Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Vacuum valve control

CHC/CCH4 AND GN2 PRESSURANT LINE MAIN VENT LINE - BACK PRESSURE CONTROL VALVE VACUUM LINE FOR CFM TEST TANK. 10 NEW EJECTOR... [Pg.170]

Another example of pressure control by variable heat transfer coefficient is a vacuum condenser. The vacuum system pulls the inerts out through a vent. The control valve between the condenser and vacuum system varies the amount of inerts leaving the condenser. If the pressure gets too high, the control valve opens to pull out more inerts and produce a smaller tube area blanketed by inerts. Since relatively stagnant inerts have poorer heat transfer than condensing vapors, additional inerts... [Pg.66]

Put the vacuum system control valves at the highest point of a horizontal run and the control valve bypass in the same horizontal plane. This is in compliance with item 8. [Pg.199]

A vacuum condenser has vacuum equipment (such as steam jets) pulling the noncondensibles out of the cold end of the unit. A system handling flammable substances has a control valve between the condenser and Jets (an air bleed is used to control nonflammable systems). The control method involves derating part of the tube surface by blajiketing it with noncondensibles that exhibit poor... [Pg.291]

The control valve allows the Jets to pull noncondensibles out of the condenser as needed for system pressure control. In addition to requiring extra surface area for control, the vacuum condenser also needs enough surface area for subcooling to ensure that the Jets do not pull valuable hydrocarbons or other materials out with the noncondensibles. To allow proper control and subcooling, some designers add approximately 50% to the calculated length. [Pg.291]

Total flow vacuum pump Exhaust control with bypass controller valve... [Pg.1291]

Connect a vacuum manifold to a vacuum source not to exceed 20 in Hg, and place a 1-L waste liquid trap between the manifold and the vacuum source. To use the manifold, remove the cover, and place a Luer hub solvent guide needle on the male Luer fitting of each flow control valve. Remove the collecting vessel rack, and replace the cover. Close all control valves on the manifold cover. [Pg.366]

Add 0.5-0.75 in of anhydrous sodium sulfate to the head of a 6-mL silica disposable column to ensure removal of any residual water, and attach a 15-mL reservoir to the top of each column. Place the desired number of silica cleanup columns into the female Luer receptacles on the cover. Turn the vacuum on at the source, and set the vacuum to about 10 inHg. Wash each column with isooctane. If the column goes dry, add an additional 10 mL of isooctane. When the solvent in the column reaches the top of the packing bed, turn the flow control valve fully off. [Pg.366]

Figure 1. Diagram of apparatus (M) monomer reservoir (F) flow meter (VG) vacuum gage (mercury manometer) (E) electrode (T) liquid nitrogen trap (P) mechanical pump (V,) needle valve (Vt) stop valve (Vs) pressure control valve (OSC) discharge frequency oscillator (AMP) amplifier (1MC) impedance matching circuit... Figure 1. Diagram of apparatus (M) monomer reservoir (F) flow meter (VG) vacuum gage (mercury manometer) (E) electrode (T) liquid nitrogen trap (P) mechanical pump (V,) needle valve (Vt) stop valve (Vs) pressure control valve (OSC) discharge frequency oscillator (AMP) amplifier (1MC) impedance matching circuit...
DC Diaphragm controller P Vacuum pump M Measuring and switching device PS Pressure sensor V1 Pump valve V2 Gas inlet valve TH Throttle... [Pg.92]

Figure 3.5. Vacuum control with steam jet ejectors and with mechanical vacuum pumps, (a) Air bleed on PC. The steam and water rates are hand set. The air bleed can be made as small as desired. This can be used only if air is not harmful to the process. Air bleed also can be used with mechanical vacuum pumps, (b) Both the steam and water supplies are on automatic control. This achieves the minimum cost of utilities, but the valves and controls are relatively expensive, (c) Throttling of process gas flow. The valve is larger and more expensive even than the vapor valve of case (a). Butterfly valves are suitable. This method also is suitable with mechanical vacuum pumps, (d) No direct pressure control. Settings of manual control valves for the utilities with guidance from pressure indicator PI. Commonly used where the greatest vacuum attainable with the existing equipment is desired. Figure 3.5. Vacuum control with steam jet ejectors and with mechanical vacuum pumps, (a) Air bleed on PC. The steam and water rates are hand set. The air bleed can be made as small as desired. This can be used only if air is not harmful to the process. Air bleed also can be used with mechanical vacuum pumps, (b) Both the steam and water supplies are on automatic control. This achieves the minimum cost of utilities, but the valves and controls are relatively expensive, (c) Throttling of process gas flow. The valve is larger and more expensive even than the vapor valve of case (a). Butterfly valves are suitable. This method also is suitable with mechanical vacuum pumps, (d) No direct pressure control. Settings of manual control valves for the utilities with guidance from pressure indicator PI. Commonly used where the greatest vacuum attainable with the existing equipment is desired.
If gas is to be delivered to a reaction flask which has an unobstructed outlet, a simple flow control valve on the high-pressure cylinder will provide adequate regulation of the gas delivery. In this case a needle valve is attached to the cylinder, or to a pressure regulator which in turn is attached to the cylinder. It also is possible to deliver gas to a closed system, such as a vacuum line, with a flow control valve. In this case the pressure within the apparatus must be carefully monitored by means of a manometer and the system should also be equipped with a means of pressure relief, such as a mercury bubbler manometer (Fig. 7.2). [Pg.280]

