Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Probe flow-through probes

For practical applications, notably, GLRS can be used in a backscattering (180°) geometry, which can be easily fabricated into a dip-type or flow through probe for processes monitoring (Fig. 9.8.2). [Pg.307]

In practice, direct insertion of samples requires a somewhat more elaborate arrangement than might be supposed. The sample must be placed on an electrode before insertion into the plasma flame. However, this sample support material is not an electrode in the usual meaning of the term since no electrical current flows through it. Heating of the electrode is done by the plasma flame. The electrode or probe should have small thermal mass so it heats rapidly, and it must be stable at the high temperatures reached in the plasma flame. For these reasons, the sort of materials used... [Pg.114]

The calibration air flows through a thin tube. The probe is placed at the exit of the tube. When the tube is long enough and the tube flow is laminar, the reference velocity for calibration can be calculated from the theoretical, fully developed laminar velocity profile. [Pg.1158]

At very high frequencies, the current is measured by assessing one of the effects that it produces. Several techniques are possible, e.g. (1) measuring the temperature rise when the current flows through a known resistance or (2) using a Hall-effect probe to measure the electromagnetic field created by the current. [Pg.236]

Make-up water. A semi-sealed section within the main tank will feed water from the treatment plant to the main tank either through a separate make-up tank or. In either case, the water level will be the same in both sections, excluding the ullage left for condense returns. Control of the make-up water level may be by float valve, float switches or conductivity probes. These methods allow water to flow through the treatment plant, although conductivity probes or switches permit a positive flow and avoid the risk of slippage. [Pg.359]

To ensure that the water flows through the whole of the system as smoothly as possible and with the minimum of turbulence, it is vital that the layout of pipework should be planned before fabrication starts. It should not be the result of haphazard improvisation to avoid more and more obstacles as construction proceeds. Pipe runs should be minimised or run as directly as possible with every effort made to avoid features that might act as turbulence raisers. For this reason the number of flow controllers, process probes, bends, branches, valves, flanges, intrusive fittings, or mechanical deformation or damage to the pipework, should be kept to a minimum. [Pg.73]

Reference electrodes The generally accepted criterion for the effectiveness of a cathodic-protection system is the structure/electrolyte potential (Section 10.1). In order to determine this potential it is necessary to make a contact on the structure itself and a contact with the electrolyte (soil or water). The problem of connection to the structure normally presents no difficulties, but contact with the electrolyte must be made with a reference electrode. (If for example an ordinary steel prol e were used as a reference electrode, then inaccuracies would result for two main reasons first, electrochemical action between the probe and the soil, and second, polarisatibn of the probe owing to current flow through the measuring circuit.)... [Pg.243]

The probes are 1-mm nickel rods with square ends. They are powered by 100-V dc power supplies. The positive side of the power supply is connected to the cell body. (This makes the probes cathodic to the case.) The control modules have 12 M3 resistance so that when contact is made, less than 10 mA flows through the circuit this is not a dangerous current. (The use of a 12-V power supply and lower resistance modules did not reliably deliver a control signal when contact was made this may have been caused by the buildup of a passive film on the probe that was not cleared by the lower voltage.)... [Pg.534]

Electrical conductance probes These can be either flush probes or wire probes. Flush probes are imbedded in a nonconducting wall, with one electrode connected to a voltage source and the second through a precision resistor to ground (Telles and Dukler, 1970 Chu and Dukler, 1974). Wire probes use closely spaced, nearly parallel conducting wires of small diameter, which are positioned normal to the flow (Brown et al., 1978). [Pg.196]

The pitot tube is a device for measuring v(r), the local velocity at a given position in the conduit, as illustrated in Fig. 10-1. The measured velocity is then used in Eq. (10-2) to determine the flow rate. It consists of a differential pressure measuring device (e.g., a manometer, transducer, or DP cell) that measures the pressure difference between two tubes. One tube is attached to a hollow probe that can be positioned at any radial location in the conduit, and the other is attached to the wall of the conduit in the same axial plane as the end of the probe. The local velocity of the streamline that impinges on the end of the probe is v(r). The fluid element that impacts the open end of the probe must come to rest at that point, because there is no flow through the probe or the DP cell this is known as the stagnation point. The Bernoulli equation can be applied to the fluid streamline that impacts the probe tip ... [Pg.294]

Cavity size (volume) Approx. 50 L Delivered power 1500 W Max. output power 1200 W Temperature control Outside IR remote sensor Immersed fiber-optic probe (optional) Pressure measurement Pneumatic pressure sensor (optional) Cooling system Air flow through cavity 100 m3 h1 External PC Optional not required as integrated key panel is standard equipment ... [Pg.41]

Currently, systems for measuring the hydrogen content in molten aluminum are commercially available. In the most common systems, an inert gas (usually nitrogen) flows through a probe and over the molten metal, such that chemical equilibrium between the hydrogen partial pressure in the gas and the concentration of hydrogen in the... [Pg.514]


See other pages where Probe flow-through probes is mentioned: [Pg.430]    [Pg.259]    [Pg.474]    [Pg.16]    [Pg.560]    [Pg.216]    [Pg.106]    [Pg.1295]    [Pg.1313]    [Pg.70]    [Pg.74]    [Pg.79]    [Pg.266]    [Pg.73]    [Pg.348]    [Pg.997]    [Pg.133]    [Pg.504]    [Pg.42]    [Pg.226]    [Pg.374]    [Pg.29]    [Pg.247]    [Pg.98]    [Pg.87]    [Pg.240]    [Pg.12]    [Pg.432]    [Pg.529]    [Pg.529]    [Pg.274]    [Pg.219]    [Pg.172]    [Pg.276]    [Pg.277]    [Pg.284]    [Pg.422]    [Pg.422]   
See also in sourсe #XX -- [ Pg.41 ]




SEARCH



Flow probes

Flow-through

© 2024 chempedia.info