Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Pressure tests manual

In order to ensure that pressure tests are carried out correctly, a Pressure Test Manual can be developed. Table 2.2 provides a suggested Table of Contents for such a manual. [Pg.38]

Safety valves are normally manually tested at periods between 1 and 6 months and pressure-tested every 12 months. Boiler operating pressures are typically 5 to 10% less than safety valve set pressures to avoid valve weeping. [Pg.79]

AATCC test method 127-2003. Water resistance hydrostatic pressure test. AATCC technical manual, vol. 82. 2007. [Pg.54]

Sample size is dictated by the test method to be used. One litre (32 oz) bottles or cans are generally used for manual vapor pressure testing. The mini-vapor pressure methods ne a much smaller sample and it can be taken in a 125 mL (4 oz) bottle. See Fig. 10. [Pg.982]

The client insisted on a test. The smoke detectors worked, and the explosive charge operated, but the cutter did not cut the rupture disc. The explosive charge could not develop enough pressure because the volume between it and the rupture disc was too great. The volume had been increased as the result of a change in design installation of a device for discharging the halon manually. [Pg.278]

Various aspects of in vitro gas production test have been reviewed by Getachew et al. [33], and these authors reported that gas measurement were centered on investigations of rumen microbial activities using manometric measurements and concluded that these methods do not have wide acceptability in routine feed evaluation since there was no provision for the mechanical stirring of the sample during incubation. Another in vitro automated pressure transducer method for gas production measurement was developed by Wilkins [34], and the method was validated by Blummel and Orskov [35] and Makkar et al. [36]. There are several other gas-measuring techniques such as (i) Flohenheim gas method or Menke s method [37] (ii) liquid displacement system [38] (iii) manometric method [39] (iv) pressure transducer systems manual [40], computerized [41], and combination of pressure transducer and gas release system [42]. [Pg.250]

A steady pressure should be maintained within the flow cell when performing the detector linearity test. Filling the flow cell with the test solution manually with a syringe sometimes leads to pressure fluctuations and hence unstable readings. An alternative is to use the pump to deliver test solution to the flow cell. However, this option requires a large volume of test solution. [Pg.183]


See other pages where Pressure tests manual is mentioned: [Pg.24]    [Pg.38]    [Pg.39]    [Pg.24]    [Pg.38]    [Pg.39]    [Pg.333]    [Pg.379]    [Pg.47]    [Pg.269]    [Pg.379]    [Pg.363]    [Pg.333]    [Pg.485]    [Pg.197]    [Pg.237]    [Pg.692]    [Pg.986]    [Pg.7]    [Pg.405]    [Pg.99]    [Pg.2305]    [Pg.447]    [Pg.118]    [Pg.167]    [Pg.745]    [Pg.244]    [Pg.24]    [Pg.97]    [Pg.260]    [Pg.280]    [Pg.418]    [Pg.132]    [Pg.375]    [Pg.318]    [Pg.141]    [Pg.299]    [Pg.116]    [Pg.15]    [Pg.73]    [Pg.99]    [Pg.128]    [Pg.128]    [Pg.766]    [Pg.4]    [Pg.390]   
See also in sourсe #XX -- [ Pg.38 , Pg.39 ]




SEARCH



Pressure test

Testing pressure

© 2024 chempedia.info