The mass spectra of the gases evolved from the deuterated SWNT sample heated in vacuum were measured with the MI 1201V mass spectrometer. Gas ionization in the ion source of the spectrometer was produced with a 70-eV electron beam. To obtain the gas phase, the sample was placed in a quartz ampoule of a pyrolyzer that was connected to the injection system of the mass spectrometer through a fine control valve. Then the ampoule was evacuated to a pressure of about 2-x 10-5 Pa in order to remove the surface and weakly bound impurities from the sample. After the evacuation, the ampoule was isolated from the vacuum system and the sample was heated to 550°C in five steps. At each step, the sample was kept at a fixed temperature for 3 h then the fine control valve was open and the mass-spectrometric analysis of the gas collected in the ampoule was performed. After the analysis, the quartz ampoule was again evacuated, the valve was closed, and the sample was heated to the next temperature. The measurements were carried out over the range 1 < m/z < 90, where m is the atomic mass and z is the ion charge. The spectrometer resolution of about 0.08% ensured a reliable determination of the gas-phase components. [Pg.228]

This cycling can be eliminated by mounting the control valve in the condensate pipe, but this creates new problems, because when the load decreases, the process is slow steam has to condense before the condensate level is affected, and when the load increases, the process is fast, because blowing out liquid condensate is fast. With such "nonsymmetrical" process dynamics, control is bound to be poor. A better option is to use lifting traps to prevent condensate accumulation. These pumping traps will make temperature control possible even when the heater is under vacuum, but will not improve the problem of low rangeability, and the possible use of two control valves in parallel can still be necessary. [Pg.281]

The SPE tubes are used in conjunction with a Visiprep D-L vacuum chamber complete with disposable flow control valve liners. [Pg.266]

A high-capacity water-seal vacuum pump downstream of an automatically controlled valve provides for chamber evacuation and pressure... [Pg.184]

The vacuum line is isolated from the forepump, and a liquid-nitrogen bath is put around trap D. The fluorine metering system is partially evacuated with the water aspirator, as discussed above. Then the vacuum line is completely evacuated with the forepump. Fluorine is allowed to expand slowly into the metal system by means of the needle valve 3. As the pressure increases above atmospheric, some fluorine is allowed to bleed into the vacuum line and storage bulbs until a pressure of 650-700 mm. is reached. First needle valve 1 on the metal system and then the fluorine control valve 3 are quickly closed. (Some fluorine may escape from the blowout manometer during this operation.) Approximately a 10% excess of fluorine (0.023 mole in this case) is condensed into the metal pressure reactor containing the thionyl fluoride. The amount of fluorine used is measured by the pressure drop in a calibrated volume in the pressure range of approximately 700-400 mm., since fluorine has a vapor pressure of approximately 400 mm. at —196°. [Pg.135]

For a condenser operating at atmospheric pressure, an adequate vent is all that is necessary. However, air is often not suitable for contact with the process due to concern about contamination or flammability. In these cases, the vent may be connected to a source of low pressure nitrogen, or other inert gas. For vacuum operation, the vent must also be connected to a vacuum pump or steam jet (eductor), as shown in Figure 3.11(B). The pressure controller adjusts the split range control valves such that as its output decreases, first PV-2 closes then PV-1 opens. Normal operation would... [Pg.46]


See other pages where Vacuum valve control is mentioned: [Pg.869]    [Pg.218]    [Pg.316]    [Pg.869]    [Pg.252]    [Pg.411]    [Pg.93]    [Pg.109]    [Pg.76]    [Pg.869]    [Pg.120]    [Pg.42]    [Pg.103]    [Pg.105]    [Pg.107]    [Pg.96]    [Pg.348]    [Pg.266]    [Pg.96]    [Pg.287]    [Pg.175]    [Pg.184]    [Pg.42]    [Pg.42]    [Pg.42]    [Pg.78]   
See also in sourсe #XX -- [ Pg.57 , Pg.58 ]




SEARCH



Control valve

Vacuum control

Vacuum controller

© 2024 chempedia.info