Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Differential pressure transducer capillary

The on-line viscosimeters currently available are adaptations of the classical dilute solution capillary viscosimeters. They work on the principle of measuring the pressure drop across a capillary with a differential pressure transducer. The pressure drop can be related to the reduced or inherent viscosity of the sample via Poiseuille s law.84 Intrinsic viscosity is determined using the equation ... [Pg.350]

Earlier experiments involved the collection of SEC effluent aliquots to measure solution viscosity in batches with the very time consuming Ubbelohde drop-time type viscometers. A continuous capillary type viscometer was first proposed for SEC by Ouano. Basically, as shown in Figure 1, a single capillary tube with a differential pressure transducer was used to monitor the viscosity of SEC effluent at the exit of the SEC column. As liquid continuously flows through the capillary (but not through the pressure transducer), the detected pressure drop (AP) across the capillary provides the measure for the fluid viscosity (h) according to the Poiseuille s viscosity law ... [Pg.82]

Details of the SEC/Viscometer detector system have been described previously.(16) The key component of the viscometer detector is a differential pressure transducer (CELESCO Model P-7D, Canoga Park, CA) with a jf25 psi pressure range. The transducer monitors the pressure drop across a section of stainless steel capillary tubing (length 2 ft., I.D. = 0.007 in.). Pump pressure fluctuations... [Pg.132]

A better approach is the use of two capillaries (Cl and C2) in series, each of which is connected to a differential pressure transducer (DPI and DP2), and a sufficiently large holdup reservoir (H) in between. With this approach, the sample viscosity q is measured from the pressure drop across the first capillary, and the solvent viscosity q0 from the pressure drop across the second capillary. Pulsations are eliminated in this setup because they appear in both transducers simultaneously. Another design is the differential viscometer, in which four capillaries are arranged similar to a Wheatstone bridge. In Fig. 10, both designs are shown schematically. [Pg.17]

Each term on the right hand-side of Eq. (2.9) is measured independently providing a direct determination of n, also under low pressures [30]. It is necessary as well to eliminate the fluctuations in the atmospheric pressure and to measure more precisely the differential pressure pg - p, (Fig. 2.12). This is done with a sensitive draft-range differential pressure transducer from Omega (Model PX750-06DI) with an accuracy of 0.5 Pa. The capillary radius r is determined by a Kondon micrometer calliper while the surface tension of the surfactant solution - by the Wilhelmy-plate technique, and the height hc in the capillary tube - with a Wild cathetometer (Model KM-274). [Pg.60]

Eluent from the SEC separation passes through a capillary or capillaries (one, two and four capillary designs are available) and the pressure drop is monitored by a differential pressure transducer. From the pressure drop and the concentration of the sample (obtained from the DRI) the intrinsic viscosity (ri) can be calculated. Molecular weight averages, Mark-Houwink (for polymer branching studies) and molecular weight distribution plots can then be obtained after suitable data treatment. [Pg.198]

The single-capillary viscometer (SCV) is represented in Fig. la. Its design is a direct extrapolation of classical viscometry measurement. It is composed of a small capillary, through which the solvent flows at a constant flow rate, and a differential pressure transducer (DPT), which measures the pressure drop across the capillary. SCV obeys PoiseuiUe s law and the pressure drop AP across the capillary depends on the geometry of the capillary, on flow rate Q, and on viscosity of the fluid 7j according to... [Pg.1714]

The last differential viscometer design is the Waters Corporation detector [9], which is in the Alliance GPCV2000 high-temperature instrument. It is composed of three capillaries, two differential pressure transducers, and two holdup reservoirs it is represented in Fig. Id. The pressure transducers are connected flow-through this eliminates the need for frequent purges. This detector provides, at the same time, relative viscosity information and relative flow information. This design does not require a perfect matching of the capillaries. [Pg.1715]

