Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Resonance radiation detectors

Ion Detectors in most mass spectrometers, ions are detected after collisions with a detector surface. The collisions cause electrons, photons, or other ions to be emitted. These can be detected by charge or radiation detectors. For example, a common detector is the electron multiplier, which was described in Section 28F-3. In ion-cyclotron resonance, the orbiting ions induce a signal whose frequencies are inversely related to the mIz values. The frequencies are decoded by Fourier transform techniques. [Pg.956]

Example of application Fluorescence spectroscopy, synchrotron radiation Rutherford back-scattering on surfaces Positron annihilation spectroscopy Muon spin resonance spectroscopy Radiation detectors, radiochemistiy Mossbauer effect Nuclide production, activation analysis... [Pg.366]

When the applied magnetic field is swept to bring the sample into resonance, MW power is absorbed by the sample. This changes the matching of the cavity to the waveguide and some power is now reflected and passes via the circulator to the detector. This reflected radiation is thus the EPR signal. [Pg.1561]

Scattered radiation. In a transmission experiment, the Mossbauer sample emits a substantial amount of scattered radiation, originating from XRF and Compton scattering, but also y-radiation emitted by the Mossbauer nuclei upon de-excitation of the excited state after resonant absorption. Since scattering occurs in 4ti solid angle, the y-detector should not be positioned too close to the absorber so as not to collect too much of this unwanted scattered radiation. The corresponding pulses may not only uimecessarily overload the detector and increase the counting dead time, but they may also affect the y-discrimination in the SCA and increase the nonresonant background noise. [Pg.45]

The formal definition of this quality factor, Q, is the amount of power stored in the resonator divided by the amount of power dissipated per cycle (at 9.5 GHz a cycle time is l/(9.5 x 109) 100 picoseconds). The dissipation of power is through the resonator walls as heat, in the sample as heat, and as radiation reflected out of the resonator towards the detector. The cycle time is used in the definition because the unit time of one second would be far too long for practical purposes within one second after the microwave source has been shut off, all stored power has long been dissipated away completely. [Pg.18]

The basic instrumentation used for spectrometric measurements has already been described in Chapter 7 (p. 277). The natures of sources, monochromators, detectors, and sample cells required for molecular absorption techniques are summarized in Table 9.1. The principal difference between instrumentation for atomic emission and molecular absorption spectrometry is in the need for a separate source of radiation for the latter. In the infrared, visible and ultraviolet regions, white sources are used, i.e. the energy or frequency range of the source covers most or all of the relevant portion of the spectrum. In contrast, nuclear magnetic resonance spectrometers employ a narrow waveband radio-frequency transmitter, a tuned detector and no monochromator. [Pg.355]


See other pages where Resonance radiation detectors is mentioned: [Pg.8]    [Pg.90]    [Pg.274]    [Pg.72]    [Pg.365]    [Pg.291]    [Pg.79]    [Pg.325]    [Pg.145]    [Pg.282]    [Pg.356]    [Pg.1215]    [Pg.37]    [Pg.49]    [Pg.423]    [Pg.380]    [Pg.174]    [Pg.791]    [Pg.904]    [Pg.26]    [Pg.39]    [Pg.40]    [Pg.41]    [Pg.58]    [Pg.534]    [Pg.541]    [Pg.164]    [Pg.528]    [Pg.15]    [Pg.60]    [Pg.331]    [Pg.361]    [Pg.82]    [Pg.5]    [Pg.106]    [Pg.2]    [Pg.63]    [Pg.105]    [Pg.618]    [Pg.89]    [Pg.380]    [Pg.536]   
See also in sourсe #XX -- [ Pg.276 , Pg.277 ]




SEARCH



Radiation detectors

Radiation resonance

Resonance detector

Resonant Detectors

© 2024 chempedia.info