Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration constraints

The area of the chromatogram for the unknown sample can also be utilized to generate a weight fraction distribution, but as a function of eluent volume, i (see Figure 4). At a constant mass fraction, the two distributions are equal and can be utilized to generate a calibration curve to check the validity of the semi-logarithmic calibration constraint. Equation 1. Figure 5 pre-... [Pg.119]

Thermal Heat Capacity - The heat capacity of SiOC-N312 BN 2-D composites was measured by differential scanning calorimetry (DSC). In this test a sample of dimensions 4.24 X 4.24 X 1 mm is placed in a calibrated heating chamber along with a known heat capacity standard, and the chamber is heated at a fixed heating rate. The temperature difference between the standard and the composite is recorded, and the heat capacity is calculated from the measured temperature difference, the heat capacity of the standard, and the calibration constraints for the system. [Pg.366]

The response surfaces in Figure 14.2 are plotted for a limited range of factor levels (0 < A < 10, 0 < B < 10), but can be extended toward more positive or more negative values. This is an example of an unconstrained response surface. Most response surfaces of interest to analytical chemists, however, are naturally constrained by the nature of the factors or the response or are constrained by practical limits set by the analyst. The response surface in Figure 14.1, for example, has a natural constraint on its factor since the smallest possible concentration for the analyte is zero. Furthermore, an upper limit exists because it is usually undesirable to extrapolate a calibration curve beyond the highest concentration standard. [Pg.667]

Random Measurement Error Third, the measurements contain significant random errors. These errors may be due to samphng technique, instrument calibrations, and/or analysis methods. The error-probability-distribution functions are masked by fluctuations in the plant and cost of the measurements. Consequently, it is difficult to know whether, during reconciliation, 5 percent, 10 percent, or even 20 percent adjustments are acceptable to close the constraints. [Pg.2550]

Referring to equation [40], we can see that we require the absorbance at each wavelength to equal zero whenever the concentrations of all the components in a sample are equal to zero. We can add some flexibility to the CLS calibration by eliminating this constraint. This will add one additional degree of freedom to the equations. To allow these non-zero intercepts, we simply rewrite equation [40] with a constant term for each wavelength ... [Pg.63]

Schwartz, L. M., Effect of Constraints on Precision of Calibration Analyses, Anal. Chem. 58, 1986, 246-250. [Pg.408]

If the final structure either deviates from the refined model or does not match the NMR restraints (8) one has to revise the experimental data and the parameters used in the DG and MD computations (9). In many cases, mistakes are made when preparing and performing the computational processes (10) or even experimental errors might be present (11). Those errors include a wrong NMR peak assignment, no precise calibration of the NOE/ROE signals, an incorrect conversion of the experimental data to constraints, and a nonfactual parameterization of the rMD and fMD trajectories. In such cases either new calculations or new experiments must be performed. [Pg.245]

The key characteristic of a RM is that the properties of interest are measured and certified on the basis of accuracy. The means of attaining the true value are varied, and several different philosophies have been utilized in the quest for the best estimate of the true value. The goal of all approaches is arrival at the best possible estimate of the true value a reliable and unassailable numerical value of the concentration of the chemical constituent, under constraints of economics, state-of-the-art analytical technologies, availability of (new and old) methods, analyst competence, availability of analysts and RM end-use requirement. The basic requirement for producing reliable data is appropriate methodology, adequately calibrated and properly used. [Pg.51]

Classical calibration procedure can only be applied when all the species that contribute to the form of the spectra are known and can be included into the calibration. Additionally, there is the constraint that no interactions between the analytes and other species (e.g. solvent) or effects (e.g. of temperature) should occur. [Pg.184]

Generally convection is calibrated by requiring that its free parameter(s) are chosen to reproduce the solar radius at the solar age. However, it is also possible to use models which do not fit the solar location, and in fact these seem to reproduce much better two observational constraints of the pre main sequence, namely the PMS Lithium depletion and the HR diagram location of some binary stars for which masses are known. [Pg.291]

