Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrument output

A typical layout controlled by the central microprocessor (CPU). Electrical inputs are received from the keyboard, mouse, or instrument. Outputs go to the video screen, printer, and the instrument. Memory and software are utilized hy the CPU on command. [Pg.308]

For non-compendial procedures, the performance parameters that should be determined in validation studies include specificity/selectivity, linearity, accuracy, precision (repeatability and intermediate precision), detection limit (DL), quantitation limit (QL), range, ruggedness, and robustness [6]. Other method validation information, such as the stability of analytical sample preparations, degradation/ stress studies, legible reproductions of representative instrumental output, identification and characterization of possible impurities, should be included [7], The parameters that are required to be validated depend on the type of analyses, so therefore different test methods require different validation schemes. [Pg.244]

Every measurement has noise—random changes in the results that is, if an instrument is left to make measurement without any sample, the baseline will not be a straight line but will be a random recording of instrument output or noise. Figure 14.1 shows the noise in the baseline of a gas chromatograph at maximum sensitivity. When an absorption or peak is vastly larger than the noise, there is little question of its authenticity. When it is not much larger than the noise, there is question of its authenticity is it real or is it noise ... [Pg.293]

A portable electronic data acquisition system was transported to the plant site and connected to the extruder panel. All available instrument outputs from the panel were connected in parallel with the acquisition system. Data collected included barrel zone temperatures, screw speed, motor current, pressure at the entry to the pump, transfer line temperature, and gear pump temperature. Process data were collected at a frequency of once every five seconds. [Pg.584]

Multiple measurements allow the analyst to detect the presence of an inter-ferent. For example, suppose an instrument outputs two responses for each sample, as shown in Figure 5-2. Assume the pure component response is obtained firom the measurement of a pure analyte of inteicst. Given any sample containing only this analyte, the same relative magnitudes for measurements and r, are observed. If the observed response shown in Figure 5.2 is encountered, the analyst will know that a problem exists. This concept similarly extends to mixtures. Tlie detection of interferents can fail if, for example, an inter-ferent has the same relative response on the two variables as one of the other species. However, the likelihood of this tjpe of failure decreases as the number of measurements increase, and as care is taken in choosing the measurements. [Pg.276]

Time between the valve switching from zero to calibrate gas and the onset of change at instrument output. [Pg.80]

Thus, if the instrument output records a reading of G = 5.50kg/min then the estimate of the true value of the rate of flow will be the corresponding value of GT 2confidence limits. The total error obtained in the calibration may be split into two parts, viz. the bias (or systematic error), i.e. 5.50 - 5.68 = - and the error due to imprecision (random error or... [Pg.534]

When applied to the performance of an instrument these terms signify the range over which an input can be varied without the instrument output responding in any way. They are usually expressed in terms of the percentage of span of the instrument and should not be confused with the concept of dead time (Section 7.6). [Pg.535]

Hayward et al. (2002) demonstrated that PTR-MS could reliably measure a wide range of VOCs and with a time resolution sufficiently fast to capture the dynamics of many environmental processes (e.g., the light dependency of isoprene emissions from vegetation). They also demonstrated that the components of the instrument output (signal plus noise) were easily characterized, enabling a simple interpretation of measurements. [Pg.73]

Fig. 2 Variation of the instrument output with the incoming gas flow... Fig. 2 Variation of the instrument output with the incoming gas flow...
The relationship between the instrument output, y, and the input analyte concentration, x, being considered as non-linear, correction polynomials of the second order are fitted to the experimental curves ... [Pg.228]

All operations performed on a sample must be recorded in a notebook or computer system. Chromatograms, spectra and other instrumental outputs must be labelled with the sample identification. [Pg.297]

A summary of all data and findings for the experiments listed in the validation protocol. This should include a complete listing of specific instrumentation, actual reagent lots, standards, equipment and supplies used in the performance of the validation. All results should be provided with references to the original notebook entries. Example chromatograms, spectra, or other instrument outputs should be provided. [Pg.177]

Finally, there is the need for proper documentation, which can be in written or electronic forms. These should cover every step of the measurement process. The sample information (source, batch number, date), sample preparation/analytical methodology (measurements at every step of the process, volumes involved, readings of temperature, etc.), calibration curves, instrument outputs, and data analysis (quantitative calculations, statistical analysis) should all be recorded. Additional QC procedures, such as blanks, matrix recovery, and control charts, also need to be a part of the record keeping. Good documentation is vital to prove the validity of data. Analyt-... [Pg.27]

A typical instrument output of mass versus increasing temperature is shown in Figure 5.3. This figure shows three distinct... [Pg.114]

Before actual data can be fit to a model, extraneous effects manifested in the trace must be removed, such as the shift in baseline as a result of the change in heat capacity of the sample during the transformation (see section 3.7.2). It may, for some device designs (e.g. post-type DTA), be difficult to purify the instrument output to represent only the latent heat from the transformation because of random baseline float. Hence, the data set fitting a particular model is a necessary but insufficient criterion for guaranteeing that the model describes the measured phenomenon. [Pg.144]

However, it is this small temperature difference between sample and reference that is measured as a function of temperature during the applied temperature scan. It is converted by the instrument s software, using instrument calibration data, to the difference between the power absorbed or released from the sample and that absorbed or released from the reference. The final instrument output is a thermogram of this differential power plotted against temperature, as for PC instruments. [Pg.734]

In the context of this discussion, noise refers to random variations in the instrument output due not only to electrical fluctuations but also to such other variables as the way the operator reads the meter, the position of the cell in the light beam, the temperature of the solution, and the output of the source. [Pg.798]

Using the procedure outlined above, we may generate a total of Ki > k inferred state histories , Z from the k original instrument outputs, z. We now sort the differential equations (DEs) defining the nominal model (equation (24.3)) into sets on the basis of the state variables they contain ... [Pg.317]

Emission Measurement. Exhaust gases leaving the vehicle tailpipe were diluted and cooled by addition to a stream of air in the constant volume sampler. Exhaust emission concentrations in the diluted stream were monitored continuously as the vehicle was operated over the 1975 FTP ( see Table II for instrumentation). Instrument output was scanned every 0.5 sec, and exhaust mass emissions corrected for temperature and pressure were calculated by numerical integration by digital computer (2). [Pg.79]

Since the aggregation process is a function of time, samples of the feed suspension were taken prior to the filtration experiments and immediately analysed with the Malvern Mastersizer. Size distribution information was directly available from the instrument output. [Pg.122]

The speed of response is generally measured by plotting the instrument output as a function of time in response to step changes in sample concentration. Such a plot is shown in Fig. 7. [Pg.161]


See other pages where Instrument output is mentioned: [Pg.247]    [Pg.672]    [Pg.38]    [Pg.16]    [Pg.377]    [Pg.570]    [Pg.275]    [Pg.532]    [Pg.467]    [Pg.32]    [Pg.40]    [Pg.17]    [Pg.734]    [Pg.85]    [Pg.128]    [Pg.17]    [Pg.778]    [Pg.779]    [Pg.531]    [Pg.95]    [Pg.209]    [Pg.768]    [Pg.771]   
See also in sourсe #XX -- [ Pg.276 ]




SEARCH



Output signals from instruments

© 2024 chempedia.info