Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Multisyringe separation system

FIGURE 9.10 Schematic diagram of a multisyringe separation system using solenoid valves rather than a multiposition valve. Normally open and normally closed ports on the solenoid valves are marked with open and closed circles, while the common port is unmarked. System is shown with wash and eluent solutions for Sr determination using a Sr-Resin separation column. [Pg.537]

Another multisyringe FI separation system design has been used in the analysis of stable and radioactive Y, using an extraction-chromatographic material containing HDEHP adsorbed on a C18 support.123 Separated samples were analyzed off-line by ICP-AES and proportional counting. This system used four syringe pumps in parallel. [Pg.536]

Fajardo, Y., Ferrer, L., Gomez, E. et al. (2008). Development of an automatic method for americium and plutonium separation and preconcentration using a multisyringe flow injection analysis multipump flow system. Anal. Chem. 80, 195-202. [Pg.113]

FIGURE 5.9 MSC system for isocratic separation. MS multisyringe burette M manometer V1-V3 solenoid valves El two-way connector E2 three-way solenoid valve MP mobile phase S sample W waste RP-18 monolithic column DAD diode array detector and SC sample coil. (Reprinted from Gonzalez-San Miguel, H. M., J. M. Alpizar-Lorenzo, V. Cerda-Martin, 2007. Talanta 72 296-300. With permission.)... [Pg.114]


See other pages where Multisyringe separation system is mentioned: [Pg.536]    [Pg.23]    [Pg.89]    [Pg.113]    [Pg.336]   
See also in sourсe #XX -- [ Pg.537 ]