Atmospheric pressure APCI

Atmospheric pressure chemical ionization (APCI) is a gas phase ionization process based on ion-molecule reactions between a neutral molecule and reactant ions [31]. The method is very similar to chemical ionization with the difference that ionization occurs at atmospheric pressure. APCI requires that the liquid sample is completely evaporated (Fig. 1.12). Typical flow rates are in the range 200-1000 xL min , but low flow APCI has also been described. First, an aerosol is formed with the help of a pneumatic nebulizer using nitrogen. The aerosol is directly formed in a heated quartz or ceramic tube (typical temperatures 200-500 °C) where the mobile phase and the analytes are evaporated. The temperature of the nebulized mobile phase itself remains in the range 120-150 °C due to evapo-... [Pg.17]

Two carotenoid contents of lycopene (3) and P-carotene (2) in kiwifruit were measured by a new method of liquid chromatography coupled with mass spectrometry (LC-MS). Their molecular ion was selected for their quantification in selective ion monitoring (SIM) mode. The chemical ionization atmospheric pressure (APCI) in positive mode was used for their target carotenoids. [Pg.43]

Closely related to ESI is APCI in that the source operates at near-atmospheric pressure. APCI produces almost molecular ions with very little fragmentation, and it provides fingerprints of herbs [22]. [Pg.62]

One of the first successful techniques for selectively removing solvent from a solution without losing the dissolved solute was to add the solution dropwise to a moving continuous belt. The drops of solution on the belt were heated sufficiently to evaporate the solvent, and the residual solute on the belt was carried into a normal El (electron ionization) or Cl (chemical ionization) ion source, where it was heated more strongly so that it in turn volatilized and could be ionized. However, the moving-belt system had some mechanical problems and could be temperamental. The more recent, less-mechanical inlets such as electrospray have displaced it. The electrospray inlet should be compared with the atmospheric-pressure chemical ionization (APCI) inlet, which is described in Chapter 9. [Pg.55]

The term nebulizer is used generally as a description for any spraying device, such as the hair spray mentioned above. It is normally applied to any means of forming an aerosol spray in which a volume of liquid is broken into a mist of vapor and small droplets and possibly even solid matter. There is a variety of nebulizer designs for transporting a solution of analyte in droplet form to a plasma torch in ICP/MS and to the inlet/ionization sources used in electrospray and mass spectrometry (ES/MS) and atmospheric-pressure chemical ionization and mass spectrometry (APCI/MS). [Pg.138]

The LC/TOF instmment was designed specifically for use with the effluent flowing from LC columns, but it can be used also with static solutions. The initial problem with either of these inlets revolves around how to remove the solvent without affecting the substrate (solute) dissolved in it. Without this step, upon ionization, the large excess of ionized solvent molecules would make it difficult if not impossible to observe ions due only to the substrate. Combined inlet/ionization systems are ideal for this purpose. For example, dynamic fast-atom bombardment (FAB), plas-maspray, thermospray, atmospheric-pressure chemical ionization (APCI), and electrospray (ES)... [Pg.163]

In one instrument, ions produced from an atmospheric-pressure ion source can be measured. If these are molecular ions, their relative molecular mass is obtained and often their elemental compositions. Fragment ions can be produced by suitable operation of an APCI inlet to obtain a full mass spectrum for each eluting substrate. The system can be used with the effluent from an LC column or with a solution from a static solution supply. When used with an LC column, any detectors generally used with the LC instrument itself can still be included, as with a UV/visible diode array detector sited in front of the mass spectrometer inlet. [Pg.167]

To achieve sufficient vapor pressure for El and Cl, a nonvolatile liquid will have to be heated strongly, but this heating may lead to its thermal degradation. If thermal instability is a problem, then inlet/ionization systems need to be considered, since these do not require prevolatilization of the sample before mass spectrometric analysis. This problem has led to the development of inlet/ionization systems that can operate at atmospheric pressure and ambient temperatures. Successive developments have led to the introduction of techniques such as fast-atom bombardment (FAB), fast-ion bombardment (FIB), dynamic FAB, thermospray, plasmaspray, electrospray, and APCI. Only the last two techniques are in common use. Further aspects of liquids in their role as solvents for samples are considered below. [Pg.279]

El = electron ionization Cl = chemical ionization ES = electrospray APCI = atmospheric-pressure chemical ionization MALDI = matrix-assisted laser desorption ionization PT = plasma torch (isotope ratios) TI = thermal (surface) ionization (isotope ratios). [Pg.280]

When mass spectrometry was first used as a routine analytical tool, El was the only commercial ion source. As needs have increased, more ionization methods have appeared. Many different types of ionization source have been described, and several of these have been produced commercially. The present situation is such that there is now only a limited range of ion sources. For vacuum ion sources, El is still widely used, frequently in conjunction with Cl. For atmospheric pressure ion sources, the most frequently used are ES, APCI, MALDI (lasers), and plasma torches. [Pg.282]

Electrospray Ionization (ES) and Atmospheric Pressure Chemical Ionization (APCI)... [Pg.283]

Thus, either the emitted light or the ions formed can be used to examine samples. For example, the mass spectrometric ionization technique of atmospheric-pressure chemical ionization (APCI) utilizes a corona discharge to enhance the number of ions formed. Carbon arc discharges have been used to generate ions of otherwise analytically intractable inorganic substances, with the ions being examined by mass spectrometry. [Pg.388]

