Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Hardware detector

Because of recent advances ia hardware, particularly detector hardware, and computer software, x-ray instmments have become very powerful. Problems that could not be solved several years ago can be solved with the newer iastmmentation. Also the instmments have become much more automatic so that the mote routine problems can be solved much faster than a few years ago. [Pg.371]

This delimits three aspects of FFC signal detection (detector hardware, excitation detection method, and data-reduction algorithm). The following paragraphs explain briefly the most popular choices we have made to handle these aspects. [Pg.454]

Owing to the lack of commercially available standards (there are only seven standards and there are 77 known MCs Table 40.3), it is virtually impossible to identify all toxins using the aforementioned methods. The use of PDA detection has undoubtedly added a degree of specificity to the method by virtue of the characteristic UV spectra obtained for MCs [124]. It is not, however, possible to differentiate between individual MCs on the basis of their UV spectra alone. While recent developments in detector hardware, combined with increased sophistication of spectral matching software, may assist in the identification of MCs by spectral match data in conjunction with retention time, the lack of the standards makes the job of creating a thorough spectral library virtually impossible [21]. [Pg.867]

Category III noise Is reduced by strategies combining optimization of both detector and non-detector hardware as In reduction of reciprocating pump pulsation, the use of low thermal coefficient, matched pairs of photodiodes for reference and sample beam optical trains, and thermal coupling of reference and sample photodiodes. Pump pulsation noise can also be reduced by proper flow cell design ( ). [Pg.109]

In order to prepare the system for 3D-CT, it is not enough to integrate a second detector array. Besides this special attention has to be paid to the computer hardware, the synchronisation between object movement and the data read out as well as to the collimator of the LINAC. The collimator has been built with 4 tungsten blocks which can be moved individually m order to shape different sht sizes for 2D-CT as well as different cone angles for 3D-CT or digital radiography. [Pg.586]

Powder diffraction studies with neutrons are perfonned both at nuclear reactors and at spallation sources. In both cases a cylindrical sample is observed by multiple detectors or, in some cases, by a curved, position-sensitive detector. In a powder diffractometer at a reactor, collimators and detectors at many different 20 angles are scaimed over small angular ranges to fill in the pattern. At a spallation source, pulses of neutrons of different wavelengdis strike the sample at different times and detectors at different angles see the entire powder pattern, also at different times. These slightly displaced patterns are then time focused , either by electronic hardware or by software in the subsequent data analysis. [Pg.1382]

Xps ndAes Instrumentation. The instmmentation required to perform xps and aes analyses is generally sophisticated and expensive (19). The need for UHV conditions in order to retain surface cleanliness for a tractable period of time was mentioned above. Beyond this requirement (and the hardware that accompanies it), the most important components of an electron spectrometer system are the source, the electron energy analyzer, and the electron detector. These will be discussed in turn below. [Pg.282]

Volume of vessel (free volume V) Shape of vessel (area and aspect ratio) Type of dust cloud distribution (ISO method/pneumatic-loading method) Dust explosihility characteristics Maximum explosion overpressure P ax Maximum explosion constant K ax Minimum ignition temperature MIT Type of explosion suppressant and its suppression efficiency Type of HRD suppressors number and free volume of HRD suppressors and the outlet diameter and valve opening time Suppressant charge and propelling agent pressure Fittings elbow and/or stub pipe and type of nozzle Type of explosion detector(s) dynamic or threshold pressure, UV or IR radiation, effective system activation overpressure Hardware deployment location of HRD suppressor(s) on vessel... [Pg.2330]

The most important hardware items appeared to be the detectors themselves. The gas detection system gave frequent spurious alarms, and on both platforms the ultraviolet (UV) fire detectors were also prone to spurious activation from distant hot work for example, and had a limited ability to detect real fires. The tmreliability of these systems had a general effect on response time and would, overall, lengthen the time to respond. The second aspect which was related to hardware was fimction and performance testing of the emergency blowdown systems. It is critical that the workers believe the systems will work when required, and this can only be achieved by occasional use or at least fimction testing. [Pg.339]

