Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Output dial

The Signal and Output dials should have the same reading. Let s assume that this is an "air-to-close valve." That means that more air pressure closes the valve. Typically, the supply air pressure is 30 psig. If the air output pressure to the control valve is 3 psig, the valve should be 100 percent open. If the air output pressure to the control valve is 15 psig, the valve would be shut. We say the valve operates... [Pg.512]

For many (but not all) control valves, you can tell if air pressure is supposed to open or close the valve. If the tubing to the diaphragm is on the top, as shown in Fig. 38.4, then the air pressure is usually (but not always) going to close the valve. When in doubt, ask the panel operator to move the valve. Touch the stem with your gloved finger and see what happens to the air pressure of the output dial. [Pg.513]

The single channel analyzer has two dial settings a LEVEL dial and a WINDOW dial. For example, when the level is set at 2 volts, and the window at 0.2 volts, the analyzer will give an output pulse only when the input pulse is between 2 and 2.2 volts. The output pulse is usually a standardized height and width logic pulse, as shown in Figure 11. [Pg.48]

Figures 5 and 6 show typical performance of Hereon equipment in dispensing flake and sticker. Output in each case is linear with the dial setting on the equipment, allowing very close control of the amount of flake and sticker dispensed in flight. Figures 5 and 6 show typical performance of Hereon equipment in dispensing flake and sticker. Output in each case is linear with the dial setting on the equipment, allowing very close control of the amount of flake and sticker dispensed in flight.
FIGURE 13. Comparison of pulsed GC-AA output continuous UV detection output. The pulsed output was obtained using a 0.85 ml injection at 382-ppm trimethylarsine. The power supply dial was set at 2000 °C and the pulse program was set to give a 7-s pulse on a 30-s interval. The background corrector of the detector was not employed and, as indicated by the arrow, molecular absorption of the 193,7-nm EDL-As can be detected. The continuous output was obtained using 0.20 ml of 2536-p.p.m. trimethylarsine with the furnace at ambient temperature and very low attenuation of signal to the recorder... [Pg.203]

Radium itself has no commercial applications. Its compounds are used mainly for medical purposes and as a source of radon these absorb some 85 per cent of the world s output 10 per cent is used for rendering dials of instruments and other objects luminous, the remaining 5 per cent being used for scientific and miscellaneous purposes. One mg of radium emits 22 2 X io8 a-particles per min. Its half-life is 1610 years. [Pg.316]

Instruments. A few instruments are commercially available. Response detection generally takes one of three forms (a) a calibrated dial to increase or decrease capacity to maintain (b) a meter to record the change in current I t (Fig. 5.13) or some other related parameter and (c) a beat-frequency oscillator where the difference between reference frequency and the working frequency is measured using a frequency-discriminator output signal. One requirement is that the oscillator have good frequency stability. [Pg.133]

Although there is a large literature on identifiability for linear systems with constant coefficients, less has been done on nonlinear systems. Two general properties should be remembered. Whereas for linear systems one can substitute impulsive inputs for the experimental inputs for die analysis of identifiability, one cannot do that for nonlinear systems. One must analyze the input-output experiment for the actual inputs used. That is a drawback. On the other hand, experience shows dial frequently the introduction of nonlinearities makes a formerly nonidentifiable model identifiable for a given input-output experiment. Two methods are available. [Pg.317]

The machine-based PSFs include characteristics of the human-system interface and the responses that the machine is designed to provide. Malfunctioning indicators are not a machine-based PSF, they are a situation-based PSF. However, poorly designed indicators are a machine-based PSF. In this PSF flame-work, HSI includes control panels and other traditional HSI elements, but it also includes the accessibility (as designed) of machine components and plant areas. This expanded definition of HSI covers humans obtaining output fl om displays and charts, and it also covers humans providing input to the machine through software, buttons and dials, mechanical tools, or other contact. [Pg.247]

As shown in Figure 1, an SFA system consists of five basic components sample changer, peristaltic pump, analytical manifold, detector, and data output system. The manifold contains the mixing coils, dial-yzer, heater, and other components through which sample and reagents are pumped and where sample clean-up and reaction takes place. The data output... [Pg.4332]

The output signal, which is proportional to the average diameter of the specimen, is fed into an existing true stress-strain computer and can be read visually on a graduated dial or, when used in conjunction with load data from the testing machine, can provide an autographic curve of true stress-strain on a Baldwin MD-2 recorder. [Pg.578]

