Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Errors dispersion

The simplest procedure, and the one almost universally applied, is to perform repeat runs at a set of conditions to establish the presence and extent of (presumably) random error. The arithmetic average of this set of repeat readings is assumed to be a better approximation of the true value. Unfortunately, each reading adds a substantial unproductive burden to the experimental program so that repeat runs are few in any investigation. The preferred procedure is to perform a few runs at one convenient experimental condition, and to calculate some measure of error dispersion at that point. Since there are rarely enough repeat measurements at any other point to do a serious evaluation of the error, a simplistic measure such as the average error is often based on the error estimate at this one point ... [Pg.27]

Now, if we want to obtain the dispersion of the individuals observed values (in a given point) from the dispersion of individual predicted model, then, on the observed dispersion one should add also the error dispersion (of predicted vs. observed values) ... [Pg.175]

In Dynamic Spatial Reconstructor at the expense of use 2D matrix of detectors there was the opportunity to use a divergent cone beam of source emission. This system had a number of lacks. In particular the number of projections is rigidly limited by the number of x-ray sources. The dispersion of source emission results in errors of data collected.. However the system confirmed basic advantages of application of conic beams and 2D matrices of detectors for collecting information about 3D object. [Pg.217]

The main sources of error which define the accuracy are counting statistics in tracer concentration measurements, the dispersion of the tracer cloud in the flare gas stream, and the stationarity of the flow during measurements. [Pg.1055]

In the next several sections, the theoretical distributions and tests of significance will be examined beginning with Student s distribution or t test. If the data contained only random (or chance) errors, the cumulative estimates x and 5- would gradually approach the limits p and cr. The distribution of results would be normally distributed with mean p and standard deviation cr. Were the true mean of the infinite population known, it would also have some symmetrical type of distribution centered around p. However, it would be expected that the dispersion or spread of this dispersion about the mean would depend on the sample size. [Pg.197]

Chemical analysis of the metal can serve various purposes. For the determination of the metal-alloy composition, a variety of techniques has been used. In the past, wet-chemical analysis was often employed, but the significant size of the sample needed was a primary drawback. Nondestmctive, energy-dispersive x-ray fluorescence spectrometry is often used when no high precision is needed. However, this technique only allows a surface analysis, and significant surface phenomena such as preferential enrichments and depletions, which often occur in objects having a burial history, can cause serious errors. For more precise quantitative analyses samples have to be removed from below the surface to be analyzed by means of atomic absorption (82), spectrographic techniques (78,83), etc. [Pg.421]

Both the wavelength dispersive and energy dispersive spectrometers are well suited for quaUtative analysis of materials. Each element gives on the average only six emission lines. Because the characteristic x-ray spectra are so simple, the process of allocating atomic numbers to the emission lines is relatively simple and the chance of making a gross error is small. [Pg.382]

The methodical elaboration is included for estimation of random and systematic errors by using of single factor dispersion analysis. For this aim the set of reference samples is used. X-ray analyses of reference samples are performed with followed calculation of mass parts of components and comparison of results with real chemical compositions. Metrological characteristics of x-ray fluorescence silicate analysis are established both for a-correction method and simplified fundamental parameter method. It is established, that systematic error of simplified FPM is less than a-correction method, if the correction of zero approximation for simplified FPM is used by preliminary established correlation between theoretical and experimental set data. [Pg.234]

Figure 7. Graph of Number of Compounds against Error % obtained from the Linear Regression of the Graph of the Corrected Peak Dispersion against the Reciprocal of the Diffusivity... Figure 7. Graph of Number of Compounds against Error % obtained from the Linear Regression of the Graph of the Corrected Peak Dispersion against the Reciprocal of the Diffusivity...
When the relieving scenarios are defined, assume line sizes, and calculate pressure drop from the vent tip back to each relief valve to assure that the back-pressure is less than or equal to allowable for each scenario. The velocities in the relief piping should be limited to 500 ft/sec, on the high pressure system and 200 ft/sec on the low pressure system. Avoid sonic flow in the relief header because small calculation errors can lead to large pressure drop errors. Velocity at the vent or flare outlet should be between 500 ft/sec and MACH 1 to ensure good dispersion. Sonic velocity is acceptable at the vent tip and may be chosen to impose back-pressure on (he vent scrubber. [Pg.379]

Errors in the molecular weight data from HPSEC are usually due to improperly prepared samples, column dispersity, or flow rate variations. The sample to be analyzed should be completely dissolved in the mobile phase and filtered prior to injection onto the column. A plugged column inlet frit will invalidate results. In addition, do not load the column with excess sample. Column overloading affects the accuracy of data by broadening peaks, reducing resolution, and increasing elution volume. For best results, the concentration of the injected sample should be as low as possible while still providing adequate... [Pg.82]

Other models (or combinations of them) are often employed when computers are used to analyze dispersal. These can give an acceptable degree of accuracy when combined with detailed weather data. Short-exposure modeling is the most difficult and is liable to the greatest degree of error. It is for this reason that such models are not accurate when dealing with odor nuisances. The problem of modeling odor dispersal is dealt with below. [Pg.760]

FIGURE 11.3 One-way ANOVA (analysis of variance). One-way analysis of variance of basal rates of metabolism in melanophores (as measured by spontaneous dispersion of pigment due to G,.-protein activation) for four experiments. Cells were transiently transfected with cDNA for human calcitonin receptor (8 j-ig/ml) on four separate occasions to induce constitutive receptor activity. The means of the four basal readings for the cells for each experiment (see Table 11.4) are shown in the histogram (with standard errors). The one-way analysis of variance is used to determine whether there is a significant effect of test occasion (any one of the four experiments is different with respect to level of constitutive activity). [Pg.231]


See other pages where Errors dispersion is mentioned: [Pg.171]    [Pg.46]    [Pg.48]    [Pg.38]    [Pg.40]    [Pg.781]    [Pg.171]    [Pg.46]    [Pg.48]    [Pg.38]    [Pg.40]    [Pg.781]    [Pg.110]    [Pg.59]    [Pg.338]    [Pg.107]    [Pg.327]    [Pg.384]    [Pg.344]    [Pg.504]    [Pg.1418]    [Pg.1427]    [Pg.1653]    [Pg.36]    [Pg.147]    [Pg.405]    [Pg.83]    [Pg.15]    [Pg.203]    [Pg.341]    [Pg.347]    [Pg.355]    [Pg.357]    [Pg.349]    [Pg.129]    [Pg.172]    [Pg.189]    [Pg.330]    [Pg.276]    [Pg.453]    [Pg.518]   
See also in sourсe #XX -- [ Pg.84 , Pg.245 , Pg.246 , Pg.247 , Pg.248 , Pg.249 , Pg.264 ]




SEARCH



Design of Controllable Dispersion-Error Techniques

Dispersion error functions

Errors dispersion phasing

© 2024 chempedia.info