Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Cantilever-mounted

Optical or laser alignment systems are based on the same principles as the reverse-dial method, but replace the mechanical components such as runout gages and cantilevered mounting arms with an optical device such as a laser. As with the reverse-dial method, offset is measured and angularity is calculated. [Pg.923]

In this case the designer has freedom of choice of both form and dimension as well as in the selection of the materials. Given this freedom, it would be desirable to examine several of the alternatives to see which would provide the best seating at the lowest cost. Obviously, there is no point in doing all of the possibilities so a selection should be made on the basis of anticipated use as well as style requirements. Three types will be analyzed. They are the single curve sheet cantilever mounted from the back, the molded pan supported on four legs, and the structural foam molding which is front supported. [Pg.251]

Earlier attempts to use the AFM for mechanically stretching chromatin fibers have run into a rather unexpected artifact. Long native chromatin fibers isolated from chicken erythrocytes, or fibers assembled in vitro from purified histones and relatively short, tandemly repeated DNA sequences were deposited on mica or glass surfaces and pulled with the AFM tip [69,70]. In such stretching experiments the scanning of the sample in the x- and y-direction used for imaging was disabled, and the cantilever-mounted tip was allowed to move only in the z-direction, i.e., upwards and downwards, away and towards the surface. When the AFM tip is pushed into the sample, it may attach to the sample by non-specific adsorption upon retraction it stretches the sample and force-extension curves are recorded (see Fig. lb for an explanation of a typical force curve). [Pg.387]

The cantilever is excited into resonance by electrically exciting the piezoelectric cantilever mount. The frequency of the excitation wave is scanned in a given frequency range, and the frequency of maximum cantilever amplitude is taken as the resonance frequency. The frequency spectrum of the cantilever response shows the fundamental frequency as well as the harmonics of cantilever vibration. The cantilevers, however, also resonate in response to ambient conditions such as room temperature or acoustic noise without requiring any external power. [Pg.250]

Figure 2 shows a schematic of a typical AFM instrument that consists of a cantilever-mounted tip, a Piezoelectric scanner, four position-sensitive photo detectors, a laser diode and a control unit. The process of operation of an AFM is relatively simple. The beam from the laser is directed against the back of the cantilever beam onto the quadrants of the photo detector. As the tip is moved across a sample, this causes the cantilever beam to bend or be twisted in manner that is proportional to the interaction force. This bending or twisting of the cantilever causes the position of the laser on the photo detector to be altered. The deflection of the cantilever beam can then be converted into a 3D topographical image of the sample surface (Gaboriaud and Dufrene, 2007 Kuznetsova et al., 2007 Lim et al., 2006). [Pg.34]

The second arrangement to meet the criteria is to have both screws end to end, running in a single casing. Cantilever-mounted screws from each end, with separate drives, then reverse at full speed to serve their nearest outlet, and run at half speed in the forward direction to feed material towards the other outlet. The hopper outlet slot is three pitches long, so half the output is taken by one screw and half by the other, whatever outlet(s) are being served (Fig. 7.1). [Pg.144]

Cantilever mount Wollaston wire Platinum core Probe tip... [Pg.988]

In dynamic loading, the probe is initially out of contact with the surface. Most scanning force microscopes perform dynamic loading under displacement control of the sample (but not the probe), e.g., the distance Z(t) between the cantilever mount and sample is reduced at constant speed V,>,. The probe and sample are then brought into contact, the contact is loaded, and then unloaded until it ruptures. Figure 3a shows typical loading behavior for a viscoelastic substrate, poly(vinylethylene) (9). [Pg.69]

Unfortunately the formula given in BS 4076 calculates the natural frequency for transverse vibration of a cantilever mounted on a rigid foundation. While this is suitable for most land-based chinmey designs it will yield an unduly optimistic estimate for an exhaust stack supported on pinned joints. The error will be increased fruther when consideration is also given to the flexibility of the supporting structure. It is clear that a more detailed analysis of the vibration behavior is required. [Pg.703]

It is interesting to note that the corresponding natural frequency for a 10-m-high cantilever mounted on a rigid foimdation is 19.6 Hz. The difference between these values emphasizes the importance of ensuring that the recommendations of the relevant standard have been interpreted correctly. [Pg.706]


See other pages where Cantilever-mounted is mentioned: [Pg.508]    [Pg.1219]    [Pg.145]    [Pg.508]    [Pg.1042]    [Pg.1223]    [Pg.56]    [Pg.129]    [Pg.135]    [Pg.146]    [Pg.150]    [Pg.151]    [Pg.151]    [Pg.369]    [Pg.266]    [Pg.476]    [Pg.271]    [Pg.347]    [Pg.216]    [Pg.353]   


SEARCH



Cantilever-mounted screws

Cantilevers

Mount

Screws cantilever-mounte

© 2024 chempedia.info