Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Determination typical analytical values

In a typical NARL study there are two or more independent measurements of the analyte concentration, each with an associated statement of traceability and estimate of uncertainty. In setting the assigned value, any significant differences between the independent measures must be considered and where possible the causes identified. If these differences are too great, it may not be possible to determine an assigned value for an analyte, in which case a consensus value or indicative value may be used. The rationale for determining the assigned value is always described in the study report. [Pg.117]

For a selection of materials and analytes, values were determined at LGC using high-accuracy IDMS methodology that had been the subject of international comparisons through the CCQM. Some typical results obtained are shown in Table 2. [Pg.179]

When compared to a typical potency assay, this validation parameter is the most variable Eq. (15.4) represents the typical calculation that is used to determine the analytical recovery for the method. The spiked amount is based upon the predetermined acceptance limit (see Section 15.3 for different methods of calculating this value). [Pg.359]

Typically in radioanalytical chemistry, the null hypothesis is the hypothesis that no analyte is in the sample. Even if no analyte is present, the net result of the measurement has uncertainty, and, if the measurement were repeated a number of times, a distribution of results about zero, including both positive and negative values, should be observed. Although results near zero are most likely, in principle there is no upper or lower bound for what the result might be. Observation of a positive result in a single measurement does not necessarily constitute strong evidence that the analyte is present. The result must exceed some positive threshold value, called the critical value, to lead one to conclude that the analyte is really present. The question is how to determine the critical value ... [Pg.205]

The resolving power of this instrument in purified nitrogen was typically 20 to 50 and could reach 75 under optimal conditions. A related work emphasized analytical values showing that PAHs as adsorbates on borosilicate glass could be determined (Figure 5.9) at 40-pg levels or 5.5 to 7 pg/mm using LDI-IMS in air at ambient pressure and 100°C for ion characterization in the drift tube. ... [Pg.112]

The terms specificity and selectivity are often used interchangeably. A method is said to be specific if it provides a response for only a single analyte. If the response in question is distinguished from all other responses, the method is said to be selective. The International Conference on Harmonization (ICH)-2 does not differentiate both terms and defines specificity or selectivity as the ability to unambiguously determine the analyte in the presence of other components whose presence is to be expected. This includes typical impurities, decomposition products, and matrix components.The specificity of the method for TLC or high-performance TLC (HPTLC) analyses was proved by identification and purity checks of the analyte spots. This can be done by measuring in sim the ultraviolet and visible spectroscopy (UV-Vis) spectra of the ana-lyte(s) and the authentic reference standard(s), those eluted on the same plate, and then by calculating their correlations (the r value should be >0.999). This correlation should be calculated on the upslope, the apex, and the downslope of the peaks. In a quality control laboratory, the selectivity... [Pg.2336]

Blank values. Ihe overall blank values of the analytical method, from weighing of the specimen to the prepeuation of the sample, are mainly caused by the added reagents, Le., the buffer solution, acids and bases, the dithiodicarbamate and, last but not least, the ultra-pure water if used instead of seawater for dilution or for the determination of the blank values. If no water of sufficient purity (ultra-pure water) is available, seawater extracted with Na-DBDTC is an accetable substitute. The use of reference seawater is another option (c.g., NASS and CASS reference material see Section 12.1.6). The blanks are derived firom the difference between the certified and the measmed values and should result in the same (similar) blank values from different reference materials. Typical blank values for Fe have been found to be between 100 and 150ng/L, for Zn between 60 and 90 ng/L, and for Cd and Pb below 10 ng/L. Variations depend on the stocks of chemicals employed. In particular the... [Pg.338]

Analytical measurements and certifications often contain a statement of traceability. Traceability describes the result or measurement whereby it can be related to appropriate standards, generally international or national standards, through an unbroken chain of comparisons (Ref. 9). Traceability typically includes the application of a reference material (RM) or a standard reference material (SRM) for instrument calibration before standardization for the analytes of interest. The true value of a measured quantity (t) cannot typically be determined. The true value is defined as characterizing a quantity that is perfectly defined. It is an ideal value which could be arrived at only if all causes of measurement uncertainty were eliminated, and the entire population was sampled. [Pg.1423]

The relative measurement error in concentration, therefore, is determined by the magnitude of the error in measuring the cell s potential and by the charge of the analyte. Representative values are shown in Table 11.7 for ions with charges of+1 and +2, at a temperature of 25 °C. Accuracies of 1-5% for monovalent ions and 2-10% for divalent ions are typical. Although equation 11.22 was developed for membrane electrodes, it also applies to metallic electrodes of the first and second kind when z is replaced by n. [Pg.495]

The limit of detection is the smallest amount of an analyte that is required for reliable determination, identification or quantitation. More mathematically, it may be defined as that amount of analyte which produces a signal greater than the standard deviation of the background noise by a defined factor. Strictly for quantitative purposes, this should be referred to as the limit of determination . The factor used depends upon the task being carried out and for quantitative purposes a higher value is used than for identification. Typical values are 3 for identification and 5 or 10 for quantitation. [Pg.26]

Quantitation of anthocyanins has become simple and fast since many anthocy-anin standards became commercially available as external standards in the past decade. When the standards are not available, individual anthocyanins or total monomeric anthocyanins can be determined by the use of a generic external standard such as commercial cyanidin-3-glucoside or other compound structurally similar to the analytes of interest. Individual and total peak areas are measured at 520 nm or their and quantified using external standards by which values are typically slightly different from those via the pH differential method. ... [Pg.486]

Applications Quantitative dry ashing (typically at 800 °C to 1200°C for at least 8h), followed by acid dissolution and subsequent measurement of metals in an aqueous solution, is often a difficult task, as such treatment frequently results in loss of analyte (e.g. in the cases of Cd, Zn and P because of their volatility). Nagourney and Madan [20] have compared the ashing/acid dissolution and direct organic solubilisation procedures for stabiliser analysis for the determination of phosphorous in tri-(2,4-di-t-butylphenyl)phosphite. Dry ashing is of limited value for polymer analysis. Crompton [21] has reported the analysis of Li, Na, V and Cu in polyolefins. Similarly, for the determination of A1 and V catalyst residues in polyalkenes and polyalkene copolymers, the sample was ignited and the ash dissolved in acids V5+ was determined photo-absorptiometrically and Al3+ by complexometric titration [22]. [Pg.594]


See other pages where Determination typical analytical values is mentioned: [Pg.86]    [Pg.554]    [Pg.44]    [Pg.261]    [Pg.2178]    [Pg.148]    [Pg.282]    [Pg.578]    [Pg.3589]    [Pg.1078]    [Pg.35]    [Pg.605]    [Pg.404]    [Pg.58]    [Pg.379]    [Pg.167]    [Pg.141]    [Pg.175]    [Pg.57]    [Pg.61]    [Pg.91]    [Pg.92]    [Pg.92]    [Pg.95]    [Pg.101]    [Pg.850]    [Pg.359]    [Pg.368]    [Pg.31]    [Pg.6]    [Pg.519]    [Pg.55]    [Pg.481]    [Pg.142]    [Pg.308]    [Pg.133]   
See also in sourсe #XX -- [ Pg.38 , Pg.47 ]




SEARCH



Analytical determinability

Analytical determinations

Analytical value

Typical values

© 2024 chempedia.info