Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Differential temperature calorimeters

In a testing context, it refers to the first detection of exothermic-activity on the thermogram. The differential scanning calorimeter (DSC) has a scan rate of I0°C/min, whereas the accelerating rate calorimeter (ARC) has a sensitivity of 0.02°C/min. Consequently, the temperature at which thermal activity is detected by the DSC can be as much as 50°C different from ARC data. [Pg.2312]

Endothermic peak temperature according to the differential scanning calorimeter method. (Speed of temperature rise 20 C/min.) he figures apply to 2-mm-thick sheet injection molded with cylinder temperature of 150°C and mold temperature 20 C. [Pg.180]

In a differential scanning calorimeter, a sample and reference material are heated in separate, but identical, metal heat sinks. The temperatures of the sample and reference material are kept the same by varying the power supplied to the two heaters. The output is the difference in power as a function of heat added. [Pg.360]

Differential Scanning Calorimeter (DSC) thermograms were obtained on a Perkin Elmer DSC-2 run at 10°C per minutes. Dynamic Mechanical Thermal Analysis (DMTA) spectra were obtained on a Polymer Labs DMTA at a frequency of 1Hz with a temperature range from -150°C to +150°C at a scan rate of 5°C per minute. [Pg.88]

Thermal Properties. The glass transition temperature (Tg) and the decomposition temperature (Td) were measured with a DuPont 910 Differential Scanning Calorimeter (DSC) calibrated with indium. The standard heating rate for all polymers was 10 °C/min. Thermogravimetric analysis (TGA) was performed on a DuPont 951 Thermogravimetric Analyzer at a heating rate of 20 °C/min. [Pg.157]

Safety studies of the graphite anode samples were performed using a Perkin-Elmer Differential Scanning Calorimeter (DSC, model Pyris 1) instrument. The temperature scanning rate was 10 C/min over a temperature range of 50 to 375°C. [Pg.301]

Adiabatic calorimeters are complex home-made instruments, and the measurements are time-consuming. Less accurate but easy to use commercial differential scanning calorimeters (DSCs) [18, 19] are a frequently used alternative. The method involves measurement of the temperature of both a sample and a reference sample and the differential emphasizes the difference between the sample and the reference. The two main types of DSC are heat flux and power-compensated instruments. In a heat flux DSC, as in the older differential thermal analyzers (DTA), the... [Pg.310]

Figure 2.6C shows the temperature difference between reference and sample as recorded by differential thermal analysis (DTA). Note also the similar differential scanning calorimeter (DSC) curve later in Figure 2.13. [Pg.21]

Fig. 1. Differential Scanning Calorimeter (DSC). Ts=sample temperature, Tr=reference temperature, T0=oven temperature. Fig. 1. Differential Scanning Calorimeter (DSC). Ts=sample temperature, Tr=reference temperature, T0=oven temperature.
Glass transition temperatures of the uv-hardened films were measured with a Perkin Elmer Model DSC-4 differential scanning calorimeter (DSC) that was calibrated with an indium standard. The films were scraped from silicon substrates and placed in DSC sample pans. Temperature scans were run from -40 to 100-200 °C at a rate of 20 ° C/min and the temperature at the midpoint of the transition was assigned to Tg. [Pg.257]

Experimental Methods Measurements of specific heat and enthalpies of transition are now usually carried out on quite small samples in a Differential scanning calorimeter (DSC). DSC is applied to two different moles of analysis, of these the one is more closely related to traditional calorimetry and is described here. In DSC an average-temperature circuit measures and controls the temperature of sample and reference holders to conform to a Organisation and Qualities... [Pg.86]

Figure 12.1 Scheme of a disk-type heat flux differential scanning calorimeter. A cell B furnace C temperature sensors S sample R reference. [Pg.172]

Figure 12.3 Schemeofa power compensation differential scanning calorimeter. A sample furnace Ar reference furnace B temperature sensor of the sample furnace Br temperature sensor of the reference furnace C resistance heater of the sample furnace Cr resistance heater of the reference furnace D cell S sample R reference. Figure 12.3 Schemeofa power compensation differential scanning calorimeter. A sample furnace Ar reference furnace B temperature sensor of the sample furnace Br temperature sensor of the reference furnace C resistance heater of the sample furnace Cr resistance heater of the reference furnace D cell S sample R reference.
G. W. H. Holme, H. K. Cammenga, W. Eysel, E. Gmelin, W. Hemminger. The Temperature Calibration of Differential Scanning Calorimeters. Thermochim. Acta 1990, 160, 1-12. [Pg.260]

H. K. Cammenga, K. Gehrich, S. M. Sarge. 4,4 -Azoxyanisolefor Temperature Calibration of Differential Scanning Calorimeters in the Cooling Mode—Yes or No . Thermochim. Acta 2006, 446, 36—40. [Pg.261]

The heat capacity is the amount of energy required to increase the temperature of a unit mass of material. It is commonly measured using a differential scanning calorimeter (DSC). The heat capacity depends on the resin type, additives such as fillers and blowing agents, degree of crystallinity, and temperature. A temperature scan for the resin will reveal the Tg for amorphous resins and the peak melting temperature and heat of fusion for semicrystalline resins. The heat capacities for LDPE and PS resins are shown in Fig. 4.15. [Pg.123]

QO2— ). Glass transition temperatures (Tg s) of copolymers I and IV were determined by the use of a differential scanning calorimeter. The samples were heated at 10 C/minute In air, to 200 C, then cooled and reheated. Glass transition data was taken from the "second heating". [Pg.157]

Microcalorimetry has gained importance as one of the most reliable method for the study of gas-solid interactions due to the development of commercial instrumentation able to measure small heat quantities and also the adsorbed amounts. There are basically three types of calorimeters sensitive enough (i.e., microcalorimeters) to measure differential heats of adsorption of simple gas molecules on powdered solids isoperibol calorimeters [131,132], constant temperature calorimeters [133], and heat-flow calorimeters [134,135]. During the early days of adsorption calorimetry, the most widely used calorimeters were of the isoperibol type [136-138] and their use in heterogeneous catalysis has been discussed in [134]. Many of these calorimeters consist of an inner vessel that is imperfectly insulated from its surroundings, the latter usually maintained at a constant temperature. These calorimeters usually do not have high resolution or accuracy. [Pg.212]


See other pages where Differential temperature calorimeters is mentioned: [Pg.1916]    [Pg.1916]    [Pg.338]    [Pg.97]    [Pg.511]    [Pg.655]    [Pg.165]    [Pg.298]    [Pg.780]    [Pg.395]    [Pg.183]    [Pg.142]    [Pg.306]    [Pg.365]    [Pg.50]    [Pg.68]    [Pg.171]    [Pg.173]    [Pg.683]    [Pg.93]    [Pg.40]    [Pg.96]    [Pg.102]    [Pg.236]    [Pg.193]    [Pg.225]    [Pg.75]    [Pg.59]   
See also in sourсe #XX -- [ Pg.841 ]




SEARCH



Calorimeter differential

Calorimeters

Differential scanning calorimeters temperature calibration

Temperature calibration of differential scanning calorimeters

Temperature differential

© 2024 chempedia.info