A tape measure can be used to measure distances from about 20 cm up to several meters. Dial calipers are useful for closely measuring extruded products, dimensions of the extruder screws, extrusion tooling, etc. A stopwatch is an indispensable tool for measuring screw speed, line speed, blender calibration, etc. A scale can be used to measure the output of an extruder. A small voltmeter is very useful in making sure that voltage and resistance levels of various components are at their required values. A millivolt source can be used for verifying thermocouple inputs and to check controller response and line continuity. [Pg.775]

Let the ideal gas equation be used to convert experimentally measured V, n, and T into output let the result be 5774 pascals. Upon the fourth measurement, the chemist finds the pressure dial to read 5654 pascals—the ideal gas equation overestimated by 120 pascals. Then if p is measured subsequent times, the results should equate with the ideal number minus a correction of 120 pascals. If the van der Waals equation is used instead to anticipate p, a similar situation arises. The correction should be smaller as account has been taken of the nonideality. [Pg.77]

Many of the electronic measuring instruments outlined in this chapter, e.g. vernier calipers, micrometers and dial test indicators, can be equipped to output data either through available data cables or by wireless transmitters/receivers. This data can be printed out or sent to a computer and collected and used by quality control to provide statistical information or for the generation of inspection certificates. [Pg.84]

Fig. 7.6 Sensitivity testing. The pacing rate is set to a rate less than the intrinsic ventricular rate and the sensitivity dial is slowly adjusted. With the sensitivity set to 7mV, the first QRS complex is sensed by the temporary pacemaker and causes inhibition of the pacing output. With the sensitivity set to 8 mV (which reduces sensitivity) the second QRS complex (arrow) is not sensed and the pacemaker delivers a pacing output at the programmed rate. The pacemaker then begins to output asynchronously the second pacing stimulus does not result in ventricular capture because it is delivered during the ventiicular refractory period. In this case the intrinsic ventricular electrogram has an amplitude between 7 and 8 mV. If no intrinsic ventricular rhjrthm is present, sensitivity testing cannot be performed. Fig. 7.6 Sensitivity testing. The pacing rate is set to a rate less than the intrinsic ventricular rate and the sensitivity dial is slowly adjusted. With the sensitivity set to 7mV, the first QRS complex is sensed by the temporary pacemaker and causes inhibition of the pacing output. With the sensitivity set to 8 mV (which reduces sensitivity) the second QRS complex (arrow) is not sensed and the pacemaker delivers a pacing output at the programmed rate. The pacemaker then begins to output asynchronously the second pacing stimulus does not result in ventricular capture because it is delivered during the ventiicular refractory period. In this case the intrinsic ventricular electrogram has an amplitude between 7 and 8 mV. If no intrinsic ventricular rhjrthm is present, sensitivity testing cannot be performed.
The automatic voltage maintenance on the output of voltage regulators is achieved by dial settings (located on the control panel) of the adjustable resistance and reactance elements of a unit called the line drop compensator. [Pg.1107]

What is the nature of the data traffic over the diai-up iine Ali interaction between the LAN server and remote PC traveis over the diai-up iink. Only keystrokes from the remote PC keyboard and output for the remote PC monitor travel over the dial-up link. [Pg.2123]

To test for this condition, the receiver cable is disconnected from the EXT. Det. connector and held against it so the cable shield circuit is made but not the inner conductor circuit. The receiver output should be quite low. The cable is then connected normally and the R dial moved from the null position just sufficiently to duplicate the receiver output observed above. The magnitude of the R dial deviation is then a good estimate of the error caused by the leakage. [Pg.9]


See other pages where Output dial is mentioned: [Pg.402]    [Pg.402]    [Pg.188]    [Pg.140]    [Pg.28]    [Pg.188]    [Pg.333]    [Pg.383]    [Pg.914]    [Pg.41]    [Pg.210]    [Pg.220]    [Pg.19]    [Pg.109]    [Pg.238]    [Pg.145]    [Pg.982]    [Pg.175]    [Pg.431]    [Pg.60]    [Pg.412]    [Pg.310]    [Pg.374]    [Pg.110]    [Pg.111]    [Pg.426]    [Pg.494]    [Pg.10]   
See also in sourсe #XX -- [ Pg.402 ]




SEARCH



DIAL

© 2024 chempedia.info