Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analytical quantity

Photochemistry in this temperature range has been restricted to analytical quantities in connection with mechanistic work. The inclusion of Figure 13-5 in this chapter may also help to encourage preparative attempts in this field. [Pg.292]

Very high sensitivity trace analysis (fg-pg analyte quantities)... [Pg.394]

Sensitivity, derivative of the measured quantity (response) y with respect to the analytical quantity x Total multicomponent sensitivity... [Pg.16]

The latent information of the measuring sample is transferred via an energetic carrier into analytical information which is manifested by signals. Their parameters correspond to the coding (encoding) process in information systems. For the formal representation of the analytical coding the following analytical quantities are introduced ... [Pg.55]

A classification of the basis of these analytical quantities with regard to their mutual relationships and dependencies from physical quantities like space co-ordinates and time will be given in Sect. 3.4 according to the dimensionality of analytical information. [Pg.56]

Fig. 2.17. The four analytical quantities in sample and signal domain and the six fundamental functions between them the functions Q = f l(x) and z = do not make... Fig. 2.17. The four analytical quantities in sample and signal domain and the six fundamental functions between them the functions Q = f l(x) and z = do not make...
For the analytical representation the signals have to be transformed from time functions into conventional measuring functions. These are characterized by analytical quantities on abscissa and ordinate axes where the values of them may be relativized in some cases (e.g. MS). Such a transformation of quantities is mostly carried out on the basis of instrument-internal adjustment and calibration. [Pg.73]

In the following chapters, these analytical quantities and parameters will be considered in more detail in the given relationships. [Pg.89]

In analytical chemistry, calibration represents a set of operations that connects quantities in the sample domain with quantities in the signal domain (see Sect. 2.3, Fig. 2.12). In Table 6.1 the real analytical quantities and properties behind the abstract input and output quantities are listed. [Pg.150]

Depending on the type of relationships between the measured quantity and the measurand (analytical quantity) it can be distinguished (Danzer and Currie [1998]) between calibrations based on absolute measurements (one calibration is valid for all1 on the basis of the simple proportion y = b x, where the sensitivity factor b is a fundamental quantity see Sect. 2.4 Hula-nicki [1995] IUPAC Orange Book [1997, 2000]), definitive measurements (b is given either by a fundamental quantity complemented by an empirical factor or a well-known empirical (transferable) constant like molar absorption coefficient and Nernst factor), and experimental calibration. [Pg.150]

As mentioned above, the random character of the input and output variables are of importance with regard to the calibration model and its estimation by calculus of regression. Because of the different character of the analytical quantity x in the calibration step (no random variables but fixed variables which are selected deliberately) and in the evaluation step (random variables like the measured values), the closed loop of Fig. 6.1 does not correctly describe the situation. Instead of this, a linear progress as shown in Fig. 6.2 takes place. [Pg.151]

The several variants deriving from the items 1 to 4 are represented in the flow sheet given in Fig. 6.6. Common calibration by Gaussian least squares estimation (OLS) can only be applied if the measured values are independent and normal-distributed, free from outliers and leverage points and are characterized by homoscedastic errors. Additionally, the error of the values in the analytical quantity x (measurand) must be negligible compared with the errors of the measured values y. [Pg.159]

Magnitude of an analytical quantity, x, measured at test samples on the one hand... [Pg.309]

Equation for the estimation of the values of a measuring quantity from given values of a analytical quantity. The calibration function may be known a priori by natural laws or estimated experimentally by means of calibration samples. [Pg.310]

It has been suggested however that isotacticity derives from polymerization occurring on colloidal particles formed by thermal decomposition of the catalysts. As stated previously, in the presence of the monomer even the allyl compounds are stable at 65°C and none of the thermal decomposition products (black to yellow solids) could be detected. As a check on these results a polymerization of propylene was carried out with Zr (benzyl) 4 in toluene at 0°C in a sealed tube. The reaction was very slow and analytical quantities of polymer could be obtained only after 312 hr. NMR analysis showed peaks assignable to isotactic sequences, and these were much stronger than the peaks assignable to syndiotactic diads. It was concluded... [Pg.300]

The 5-mm tube held in the sample probe is indicative of the sample size used in an NMR instrument— typically some fraction of a milliliter. Typical analyte quantities dissolved in this volume range from 10... [Pg.287]

Determination of Accuracy and Precision of the Analytical Procedure. The desorption efficiency was determined for each method at widely separated analyte quantities to establish the average recovery to be expected. The spiking and analysis procedures for these tests were similar to those described earlier for the preliminary desorption efficiency tests. For HCCP and... [Pg.61]

