Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Recommended Spectral Range

It is shown, that the offered approach allows to optimize a choice of a spectral range for cairying out of measurings and to recommend for a selected spectral range of value of the pai ameters of the equipment. It s permitting to implement measurings most efficiently. [Pg.458]

One of the problems which must be solved for quantitative measurements by emission is the need for a blackbody source at the temperature of measurement. And a variety of blackbody references have been used including a V-shaped cavity of graphite 164), a metal plate covered with a flat black paint1S6 160) and a cone of black paper l53). However, none of these methods of producing a blackbody reference spectrum are adequate. In most cases the efficiency of the reference has not been established. The most recent recommendation 1S0) is an aluminium cup painted with an Epley-Parsons solar black lacquer which has an emittance of greater than 98% over the infrared spectral range. [Pg.115]

Adhering to these rather strict requirements for optimal optical density, spectral range and minimizing spectral artifacts requires permanent verifications of the reality of the CD signal measured. Good CD results require formulation and keeping of the measurement protocol. It is recommended that the vhole series of samples that are to be compared is measured under the same conditions. For example, a broken sample cell can distort the vhole series. [Pg.272]

Since increasing the number of scans can improve spectral quality, in general the user of the instrument should be allowed to adjust this parameter to optimize the spectrum of each sample. Thus, when writing a standard operating procedure it would be best to recommend a range for the number of scans or include a suggested value, but leave the instrument operator free to alter the number of scans as needed. You may have noticed that many FTIR operators routinely use a number of scans that is a multiple of 2, such as 16 or 32 scans. There is no magic in these numbers. In most cases there is no reason why 15 or 33 scans could not be used instead. It s just that these multiples of 2 are small, round numbers that are easy to remember. [Pg.30]

Fig. 4.14 Spectral range of availability of the most common actinometers the dashed lines indicate the wavelength ranges where the use is possible, but not recommended or simple... Fig. 4.14 Spectral range of availability of the most common actinometers the dashed lines indicate the wavelength ranges where the use is possible, but not recommended or simple...
Any object whose temperature Is greater than absolute zero emits radiation. Since radiation emitted by the human body falls within a well-defined spectral range. It Is theoretically possible to detect the Intrusion of humans Into a robotic workplace by using Infrared sensors. Unfortunately, many other objects such as electric light bulbs emit radiation In the same spectral range as the human body. Therefore, Infrared devices are subject to frequent false triggerings. These devices have been examined by a few researchers who do not recommend their use because of the above-mentioned reasons. [Pg.356]

Quantum-Yield and Quantum-Counter Standards. The materials recommended as quantum-yield and quantum-counter standards are summarized In Table IV. As can be seen, several of the materials suggested as spectral responslvlty standards have also been suggested as quantum-yield and quantum-counter standards. The materials listed In Table IV cover the wavelength range from 200 to almost 800 nm, giving an excellent choice for the researcher. Quantum counters that operate further Into the red are still needed, especially for the area of... [Pg.103]

For the confirmatory procedure, it is recommended that the sponsor develop spectral data based on at least three structurally specific ions that completely define the marker residue molecule. These ions may or may not include the molecular ion. The use of water loss and isotopic ions is usually unacceptable and CVM concurrence should be sought when water loss ions or isotopic ions are selected for the confirmatory analysis. The proposed fragment ion structures should be consistent with the fragmentation pattern, and justification for specificity of selected ions or scan range should be included. All confirmation criteria should be specified in the standard operating procedure. [Pg.86]

Figure 7 demonstrates the effects of window functions on the signal to noise ratio of the 5r,r spectrum (for D-HMBC) and 5r,i spectrum. 5r,r spectra shown in fig. 7(c) revealed marked decrease of -noise in the range between 0.8 to 1.1 ppm. Processing of S r r spectral data, however, requires somewhat tedious phase adjustment of the Fi spectrum. In order to achieve this easily, it is recommended to take a ID-NMR spectrum at first under... [Pg.183]

In view of the subsequent developments118, the recommendations for a potential user would be for decoupled spectra when only enhancement is required—use INEPT (it is shortest) for spectral editing—use DEPT (less cross talk) for coupled spectra use INEPT- or, if antiphase multiples can be tolerated, use INEPT. To enhance the uniformity of excitation over larger 7 ranges, experiments measured with different r values can be added. [Pg.248]

The CIE (Commission Internationale de l Eclairage) has selected a number of light sources from the wide range available with different spectral distribution and color temperature. The use of these standard illuminants is recommended and those used most commonly are listed here ... [Pg.7]

Verbally, the absorptivity computed from Eq. (5-141) by using the correlations in Table 5-5 is based on a value for gas emissivity Eg calculated at a temperature 7 and at a partial-pressure path-length product of (pc+pw)L7y77. The absorptivity is then equal to this value of gas emissivity multiplied by (Tg/Tj). It is recommended that spectrally based models such as RADCAL (loc. cit.) be used particularly when extrapolating beyond the temperature, pressure, or partial-pressure-length product ranges presented in Table 5-5. [Pg.32]

Calibration models are developed to determine the relationship between calibration set spectra and the constituent value of interest for those samples. Calibration involves taking spectra from many samples varying over the measurement range and also measuring the desired parameters. A rugged chemometric model for a complex sample may require hundreds to thousands of samples taken from all possible situations, in and out of specification, that it may encounter. Samples selected for calibration must contain all of the variables affecting the chemical and physical properties of the samples to be analyzed. To characterize each source of variation, it is recommended that 15 to 20 samples be run for each variable. Application of a math treatment, such as second derivative, prepares the raw spectral data for use in a regression and subsequent development of a calibration equation. This type of treatment results in a data file that will yield more information more easily than a raw data file. [Pg.3631]


See other pages where Recommended Spectral Range is mentioned: [Pg.86]    [Pg.86]    [Pg.164]    [Pg.82]    [Pg.393]    [Pg.30]    [Pg.298]    [Pg.308]    [Pg.3379]    [Pg.58]    [Pg.278]    [Pg.143]    [Pg.722]    [Pg.190]    [Pg.325]    [Pg.881]    [Pg.190]    [Pg.147]    [Pg.198]    [Pg.86]    [Pg.225]    [Pg.360]    [Pg.102]    [Pg.402]    [Pg.254]    [Pg.282]    [Pg.1013]    [Pg.248]    [Pg.88]    [Pg.404]    [Pg.411]    [Pg.137]    [Pg.1219]    [Pg.6087]    [Pg.148]    [Pg.1502]    [Pg.310]    [Pg.71]   


SEARCH



Spectral range

© 2024 chempedia.info