Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Columns examples

Flow markers are often chosen to be chemically pure small molecules that can fully permeate the GPC packing and elute as a sharp peak at the total permeation volume (Vp) of the column. Examples of a few common flow markers reported in the literature for nonaqueous GPC include xylene, dioctyl phthalate, ethylbenzene, and sulfur. The flow marker must in no way perturb the chromatography of the analyte, either by coeluting with the analyte peak of interest or by influencing the retention of the analyte. In all cases it is essential that the flow marker experience no adsorption on the stationary phase of the column. The variability that occurs in a flow marker when it experiences differences in how it adsorbs to a column is more than sufficient to obscure the flow rate deviations that one is trying to monitor and correct for. [Pg.549]

In the distillation column example, the manipulated variables correspond to all the process parameters that affect its dynamic behavior and they are normally set by the operator, for example, reflux ratio, column pressure, feed rate, etc. These variables could be constant or time varying. In both cases however, it is assumed that their values are known precisely. [Pg.12]

Multivariate data are represented by one or several matrices. Variables (scalars, vectors, matrices) are written in italic characters scalars in lower or upper case (examples n, A), vectors in bold face lower case (example b). Vectors are always column vectors row vectors are written as transposed vectors (example bv). Matri ces are written in bold face upper case characters (example X). The first index of a matrix element denotes the row, the second the column. Examples x,- - or x(i, j) is an element of matrix X, located in row i and column / xj is the vector of row i xy is the vector of column j. [Pg.311]

The transpose (AT or A) of a matrix A is made by exchanging rows and columns. Examples ... [Pg.113]

Reducing the sample volume, however, means that the mass of sample loaded onto the narrow-bore column is reduced, and therefore the mass of sample reaching the detector is decreased. This imposes difficulties on the detection scheme because less sample is available for detection. Therefore, it is important that the entire instrument be optimized for use with small-bore columns. Examples of reduced sample volumes for each of the column categories are listed in Table 8.2. [Pg.246]

FIG URE 9.3 Overall mass balance on extraction column (Example 9.1). [Pg.386]

As a final note on the reactor/column example, we might have consid-... [Pg.155]

The steps we take to obtain an empirical formula from percent composition data are given in the left column (Steps) that follows. In the right column (Example), the empirical formula of a compound containing 39.2% phosphorus and 60.8% sulfur is calculated. [Pg.205]

Fig. 32. Chromatography of normal acid gastric juice on Amberlite IRC-50 column. Example of Co8 -vitamin Bjj binding of effluent obtained from column chromatographic separation of an individual gastric juice specimen. The vitamin Bjg binding is expressed as mug Co 0-vitamin Bjg bound/6 ml effluent. From Welsh et al. (Wll). Fig. 32. Chromatography of normal acid gastric juice on Amberlite IRC-50 column. Example of Co8 -vitamin Bjj binding of effluent obtained from column chromatographic separation of an individual gastric juice specimen. The vitamin Bjg binding is expressed as mug Co 0-vitamin Bjg bound/6 ml effluent. From Welsh et al. (Wll).
FIGURE 3.4 Performance of two-stage distillation column (Example 3.2). [Pg.153]

Reboiler Duty for a Single-Feed Column (Example 9.4)... [Pg.298]

FIGURE 9.7 Composition profiles in the ammonia-acetone-water column (Example 9.7). [Pg.309]

Figure 4.4 Graphical solution of enriching column, Example 4.2. Figure 4.4 Graphical solution of enriching column, Example 4.2.
To demonstrate some of the numerical characteristics of the 6 method for complex columns, Example 3-2 was selected. The statement of this example is given in Table 3-1 and the geometry of the column is depicted in Fig. 3-4. In addition to the distillate and bottoms, the column has one sidestream withdrawn. Thus, the 9 method has two 6 multipliers, 90 and 0U which are defined by the first two expressions given by Eq. (3-10). Since D and are specified (see Table 3-1) the two g functions to be used for computing these 0 s are given by Eqs. (3-21) and (3-22). [Pg.98]

A summary of the computer times and the number of trials required to solve a variety of examples by the methods enumerated above is presented in Table 4-17. These results show that for absorber-type columns (Examples 4-2, 4-3, and 4-4), procedure 1 requires fewer trials than do procedures 2 and 3. For columns which have a relatively small number of plates, all three procedures require about the same amount of computer time. As the number of plates and components are increased (see Example 4-12), the speed advantage of procedures 2 and 3 over procedure 1 becomes more pronounced. [Pg.172]

For distillation-type columns (Examples 4-5 through 4-7) the 0 method, which is presented in Chaps. 2 and 3, is seen to be from 5 to 20 times faster than the 2N Newton-Raphson method. [Pg.172]

A statement of the monoethanolamine example, Example 8-2, is given in Table 8-4. The solution is given in Table 8-5, and convergence characteristics of procedures 1 and 2 are given in Table 8-6. To demonstrate the use of procedures 1 and 2 in the solution of problems involving conventional distillation columns, Example 8-1 was solved. A statement of this example is... [Pg.291]

Capillary electrophoresis (CE) has emerged as an efficient and rapid separation technique in recent years. Its high efficiency has been employed in many applications such as in the analysis of environmental pollutants.Different approaches have been adopted to enhance selectivity for the analysis of different types of compounds. There are two approaches most commonly used to improve CE separations the addition of modifiers into the electrophoretic medium and the modification of the column. Examples of the first approach include the addition of surfactants into the electrophoretic medium as in micellar electrokinetic chromatography (MEKC), and the use of organic solvents,cyclodextrines, " " or bile salts " as buffer modifiers. Examples of the second method include the use of gel-filled columns (capillary gel electrophoresis) " and the coating of the capillary wall surface. " " ... [Pg.430]

This section has thus presented a quick synthesis and analysis method for two simplified infinite reflux cases. Just like simple columns, it can be generally stated that if a design is considered feasible at infinite reflux conditions, then a feasible design can be found at finite reflux too. This fact is particularly useful for nonideal systems. An illustration of a more complex infinite Petlyuk column example is given in the following example for the azeotropic acetone/benzene/chloroform system. [Pg.215]

Fig. 6.6 Cumulative Distribution Function for y in Steel Column Example. Fig. 6.6 Cumulative Distribution Function for y in Steel Column Example.
Figure 4.B.1 Separation of an acetone, methanol and toluene mixture using three distillation columns example in Aspen Plus V8.4. Figure 4.B.1 Separation of an acetone, methanol and toluene mixture using three distillation columns example in Aspen Plus V8.4.
EXAMPLE 21.9 Tuning PI Control Loops for a Binary Distillation Column (Example 21.8 Revisited)... [Pg.735]


See other pages where Columns examples is mentioned: [Pg.410]    [Pg.32]    [Pg.177]    [Pg.23]    [Pg.28]    [Pg.1753]    [Pg.434]    [Pg.153]    [Pg.288]    [Pg.305]    [Pg.164]    [Pg.434]    [Pg.313]    [Pg.67]    [Pg.331]    [Pg.137]    [Pg.172]    [Pg.1747]    [Pg.486]    [Pg.489]    [Pg.742]    [Pg.64]    [Pg.351]    [Pg.485]    [Pg.725]   
See also in sourсe #XX -- [ Pg.374 , Pg.375 , Pg.397 , Pg.398 , Pg.399 ]

See also in sourсe #XX -- [ Pg.251 ]




SEARCH



Chromatographic separations example, 125 mm column

Column Sizing Example

Column Sizing Example Trays or Packings

Columns, reversed HPLC examples

Examples for a Single Batch Chromatographic Column

© 2024 chempedia.info