Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Compensating control

Compensating control The process of au tomatically adjusting the control point of a controller to compensate for changes in a second measured variable. [Pg.1423]

Fig. 3. Bode diagram of the closed-loop under the estimation/compensation control approach. Closed-loop system is expected to be non-sensitive to high frequency signais. Arrows indicate how the frequency response of the ciosed-ioop as Kc is increased. Fig. 3. Bode diagram of the closed-loop under the estimation/compensation control approach. Closed-loop system is expected to be non-sensitive to high frequency signais. Arrows indicate how the frequency response of the ciosed-ioop as Kc is increased.
Here, the performance of the heat compensation controller is tested. To this end, the model parameters were taken as follows [16] x = 862.5 mol m. ... [Pg.47]

Compensation Control FL1 antibody BerEP-4 FITC antibody ... [Pg.330]

Such measurements require a pH meter and probe with temperature compensator control and calibration with standard buffers at the desired temperature. [Pg.318]

Acquire all single-stained compensation controls. If necessary, adjust the PMTV gains for each detector starting from those defined by CS T procedure. [Pg.450]

The unit consists of the reactor, housing the core with actuators of the scram and compensation control rod groups, two steam generators that are connected to the reactor vessel with short nozzles, and two hydraulic chambers accommodating the electric pumps. [Pg.289]

One peculiarity of electrical compensation should be mentioned, too every electronic temperature controller needs a minimal but nonzero temperature difference to react. Furthermore, the heat produced by the sample and that produced by the electrical heater are not at the same place, and there must be some temperature gradient to allow a heat exchange between them. Therefore, such calorimeters operate not in a strictly isothermal but in a quasi-isothermal mode. The temperature is generally kept constant but may differ locally. The deviations depend on the heat flow rate as well as on the quality of the compensation controller. Normally, these differences can be neglected in practice. [Pg.154]

Answer About five hours after startup, the production of additional xenon from decay of the increasing iodine concentration catches up with the rate of xenon decay plus burnout, and the xenon passes through 0 minimum or turnaround point. The compensating control rods therefore pass through 0 maximum insertion point at time of farnarcand. [Pg.50]

Control of Preceding Reactivity Changes (Transients) - Size and Timing of ansients Compensating Control Required. . . ... [Pg.64]

As shown in the basic circuit diagram, Figure 2.1, the temperature compensation control changes the output slope (mV/pH unit)... [Pg.23]

If the temperature compensation control is left at the standardization temperature and the sample is at a different temperature, the correct sample pH can be calculated with the formula... [Pg.24]

Manufacturers of pH meters may publish specifications such as those listed in Table 2.4. Some specifications may be misleading, because the conditions for which they are valid are not always obvious. For example, a relative accuracy specification of 0.07 pH may be more accurate than one stated as 0.03 pH within 2 pH units of standardization. In the first case, the manufacturer may include the entire meter scale linearity, while in the second case only a small portion of the scale is considered. A specification on a temperature compensating control is usually better between 10° and 60°C than it is outside this range since the extremes of this potentiometer are not as accurate. [Pg.33]

Also remember that the temperature compensator setting becomes more critical as the pH value deviates from 7. As discussed in Chapter 2, a meter is designed with zero millivolts at pH 7 and the temperature compensator control has no effect at this point. Since this point is the isopotential point of the electrode, any change in slope (mV/pH) due to temperature change has no effect. [Pg.102]

Then, set the display to read pH 6.00 with the standardization control. When the temperature compensating control is rotated to 0°C, the display should read about 5.9. When it is rotated to lOO C, it should read about 6.2. Reset the temperature control to 30°C and rotate the slope control. At 80% slope, the reading should be lower than pH 5.8. [Pg.140]

Ingimundarson, I., and T. Hagglund, Performance Comparison Between PID and Dead-Time Compensating Controllers, /. Process Control, 12, 887 (2002). [Pg.308]

Compensator Control element Controlled variable Controller... [Pg.254]


See other pages where Compensating control is mentioned: [Pg.64]    [Pg.279]    [Pg.280]    [Pg.45]    [Pg.26]    [Pg.350]    [Pg.2071]    [Pg.136]    [Pg.55]    [Pg.2059]    [Pg.282]    [Pg.1161]    [Pg.455]    [Pg.474]    [Pg.198]    [Pg.1189]    [Pg.173]    [Pg.359]    [Pg.228]    [Pg.17]    [Pg.27]    [Pg.37]    [Pg.135]    [Pg.217]   
See also in sourсe #XX -- [ Pg.1423 ]




SEARCH



© 2024 chempedia.info