Samples containing mixtures of peptides can be analyzed directly by electrospray. Alternatively, the peptides can be separated and analyzed by LC/MS coupling techniques such as electrospray or atmospheric pressure chemical ionization (APCI). [Pg.417]

A liquid chromatography-mass spectrometry (LC-MS) method that can quantitatively analyze urinar y normal and modified nucleosides in less than 30 min with a good resolution and sufficient sensitivity has been developed. Nineteen kinds of normal and modified nucleosides were determined in urine samples from 10 healthy persons and 18 breast cancer patients. Compounds were separ ated on a reverse phase Kromasil C18 column (2.1 mm I.D.) by isocratic elution mode using 20 mg/1 ammonium acetate - acetonitrile (97 3 % v/v) at 200 p.l/min. A higher sensitivity was obtained in positive atmospheric pressure chemical ionization mode APCI(-i-). [Pg.351]

An on-line chromatography/atmospheric pressure chemical ionization tandem mass spectrometry (LC-APCI/MS/MS) methods was developed for rapid screen of pharmacokinetics of different drugs, including 5 (98RCM1216). The electron impact mass spectrum of 5 and ethyl 9,10-difluoro-3-methyl-7-oxo-2,3-dihydro-7Ff-pyrido[l,2,3- fe]-l,4-benzoxazine-6-carboxylate was reported (97MI28). Electron impact/Fourier transform... [Pg.268]

Atmospheric pressure chemical ionization (APCI) Chemical ionization at atmospheric pressure. [Pg.372]

The pump must provide stable flow rates from between 10 ttlmin and 2 mlmin with the LC-MS requirement dependent upon the interface being used and the diameter of the HPLC column. For example, the electrospray interface, when used with a microbore HPLC column, operates at the bottom end of this range, while with a conventional 4.6 mm column such an interface usually operates towards the top end of the range, as does the atmospheric-pressure chemical ionization (APCI) interface. The flow rate requirements of the different interfaces are discussed in the appropriate section of Chapter 4. [Pg.27]

Ionization methods that may be utihzed in LC-MS include electron ionization (El), chemical ionization (Cl), fast-atom bombardment (FAB), thermospray (TSP), electrospray (ESI) and atmospheric-pressure chemical ionization (APCI). [Pg.52]

Atmospheric-pressure chemical ionization (APCI) is another of the techniques in which the stream of liquid emerging from an HPLC column is dispersed into small droplets, in this case by the combination of heat and a nebulizing gas, as shown in Figure 4.21. As such, APCI shares many common features with ESI and thermospray which have been discussed previously. The differences between the techniques are the methods used for droplet generation and the mechanism of subsequent ion formation. These differences affect the analytical capabilities, in particular the range of polarity of analyte which may be ionized and the liquid flow rates that may be accommodated. [Pg.180]

Cl is an efficient, and relatively mild, method of ionization which takes place at a relatively high pressure, when compared to other methods of ionization used in mass spectrometry. The kinetics of the ion-molecule reactions involved would suggest that ultimate sensitivity should be obtained when ionization takes place at atmospheric pressure. It is not possible, however, to use the conventional source of electrons, a heated metallic filament, to effect the initial ionization of a reagent gas at such pressures, and an alternative, such as Ni, a emitter, or a corona discharge, must be employed. The corona discharge is used in commercially available APCI systems as it gives greater sensitivity and is less hazardous than the alternative. [Pg.181]

Particular emphasis has been placed upon electrospray and atmospheric-pressure chemical ionization (APCI) which, in addition to being the currently most widely used interfaces, are ionization techniques in their own right. [Pg.184]

Figure 5.1 Pesticides included in the systematic investigations on APCI-MS signal response dependence on eluent flow rate the parameter IsTow represents the distribution coefficient of the pesticide between n-octanol and water. Reprinted from J. Chromatogr, A, 937, Asperger, A., Efer, 1., Koal, T. and Engewald, W., On the signal response of various pesticides in electrospray and atmospheric pressure chemical ionization depending on the flow rate of eluent applied in liquid chromatography-mass spectrometry , 65-72, Copyright (2001), with permission from Elsevier Science. Figure 5.1 Pesticides included in the systematic investigations on APCI-MS signal response dependence on eluent flow rate the parameter IsTow represents the distribution coefficient of the pesticide between n-octanol and water. Reprinted from J. Chromatogr, A, 937, Asperger, A., Efer, 1., Koal, T. and Engewald, W., On the signal response of various pesticides in electrospray and atmospheric pressure chemical ionization depending on the flow rate of eluent applied in liquid chromatography-mass spectrometry , 65-72, Copyright (2001), with permission from Elsevier Science.
Atmospheric-pressure chemical ionization (APCI) and electrospray ionization are both soft ionization techniques which give rise, almost exclusively, to the production of molecular species. Structural information. [Pg.205]

Atmospheric-pressure chemical ionization (APCI) An ionization memod in which a liquid stream is passed through a heated capillary and a concentric flow of a nebulizing gas. Ions are formed by ion-molecule reactions between me analyte and species derived from me HPLC mobile phase. [Pg.303]


See other pages where Atmospheric pressure APCI is mentioned: [Pg.360]    [Pg.268]    [Pg.369]    [Pg.871]    [Pg.543]    [Pg.547]    [Pg.360]    [Pg.268]    [Pg.369]    [Pg.871]    [Pg.543]    [Pg.547]    [Pg.38]    [Pg.60]    [Pg.71]    [Pg.73]    [Pg.163]    [Pg.277]    [Pg.283]    [Pg.248]    [Pg.218]    [Pg.14]    [Pg.197]   