However, it is not the simultaneous use of all detector functions that makes this detector so useful. The real advantage of the trifunctional detector is that it allows the analyst a choice of the three most useful detector functions in one detecting system. Furthermore, any of the three functions can be chosen at the touch of a switch and without any changes in hardware. [Pg.190]

Fig. 2.6.5 Hardware for high field NMR remote probe in (c) contains a relatively large saddle-detection. Photographs (a) and (b) show la- coil and is used for (flow) imaging. The detec-boratory-built remote detection probes with tor probe in (d) contains a microsolenoid coil both rf coils built into the same body (c), (d) for optimized mass sensitivity, which is parti-and (e) are detector-only remote probes that cularly useful for microfluidic NMR applica-can be inserted from the top or bottom into the tions. The same probe is shown in (e) with a NMR imaging assembly, so that the well mounted holder for a microfluidic chip that is... Fig. 2.6.5 Hardware for high field NMR remote probe in (c) contains a relatively large saddle-detection. Photographs (a) and (b) show la- coil and is used for (flow) imaging. The detec-boratory-built remote detection probes with tor probe in (d) contains a microsolenoid coil both rf coils built into the same body (c), (d) for optimized mass sensitivity, which is parti-and (e) are detector-only remote probes that cularly useful for microfluidic NMR applica-can be inserted from the top or bottom into the tions. The same probe is shown in (e) with a NMR imaging assembly, so that the well mounted holder for a microfluidic chip that is...
Some analytical methods are highly mature (NAA, XRD, XRF, XAS), the theory is well assessed, and just instrumental and incremental improvements (more intense sources, better detectors) may be expected. However, in many other areas the sharply increasing power of analytical instrumentation (with regard to both hardware and software) and its transformation into tools for in-process control (such as NIRS, LR-NMR, etc.) are most appropriately considered as breakthroughs. [Pg.728]

Topics which will be presented in this chapter include the hardware, software, automation, valve and column configurations, and integration used in comprehensive 2DLC. Aspects of the 2DLC experiment in conjunction with multichannel detectors such as UV diode array optical detectors and mass spectrometers are discussed along with the handling of the data, which is expected to expand in scope in the future as chemometric methods are more widely used for data analysis. [Pg.97]

Since NIR was developed by the USDA for food products, the first (commercial) mode developed was diffuse reflection. The earliest work was performed on instruments which had, in essence, NIR as an extra. The UV/Vis hardware (e.g., Cary model 10) had an additional detector and could be used through the NIR. This one fact explains why so much literature uses nanometers for units instead of wave numbers. [Pg.171]

Buffers are used mainly to control the pH and the acid-base equilibrium of the solute in the mobile phase. They can also be used to influence the retention times of ionizable compounds. The buffer capacity should be maximum and should be uniform in the pH range of 2-8 commonly used in HPLC. The buffers should be soluble, stable, and compatible with the detector employed, e.g., citrates are known to react with certain HPLC hardware components. [Pg.556]


See other pages where Hardware detector is mentioned: [Pg.76]    [Pg.257]    [Pg.102]    [Pg.227]    [Pg.6]    [Pg.60]    [Pg.142]    [Pg.522]    [Pg.484]    [Pg.76]    [Pg.257]    [Pg.102]    [Pg.227]    [Pg.6]    [Pg.60]    [Pg.142]    [Pg.522]    [Pg.484]    [Pg.66]    [Pg.398]    [Pg.111]    [Pg.216]    [Pg.2327]    [Pg.90]    [Pg.243]    [Pg.113]    [Pg.411]    [Pg.364]    [Pg.828]    [Pg.262]    [Pg.83]    [Pg.379]    [Pg.396]    [Pg.443]    [Pg.83]    [Pg.101]    [Pg.299]    [Pg.86]    [Pg.733]    [Pg.78]    [Pg.356]    [Pg.186]   
See also in sourсe #XX -- [ Pg.65 , Pg.70 , Pg.75 ]




SEARCH



Hardware

© 2024 chempedia.info