Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Perturbation, Fourier analysis

Figure 3. Frequency spectra obtained by the Fourier analysis of the progressions in Fig. 2 Curve (b) shows the characteristic spectrum of an anharmonic progression while curve (a) is obviously perturbed [7],... Figure 3. Frequency spectra obtained by the Fourier analysis of the progressions in Fig. 2 Curve (b) shows the characteristic spectrum of an anharmonic progression while curve (a) is obviously perturbed [7],...
The off-diagonal elements depend on the x and y components of the local field, which contain many fluctuating components oscillating at different frequencies. The parts which oscillate at the ESR frequency ( ) induce transition between the a- and P-states. By making a Fourier analysis of V(f) and using time-dependent perturbation theory [22], the transition probability between the a- and p-states (P< ) is given by... [Pg.108]

The first widely used global method was the Fourier Amplitude Sensitivity Test (FAST) (for a review see [83]). In the FAST method, all rate parameters were simultaneously perturbed by sine functions with incommensurate frequencies. Fourier analysis of the solution of the model provided the variance crf(t) of concentration i, and also the variance o- (t) of c, arising from the uncertainty in the /th parameter. Their ratio... [Pg.323]

Figure 8.4 Impedance results obtained for the time-domain results presented in Figure 8.3 by use of the Fourier analysis presented in Section 7.3.3 with potential perturbation amplitude AV as a parameter a) Nyquist representation b) real part of the impedance as a function of frequency and c) imaginary part of the impedance as a function of frequency. Figure 8.4 Impedance results obtained for the time-domain results presented in Figure 8.3 by use of the Fourier analysis presented in Section 7.3.3 with potential perturbation amplitude AV as a parameter a) Nyquist representation b) real part of the impedance as a function of frequency and c) imaginary part of the impedance as a function of frequency.
Fig. 4. Fourier-analysis of the spectrum shown in Pig. 3. The sequence exhibits around 65 cm the fundamental frequency and at 130 cm and 195 cm the corresponding second and third harmonic. The insert in the figure shows the unharmonic progression at a higher resolution. The spacings indicate that the sequence is perturbed. Fig. 4. Fourier-analysis of the spectrum shown in Pig. 3. The sequence exhibits around 65 cm the fundamental frequency and at 130 cm and 195 cm the corresponding second and third harmonic. The insert in the figure shows the unharmonic progression at a higher resolution. The spacings indicate that the sequence is perturbed.
In order to evaluate the changes to the performance of the ion trap, caused by the perturbation to the trapping field due to the addition of holes in the ring electrode to enable our fluorescence measurements, we have carried out ion trajectory calculations in several models of the modified Esquire 3000-1- QIT. Franzen [87,91,92] demonstrated previously the utility of ion trajectory calculations in investigation of the modified hyperbolic angle trap, but considered only ion motion in the axial direction. Here, we discuss results from several SIMION models, which have been constructed with different numbers and sizes of holes in the ring electrode. Fourier analysis of ion trajectories is used to determine secular frequencies, frequency shifts. [Pg.263]

A number of other operational problems exist when using the FFT algorithm. The most important of these, as far as electrochemistry is concerned, is due to the inherently nonlinear nature of the system. When Eq. (56) is used to measure the impedance with an arbitrary time domain input function (i.e. not a single-frequency sinusoidal perturbation), then the Fourier analysis will incorrectly ascribe the harmonic responses due to system nonlinearity, to input signal components which may or may not be present at higher frequencies. As a consequence, the measured impedance spectrum may be seriously in error. [Pg.165]

