 2003). [Pg.36]

The relatively rapid nature of the separation, e.g., routinely can take as little as seconds with analytical systems and between minutes to hours for semipreparative or preparative systems with high-performance, pressure-stable, and mechanically stable sorbents with particle diameters in the range of 3-65 /rm rather than days with the classical types of soft gels, where the particle sizes are usually greater than 150 /un. [Pg.107]

Pseudomolecular Ions. In contrast to the traditional MS, the highest mass peaks in ESI/APCI spectra are not always the molecular ion of interest. Instead, pseudomolecular ions, or noncovalent complex ions, are commonly observed. The pseudomolecular ions are generally formed by the analyte-adduct interaction in the solution system that is preserved as a result of the soft ionization of the ESI/APCI process. These ions are also formed by analyte-adduct gas-phase collisions in the spray chamber [49]. The exact mechanisms of how the analyte adducts are formed in ESI/APCI still remain unresolved at this point. More often than not, the adduct ion formation is a major cause for the low detection limit for ESEAPCI MS. However, these associative processes have also created interest in the study of drug-protein/ drug-oligonucleotide gas-phase complexes that benefit from the ability of ESI/APCI MS analysis. [Pg.306]


See other pages where SOFT SYSTEMS ANALYTIC is mentioned: [Pg.88]    [Pg.104]    [Pg.88]    [Pg.104]    [Pg.16]    [Pg.115]    [Pg.377]    [Pg.207]    [Pg.139]    [Pg.165]    [Pg.681]    [Pg.52]    [Pg.95]    [Pg.155]    [Pg.578]    [Pg.637]    [Pg.68]    [Pg.637]    [Pg.668]    [Pg.22]    [Pg.70]    [Pg.355]    [Pg.728]    [Pg.287]    [Pg.224]    [Pg.377]    [Pg.35]    [Pg.74]    [Pg.34]    [Pg.542]    [Pg.557]    [Pg.201]    [Pg.120]    [Pg.670]    [Pg.435]    [Pg.188]    [Pg.88]    [Pg.286]    [Pg.22]    [Pg.70]    [Pg.355]    [Pg.81]   


SEARCH



Analytical systems

SOFT SYSTEMS

SOFT SYSTEMS ANALYTIC FRAMEWORK

© 2024 chempedia.info