Sector field ICP

Kim et al. [116] determined plutonium isotopes in seawater by an online sequential injection technique with sector field ICP-MS. [Pg.355]

M. P. Field, J. T. Cullen, and R. M. Sherrell. Direct Determination of 10 Trace Metals in 50 xL Samples of Coastal Seawater Using Desolvating Micronebulization Sector Field ICP-MS. J. Anal. Atom. Spectrom., 14(1999) 1425-1431. [Pg.241]

Figure 5.2 Overview of ICP-MS instrumentation quadrupole-based ICP-MS with and without colli-sion/reaction cell, sector field ICP-MS with single ion collector (ICP-SFMS) and multiple ion collector ICP-MS (MC-ICP-MS) and time-of-flight ICP-MS. Figure 5.2 Overview of ICP-MS instrumentation quadrupole-based ICP-MS with and without colli-sion/reaction cell, sector field ICP-MS with single ion collector (ICP-SFMS) and multiple ion collector ICP-MS (MC-ICP-MS) and time-of-flight ICP-MS.
As an alternative to commercial quadrupole based ICP-MS measurements at low mass resolution ( / 300), in 1996 Yiang and Douglas15 proposed a quadrupole ICP-MS system which allows a maximum mass resolution of 9000. At a mass resolution of 5000 the sensitivity was comparable to that of a commercial double-focusing sector field ICP-MS operated at the same mass resolution. Due to the very high continuum instrumental background of about lOOOcps no commercial high resolution quadrupole instrumentation with an inductively coupled plasma source exists and the development of high resolution quadrupole based ICP-MS has ceased. [Pg.123]

Double-focusing Sector Field ICP Mass Spectrometers with Single Ion Collector (ICP-SFMS)... [Pg.131]

Figure 5.9 Double-focusing sector field ICP-MS (ICP-SFMS, Element XR, Thermo Fisher Scientific, Bremen) with extended dynamic range up to 1 013 (a), arrangement of ion detector systems is shown at top right, (b). (Reproduced by permission of Thermo Fisher Scientific, Bremen.)22... Figure 5.9 Double-focusing sector field ICP-MS (ICP-SFMS, Element XR, Thermo Fisher Scientific, Bremen) with extended dynamic range up to 1 013 (a), arrangement of ion detector systems is shown at top right, (b). (Reproduced by permission of Thermo Fisher Scientific, Bremen.)22...
The analytical performance of selected commercially available sector field ICP-MS with multiple ion collection compared to double-focusing sector field ICP-MS with a single ion collector are compared in Table 5.2. [Pg.138]

Sector field ICP-MS with single and multiple ion collectors ... [Pg.139]

The experimental arrangement of LA-ICP-MS using a powerful laser ablation system coupled to a double-focusing sector field ICP-MS (Element, Thermo Fisher Scientific, Bremen) is shown in Figure 5.22. A cooled laser ablation chamber using two Peltier elements in serial connection... [Pg.151]

Figure 5.22 Schematic of laser ablation inductively coupled plasma mass spectrometer a laser ablation system (wa velength 213 nm) is coupled to a double-focusing sector field ICP-MS (Element, Thermo Fisher Scientific, Bremen). The cooled laser ablation chamber using two Peltier elements was developed in the author s laboratoryJ26 (M. V. Zoriy, M. Kayser, A. Izmer, C. Pickhardt and J. S. Becker, Int. J. Mass Spectrom., 242, 297 (2005). Reproduced by permission of Elsevier.)... Figure 5.22 Schematic of laser ablation inductively coupled plasma mass spectrometer a laser ablation system (wa velength 213 nm) is coupled to a double-focusing sector field ICP-MS (Element, Thermo Fisher Scientific, Bremen). The cooled laser ablation chamber using two Peltier elements was developed in the author s laboratoryJ26 (M. V. Zoriy, M. Kayser, A. Izmer, C. Pickhardt and J. S. Becker, Int. J. Mass Spectrom., 242, 297 (2005). Reproduced by permission of Elsevier.)...
Figure 6.2 Mass spectrum of hafnium isotopes measured by double-focusing sector field ICP-MS. Figure 6.2 Mass spectrum of hafnium isotopes measured by double-focusing sector field ICP-MS.
Figure 6.5a illustrates part of a real mass spectrum of a radioactive waste solution in the mass range 180-250 u measured by double-focusing sector field ICP-MS (ELEMENT, Thermo Fisher Scientific) at low mass resolution. In this mass spectrum, atomic ions of Re at mass 185 u and 187 u with isotope abundances of 37.4 % and 62.6 %, respectively (Re was used for "Tc precipitation),... [Pg.182]


See other pages where Sector field ICP is mentioned: [Pg.56]    [Pg.326]    [Pg.28]    [Pg.36]    [Pg.100]    [Pg.127]    [Pg.131]    [Pg.132]    [Pg.132]    [Pg.132]    [Pg.135]    [Pg.137]    [Pg.138]    [Pg.138]    [Pg.144]    [Pg.151]    [Pg.152]    [Pg.182]    [Pg.217]    [Pg.217]    [Pg.228]    [Pg.231]    [Pg.231]    [Pg.241]    [Pg.242]    [Pg.243]    [Pg.244]    [Pg.247]    [Pg.264]    [Pg.270]    [Pg.272]    [Pg.285]    [Pg.309]    [Pg.309]    [Pg.311]    [Pg.312]   


SEARCH



Double-focusing sector field ICP mass spectrometry