The use of radiolabeled tracer analyte (quantity too small to affect the assay)... [Pg.113]

Before proceeding, a few general comments will be helpful. In most cases at least one of the four carbonate system analytical quantities (total alkalinity, total CO2, carbon dioxide partial pressure, and pH) will be known. The equilibrium equations relating these quantities, along with those for water and calcite, will be used frequently in calculations. Also, in cases where calcite dissolution and... [Pg.55]

As with univariate and multivariate calibration, three-way calibration assumes linear additivity of signals. When the sample matrix influences the spectral profiles or sensitivities, either care must be taken to match the standard matrix to those of the unknown samples, or the method of standard additions must be employed for calibration. Employing the standard addition method with three-way analysis is straightforward only standard additions of known analyte quantity are needed [42], When the standard addition method is applied to nonbilinear data, the lowest predicted analyte concentration that is stable with respect to the leave-one-out cross-validation method is unique to the analyte. [Pg.496]

Many examples of calibration curves will be found in later pages of this text. A good calibration curve will be linear over a good range of analyte quantities. Ultimately curvature must be anticipated at the higher ranges and uncertainty and deteriorating precision at the low ones. [Pg.32]

In Sec. 3.3.4 the detection limit and related analytical quantities are described in detail. In this section we define the detection limit as the concentration c we get from a signal which is three times as high as the standard deviation of the background (Kaiser, 1965). The signal should be the absorbance which is proportional to the concentration. The mean value of the noise amplitude will be approximately 5 times the standard deviation (Doerffel, 1988). With this assumption we get... [Pg.612]

Analyte distribution function for the mass balance is the sum of the analyte quantities in all phases ... [Pg.65]

If Wq is too small, a biased sample deficient in coarse particles, results. For this reason Wq should be at least 3d where d is the diameter of the largest particle present in the bulk. ISO 3081 suggests a minimum incremental mass based on the maximum particle size in mm. These values are given in Table 1.8. Secondary samplers then reduce this to analytical quantities. [Pg.52]

Potentiometric transducers measure the potential under conditions of constant current. This device can be used to determine the analytical quantity of interest, generally the concentration of a certain analyte. The potential that develops in the electrochemical cell is the result of the free-energy change that would occur if the chemical phenomena were to proceed until the equilibrium condition is satisfied. For electrochemical cells containing an anode and a cathode, the potential difference between the cathode electrode potential and the anode electrode potential is the potential of the electrochemical cell. If the reaction is conducted under standard-state conditions, then this equation allows the calculation of the standard cell potential. When the reaction conditions are not standard state, however, one must use the Nernst equation to determine the cell potential. Physical phenomena that do not involve explicit redox reactions, but whose initial conditions have a non-zero free energy, also will generate a potential. An example of this would be ion-concentration gradients across a semi-permeable membrane this can also be a potentiometric phenomenon and is the basis of measurements that use ion-selective electrodes (ISEs). [Pg.540]

The use of solid standards introduces a different concept of calibration. Traditional calibration is done by changing the analyte content. With micro-weighings, accurately replicating the same sample mass is difficult. Therefore, calibration with solid standards for ETA is done with powdered reference materials by varying the sample mass. Conventional two-dimensional calibration focuses on the analyte mass without considering the corresponding matrix mass. The analyte mass is an analytical quantity derived... [Pg.375]

Column chromatography with hexanes on flash silica gel gave a few fractions of with 95% purity, as determined by HPLC, along with later fractions containing mixtures of Cio/C a in various ratios. Because of the poor solubility of C o and C70 in these alkanes, only limited amounts of pure C o were made available this way, in insufficient quantity for a reliable C NMR spectrum to be measured (see below). However, column chromatography on neutral alumina with hexanes gave an excellent separation in analytical quantities. Thus, pure fractions containing C (99.85%) and C70 (>99%) were obtained, as indicated by mass spectrometric measurements described below. [Pg.39]


See other pages where Analytical quantity is mentioned: [Pg.341]    [Pg.78]    [Pg.177]    [Pg.308]    [Pg.17]    [Pg.38]    [Pg.37]    [Pg.370]    [Pg.17]    [Pg.41]    [Pg.28]    [Pg.349]    [Pg.314]    [Pg.1319]    [Pg.182]    [Pg.259]    [Pg.52]   
See also in sourсe #XX -- [ Pg.32 , Pg.124 , Pg.284 ]

See also in sourсe #XX -- [ Pg.32 , Pg.124 , Pg.284 ]




SEARCH



The Ab Initio Analytical Approach of Vibronic Quantities Application to Inorganic Stereochemistry

© 2024 chempedia.info