Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Differential column

Deactivating catalyst 319 Dead zones 159, 162, 163 Degree of segregation 471 Density influences 492 Desorption of solute 578, 579 Difference differential equation 579 Difference formulae for partial differential equations 268 Differential column 167... [Pg.693]

For a system with n components (including nonad-sorbable inert species) there are n — 1 differential mass balance equations of type (17) and n — 1 rate equations [Eq. (18)]. The solution to this set of equations is a set of n — 1 concentration fronts or mass transfer zones separated by plateau regions and with each mass transfer zone propagating through the column at its characteristic velocity as determined by the equilibrium relationship. In addition, if the system is nonisothermal, there will be the differential column heat balance and the particle heat balance equations, which are coupled to the adsorption rate equation through the temperature dependence of the rate and equilibrium constants. The solution for a nonisothermal system will therefore contain an additional mass transfer zone traveling with the characteristic velocity of the temperature front, which is determined by the heat capacities of adsorbent and fluid and the heat of adsorption. A nonisothermal or adiabatic system with n components will therefore have n transitions or mass transfer zones and as such can be considered formally similar to an (n + 1)-component isothermal system. [Pg.39]

Control system. electronic control for oil bath according to temperature or differential column pressure — electronic control or follow-up control for heating jacket — reflux control with vapour division — vacuum control — fraction collector control — recording of all data — protection against superheating and interruption of cooling water supply... [Pg.354]

Differential columns True countercurrent flaw ofgas and liquid. [Pg.7]

Solution The first step is to write a mass balance on the water in a differential column height Az ... [Pg.253]

As in Eq. 15.4-4, the left-hand side of this relation represents the accumulation of the solute within the solution in a differential column volume. The first term on the right-hand side represents solute convection in minus out and is also common with the earher analysis. The second term on this side represents material lost from the solution into the adsorbent. The overall mass transfer coefficient k may include not only diffusion from the bulk solution to the adsorbent but both diffusion and reaction within the adsorbent. The surface area per volume a usually is taken to represent external adsorbent area per volume of bed however, if the mass transfer is dominated by reaction within the particle, this area can be more conveniently defined as the actual surface area of the microporous adsorbent per volume of bed. The concentration y is as usual that concentration that would exist in the liquid if it were at equilibrium with the solid. It is this concentration that includes the nonlinear isotherms. [Pg.444]

At first we tried to explain the phenomenon on the base of the existence of the difference between the saturated vapor pressures above two menisci in dead-end capillary [12]. It results in the evaporation of a liquid from the meniscus of smaller curvature ( classical capillary imbibition) and the condensation of its vapor upon the meniscus of larger curvature originally existed due to capillary condensation. We worked out the mathematical description of both gas-vapor diffusion and evaporation-condensation processes in cone s channel. Solving the system of differential equations for evaporation-condensation processes, we ve derived the formula for the dependence of top s (or inner) liquid column growth on time. But the calculated curves for the kinetics of inner column s length are 1-2 orders of magnitude smaller than the experimental ones [12]. [Pg.616]

In many applications in mass spectrometry (MS), the sample to be analyzed is present as a solution in a solvent, such as methanol or acetonitrile, or an aqueous one, as with body fluids. The solution may be an effluent from a liquid chromatography (LC) column. In any case, a solution flows into the front end of a mass spectrometer, but before it can provide a mass spectrum, the bulk of the solvent must be removed without losing the sample (solute). If the solvent is not removed, then its vaporization as it enters the ion source would produce a large increase in pressure and stop the spectrometer from working. At the same time that the solvent is removed, the dissolved sample must be retained so that its mass spectrum can be measured. There are several means of effecting this differentiation between carrier solvent and the solute of interest, and thermospray is just one of them. Plasmaspray is a variant of thermospray in which the basic method of solvent removal is the same, but the number of ions obtained is enhanced (see below). [Pg.71]

Rota.ryAgita.ted Columns. Because of the mechanical advantages of rotary agitation, most modem differential contactors employ this method. The best known of the commercial rotary agitated contactors are shown in Eigure 15. Eeatures and appHcations of these columns ate given in Table 3. [Pg.75]

The first differential centrifugal extractor to be used in industry was the PodbieHiiak extractor which was introduced in the 1950s (209,210) and can be regarded as a perforated-plate column wrapped around a rotor shaft. Rotation creates a centrifugal force which results in a great reduction in the equivalent height and contact time that would be needed in a conventional perforated-plate column. [Pg.77]

Plutonium solutions that have a low activity (<3.7 x 10 Bq (1 mCi) or 10 mg of Pu) and that do not produce aerosols can be handled safely by a trained radiochemist in a laboratory fume hood with face velocity 125—150 linear feet per minute (38—45 m/min). Larger amounts of solutions, solutions that may produce aerosols, and plutonium compounds that are not air-sensitive are handled in glove boxes that ate maintained at a slight negative pressure, ca 0.1 kPa (0.001 atm, more precisely measured as 1.0—1.2 cm (0.35—0.50 in.) differential pressure on a water column) with respect to the surrounding laboratory pressure (176,179—181). This air is exhausted through high efficiency particulate (HEPA) filters. [Pg.204]

A vapor feed is favored when the stream leaves the upstream unit as a vapor or when most of the column feed leaves the tower as overhead product. The use of a vapor feed was a key component in the high efficiency cited previously for the spHtter, where most of the feed goes overhead. Low Column Pressure Drop. The penalty for column pressure drop is an increase in temperature differential ... [Pg.85]

Generally, q is small because the outside area is not large in comparison to the amount of heat being transferred, and the energy balance can be simplified. In these conditions it is also convenient to write balances over a differential section of the column. These balances yield the following ... [Pg.100]


See other pages where Differential column is mentioned: [Pg.167]    [Pg.694]    [Pg.694]    [Pg.130]    [Pg.568]    [Pg.159]    [Pg.542]    [Pg.141]    [Pg.141]    [Pg.403]    [Pg.6]    [Pg.167]    [Pg.694]    [Pg.694]    [Pg.130]    [Pg.568]    [Pg.159]    [Pg.542]    [Pg.141]    [Pg.141]    [Pg.403]    [Pg.6]    [Pg.57]    [Pg.60]    [Pg.36]    [Pg.55]    [Pg.48]    [Pg.57]    [Pg.134]    [Pg.260]    [Pg.263]    [Pg.503]    [Pg.67]    [Pg.68]    [Pg.75]    [Pg.11]    [Pg.201]    [Pg.207]    [Pg.242]    [Pg.429]    [Pg.204]    [Pg.431]    [Pg.20]    [Pg.20]    [Pg.496]    [Pg.41]    [Pg.106]   
See also in sourсe #XX -- [ Pg.130 ]

See also in sourсe #XX -- [ Pg.159 ]




SEARCH



© 2024 chempedia.info