The experimental apparatus required to perform UK experiments is relatively simple (Figure 1). The two isotopic feeds are routed via a four-way valve to either a vent or the reactor. Their flow rates are set as closely as possible using standard mass flow controllers. The pressures of the two gas streams are then equalised using a differential pressure transducer and fine needle valve at the vent outlet. The reactor effluent is monitored continuously by a quadmpole mass spectrometer (Fisons Instruments GasLab 300) via a capillary sampling system. [Pg.381]

Fig. 5.18 Principle of a growing drop experiment according to Passerone et al. (1991) S - motor driven syringe, C - capillary, DPT - differential pressure transducer, LI and L2 - the two liquids... Fig. 5.18 Principle of a growing drop experiment according to Passerone et al. (1991) S - motor driven syringe, C - capillary, DPT - differential pressure transducer, LI and L2 - the two liquids...
The automatic relative viscometer is ideally suited for measuring dilute polymer viscosities. It provides faster analysis and greater precision than is obtainable with conventional glass tube viscometers (Ubbelohde or Cannon-Fenske), which it replaces. The principle of operation is based on measurement of pressure drops due to the continuous forced flow of solvent and sample through two stainless steel capillary tubes placed in series. The pressure drop across each capillary tube obeys Poiseuille s law. The pressme drop is measured by a differential pressure transducer. The sample solution is loaded into a sample loop via a syringe pump and then pushed into one of the two capillaries. A steady-state condition is reached when the sample solution completely fills capillary 2, solvent remaining in capillary 1 at all times. The relative viscosity of the sample solution is determined simply and directly by the ratio of the pressure drops. From the measured relative viscosity, all other solution viscosity measurements can be calculated. Solution viscosities are determined by the viscosity of the sample relative to the reference solvent. The relative viscometer measures the solvent and sample viscosity simultaneously, so errors due to temperatme fluctuation and solvent variations are avoided. The main advantages of this approach are ... [Pg.29]

Another automatic capillary viscometer which directly detects the pressure difference AP (eqn (5.42)) has recently come on the market. The Relative Viscometer manufactured by the Viscotek Co. includes two capillaries (capillary 1 and 2) connected in series, with the sample injection valve located between the two capillaries. Differential pressure transducers (DPT) are connected in parallel to each capillary. The sample is injected into capillary 2 and a pressure change is detected by the DPT. The viscosity ratio v r is determined as the ratio of the pressures divided by the instrument constant. [Pg.140]

Single capillary viscometers are remarkably easy and inexpensive to construct using a modem differential pressure transducer with a double T-network and capillary of given L and/ . [Pg.236]

Another device that simplifies the measurement of dilute solution viscosity is the differential viscometer. One design [31] consists of two capillary tubes in series (Figure 7.9). When fluid is pumped at a constant rate through the two tubes, the pressure drops are sensed by differential pressure transducers, DPT-1 and DPT-2. With... [Pg.282]

Sample solution is introduced into the solvent flow stream from a sample loop via a sample injection valve. The solvent flow pushes the sample solution through the analytical capillary vdiere the viscosity of the sample is detected. The AP signal from each pressure transducer is fed to a differential logarithmic amplifier as shown in the Figure. The viscometer output is the In signal of the sample solution. A pump is used to... [Pg.86]


See other pages where Differential pressure transducer capillary is mentioned: [Pg.16]    [Pg.131]    [Pg.16]    [Pg.131]    [Pg.504]    [Pg.55]    [Pg.17]    [Pg.377]    [Pg.236]    [Pg.166]    [Pg.21]    [Pg.3108]    [Pg.105]    [Pg.30]    [Pg.470]    [Pg.71]    [Pg.137]    [Pg.1916]    [Pg.421]    [Pg.43]    [Pg.151]    [Pg.210]    [Pg.80]    [Pg.84]    [Pg.105]    [Pg.218]    [Pg.72]    [Pg.806]    [Pg.35]    [Pg.43]    [Pg.18]    [Pg.249]   


SEARCH



Capillary pressure

Capillary transducers

Differential pressure transducer

Differential pressure transducer capillary viscometer

Transducer, transducers

© 2024 chempedia.info