Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

In-circuit measurement

However, the best method to measure leakage is actually an in-circuit measurement — so that we include the secondary-side PCB traces in the measurement. The recommended procedure is as follows. [Pg.140]

Any main CT that is itndeiioaded will also add to the error in the measurement. Similarly, if provision is tnade in the primary of the summation CT to accommodate fntnre circuits but is not being utilized it tnust be left open, otherwise it will also add to the error. The impedance of the shorting terminals will add to the impedance of the circuit and will increase the total error. [Pg.477]

Strictly, the strain gauges referred to above come into this category, since in such cases the change in the measured quantity causes a corresponding change in the resistance of the element. However, the principle has a much wider application, using changes in either the inductive or capacitive reactance of electrical circuit elements. [Pg.244]

In making measurements of current flowing within a structure, it is extremely important that additional resistance, as for example a shunt, is not introduced into the circuit, as otherwise erroneous results will be obtained. One method is to use a tong test meter. Such instruments are, however, not particularly accurate, especially at low currents, and are obviously jmpracticablein thecaseof, say, a 750 mm diameter pipeline. A far moreaccurate method and onethat can beapplied to ail structures, isthe zero-resistance ammeter or, as it is sometimes called, the zero-current ammeter method. The basic circuit of such an instrument is shown in Fig. 10.47. [Pg.249]

A typical electrochemical promotion experiment includes kinetic measurements under open and closed circuit conditions as well as study of the effect of catalyst potential or work function on catalytic rate and selectivity under steady state and transient conditions. In kinetic measurements one should change the partial pressure of each reactant while... [Pg.554]

The ends of a correctly constructed electrochemical circuit measuring electrical potential difference must always have metals or conductors with identical chemical composition. It is usually reached by simple connection of two metals by copper wires. The inclusion between two metal conductors of a third metal conductor according to Volta s law does not change the difference of potentials at the output of a circuit. The difference of potentials in an electrochemical circuit at equilibrium is caused by the change of Gibbs free energy during the appropriate electrochemical reaction ... [Pg.655]

The arrival of integrated circuits with very good performance/price ratios and relatively low-cost microprocessors and memories has had a profound influence on many areas of technical endeavour. Also in the measurement and control field, modem electronic circuits were introduced on a large scale leading to very sophisticated systems and novel solutions. However, in these measurement and control systems, quite often sensors and actuators were applied that were conceived many decades ago. Consequently, it became necessary to improve these devices in such a way that their performance/price ratios would approach that of modem electronic circuits. [Pg.406]

Measure the in-circuit primary-side leakage inductance Llk in Henries. [Pg.229]

Reference electrodes provide a standard for the electrochemical measurements. For potentiometric sensors, an accurate and stable reference electrode that acts as a halfcell in the measurement circuit is critical to providing a stable reference potential and for measuring the change in potential difference across the pH sensitive membrane as the pH concentration changes. This is especially important in clinical applications such as pH measurements in the blood, heart, and brain, where the relevant physiological pH range is restricted to a very small range, usually less than one unit. [Pg.301]

Experimental Methods Measurements of specific heat and enthalpies of transition are now usually carried out on quite small samples in a Differential scanning calorimeter (DSC). DSC is applied to two different moles of analysis, of these the one is more closely related to traditional calorimetry and is described here. In DSC an average-temperature circuit measures and controls the temperature of sample and reference holders to conform to a Organisation and Qualities... [Pg.86]

To understand the impact of a CMP process on a certain product with a unique integrated circuit pattern, it is desirable to measure areas with different feature sizes and shapes. Since CMP polish rate may be affected by pattern density, areas encompassing various features should be included in the measurement program. The within-die thickness nonuniformity will indicate the planarization capability of a CMP process. [Pg.224]

Fig. 19 (a) The device schematic for a pseudo Y-junction transistor, (b) SEM micrograph of the overall circuit arrangement used in the measurement of the electrical characteristics, with Au contact pads and an FIB-patterned Pt wire contacting the Au pads and the Y-junction. (c) The ambipolar I-V curves resemble that of an n-type semiconductor at a positive gate potential, and a p-type semiconductor at a negative gate potential top), and the equivalent circuit for a pseudo Y-junction SWNT device bottom). (Reprinted with permission from [170, 171])... [Pg.148]


See other pages where In-circuit measurement is mentioned: [Pg.131]    [Pg.116]    [Pg.116]    [Pg.131]    [Pg.116]    [Pg.116]    [Pg.1122]    [Pg.202]    [Pg.67]    [Pg.765]    [Pg.475]    [Pg.491]    [Pg.663]    [Pg.529]    [Pg.60]    [Pg.1151]    [Pg.241]    [Pg.251]    [Pg.1018]    [Pg.1019]    [Pg.565]    [Pg.101]    [Pg.591]    [Pg.627]    [Pg.168]    [Pg.216]    [Pg.373]    [Pg.296]    [Pg.781]    [Pg.180]    [Pg.362]    [Pg.479]    [Pg.523]    [Pg.324]    [Pg.117]    [Pg.448]    [Pg.19]    [Pg.427]   


SEARCH



© 2024 chempedia.info