In practice, the time constraint condition may be met by a careful choice of appropriate experimental conditions for a given transducer. Kinetic data for the process of interest is of great help in choosing those conditions. The validity of r <time profile of the calibration and sample waveforms or by varying conditions (e.g., sample concentration) until the experimental waveform reaches a maximum in amplitude. [Pg.195]

As mentioned previously, a photoacoustic calorimetry experiment consists of two consecutive runs, the calibration and the experiment. In the first one, a photoacoustic calibrant is used to determine the proportionality constant K in equation 13.9. Then the procedure is repeated with the sample of interest, ensuring that the time constraint referred to is met and also that the experimental conditions are as close as possible to the calibration (maintaining constant the factors that affect K). Because the procedure is identical for both calibration and experiment, it will be illustrated here for the simpler case of the calibration (0bs = 1) for which equation 13.9 reduces to equation 13.17 ... [Pg.200]

What are the main error sources in PAC experiments One of them may result from the calibration procedure. As happens with any comparative technique, the conditions of the calibration and experiment must be exactly the same or, more realistically, as similar as possible. As mentioned before, the calibration constant depends on the design of the calorimeter (its geometry and the operational parameters of its instruments) and on the thermoelastic properties of the solution, as shown by equation 13.5. The design of the calorimeter will normally remain constant between experiments. Regarding the adiabatic expansion coefficient (/), in most cases the solutions used are very dilute, so the thermoelastic properties of the solution will barely be affected by the small amount of solute present in both the calibration and experiment. The relevant thermoelastic properties will thus be those of the solvent. There are, however, a number of important applications where higher concentrations of one or more solutes have to be used. This happens, for instance, in studies of substituted phenol compounds, where one solute is a photoreactive radical precursor and the other is the phenolic substrate [297]. To meet the time constraint imposed by the transducer, the phenolic... [Pg.201]

MCR methods [88-92] use the exact same Beer s Law model as the CLS method (Equation 12.35). However, unlike the CLS method, the MCR method can be used without knowing the concentrations of all of the constituents in the calibration samples. Given only the spectral data (X), an estimate of the number of chemically varying and spectrally active components represented in the data (A), and some constraints regarding the nature of the expected structures of the pure component prohles (C) and pure component spectra (K), this method provides estimates of the pure component spectra (K) and the pure component concentrations (C). [Pg.403]

The linear calibration method has utility for characterization of water-soluble polymers due to the constraints imposed in aqueous SEC towards universal calibration methodology. A cursory evaluation of the linear calibration method for aqueous SEC indicates the method can be used with a high degree of accuracy to calculate molecular weight distribution values. [Pg.94]

Figure 1. Schematic for GPC calibration subject to Poisson constraints. Figure 1. Schematic for GPC calibration subject to Poisson constraints.
One of the simplest calorimetric methods is combustion bomb calorimetry . In essence this involves the direct reaction of a sample material and a gas, such as O or F, within a sealed container and the measurement of the heat which is produced by the reaction. As the heat involved can be very large, and the rate of reaction very fast, the reaction may be explosive, hence the term combustion bomb . The calorimeter must be calibrated so that heat absorbed by the calorimeter is well characterised and the heat necessary to initiate reaction taken into account. The technique has no constraints concerning adiabatic or isothermal conditions hut is severely limited if the amount of reactants are small and/or the heat evolved is small. It is also not particularly suitable for intermetallic compounds where combustion is not part of the process during its formation. Its main use is in materials thermochemistry where it has been used in the determination of enthalpies of formation of carbides, borides, nitrides, etc. [Pg.82]


See other pages where Calibration constraints is mentioned: [Pg.271]    [Pg.120]    [Pg.71]    [Pg.439]    [Pg.486]    [Pg.78]    [Pg.107]    [Pg.288]    [Pg.56]    [Pg.57]    [Pg.194]    [Pg.16]    [Pg.64]    [Pg.450]    [Pg.366]    [Pg.417]    [Pg.16]    [Pg.54]    [Pg.55]    [Pg.272]    [Pg.177]    [Pg.1]    [Pg.285]    [Pg.245]    [Pg.246]    [Pg.15]    [Pg.95]    [Pg.15]    [Pg.12]    [Pg.180]    [Pg.197]   


SEARCH



© 2024 chempedia.info