New techniques for data analysis and improvements in instrumentation have now made it possible to carry out stmctural and conformational studies of biopolymers including proteins, polysaccharides, and nucleic acids. NMR, which may be done on noncrystalline materials in solution, provides a technique complementary to X-ray diffraction, which requires crystals for analysis. One-dimensional NMR, as described to this point, can offer structural data for smaller molecules. But proteins and other biopolymers with large numbers of protons will yield a very crowded spectrum with many overlapping lines. In multidimensional NMR (2-D, 3-D, 4-D), peaks are spread out through two or more axes to improve resolution. The techniques of correlation spectroscopy (COSY), nuclear Overhausser effect spectroscopy (NOESY), and transverse relaxation-optimized spectroscopy (TROSY) depend on the observation that nonequivalent protons interact with each other. By using multiple-pulse techniques, it is possible to perturb one nucleus and observe the effect on the spin states of other nuclei. The availability of powerful computers and Fourier transform (FT) calculations makes it possible to elucidate structures of proteins up to 40,000 daltons in molecular mass and there is future promise for studies on proteins over 100,000... [Pg.165]

Another more sophisticated approach is to make a Fourier Transform analysis of the response in the way proposed by Bond et al. [84, 85]. In this case, the perturbation is a continuous function of time (a ramped square wave waveform) which combines a dc potential ramp with a square wave of potential that can be described as a combination of sinusoidal functions. Under these conditions, the faradaic contribution to the response generates even harmonics only (i.e., the non-faradaic current goes exclusively through odd harmonics). Thus, the analysis of the even harmonics will provide excellent faradaic-to-non-faradaic current ratios. [Pg.557]

Lastly, it should be pointed out that there are often significant differences between H positions determined by X-ray analysis and those determined from neutron data. Neutron diffraction provides true nuclear positions, whereas X-ray diffraction measures the electron density distribution. Thus, X-ray Fourier maps often give H peaks that, because of the perturbing influence of the M-H bonding electrons, appear closer to the M atoms than they really are A thorough analysis of this effect has... [Pg.7]

The traditional way is to measure the impedance curve, Z(co), point-after-point, i.e., by measuring the response to each individual sinusoidal perturbation with a frequency, to. Recently, nonconventional approaches to measure the impedance function, Z(a>), have been developed based on the simultaneous imposition of a set of various sinusoidal harmonics, or noise, or a small-amplitude potential step etc, with subsequent Fourier- and Laplace transform data analysis. The self-consistency of the measured spectra is tested with the use of the Kramers-Kronig transformations [iii, iv] whose violation testifies in favor of a non-steady state character of the studied system (e.g., in corrosion). An alternative development is in the area of impedance spectroscopy for nonstationary systems in which the properties of the system change with time. [Pg.189]

For the purpose of linear analysis, we represent the perturbation quantities by their Fourier- Laplace transform via... [Pg.26]

Cd(II) plastocyanin does not, nnder the given experimental conditions. This is an illnstration of an application of PAC spectroscopy to the stndy of protem-protem interactions, and it is even possible to estimate the dissociation constant directly from the spectroscopic data. Note that in this example the pertnrbation function and not the Fourier transformed data, is displayed in the figure, as the perturbation function illustrates the effect of dynamics more clearly. Note that the data analysis is almost always carried out on the perturbation function. [Pg.6277]

We assume that the flat interface between the two fluids, now designated as z = 0, is perturbed with an arbitrary infinitesimal perturbation of shape. As usual, for a linear stability analysis, we consider only a single Fourier mode in each of the x and y directions, with the wave number (or wavelength) as a parameter in the stability analysis. Hence we consider a perturbation of the form... [Pg.826]


See other pages where Perturbation, Fourier analysis is mentioned: [Pg.164]    [Pg.325]    [Pg.93]    [Pg.131]    [Pg.121]    [Pg.139]    [Pg.179]    [Pg.100]    [Pg.1219]    [Pg.172]    [Pg.173]    [Pg.409]    [Pg.34]    [Pg.25]    [Pg.85]    [Pg.133]    [Pg.46]    [Pg.269]    [Pg.59]    [Pg.169]    [Pg.247]    [Pg.282]    [Pg.581]    [Pg.357]    [Pg.174]    [Pg.18]    [Pg.395]    [Pg.113]    [Pg.251]    [Pg.584]    [Pg.586]    [Pg.65]    [Pg.357]    [Pg.24]   
See also in sourсe #XX -- [ Pg.280 ]




SEARCH



Fourier analysis

© 2024 chempedia.info