Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

A.c. radiometer

In the stepping mode, the signal is integrated at each rest position for a certain time, the dwell time. After that, the Michelson mirror advances to the following position and the next point is recorded. As in the case of an a.c. radiometer an external chopper may be used to modulate the incoming radiation. The stepping technique has been perfected by J. and P. Coimes. [Pg.223]

A spectro-radiometer-luminometer for chemiluminescence and fluorescence quantum yield studies has been described by B. G. Roberts and H. C. Hirt 187>. To obtain emission spectra from very weak chemiluminescence reactions, a large-aperture spectrograph combined with a sensitive image-intensifier tube has been used68 this was developed from a device previously constructed by Bass and Kessler 188>. With it, it was possible to record the very weak emission of singlet oxygen dimer... [Pg.131]

Fig. 2. (A) Left lateral gamma images of dogs with acute experimental MI. Images a, c, e, and g were obtained after iv administration of conventionally modified " In-DTPA-antimyosin Fab (AM-Fab) images b, d, f, and h were obtained after administration of AM-Fab modified with 11 In-loaded chelating polymer (PL-AM Fab). The heavy load of chelating polymer-modified AM Fab with reporter radiometal permits one to obtain a clear infarct image in the left lateral oblique position only in 3 h (bright spot on image h). Whereas the use of traditionally labeled antibody requires <24 h to yield images of a comparable quality. From ref. 21. Fig. 2. (A) Left lateral gamma images of dogs with acute experimental MI. Images a, c, e, and g were obtained after iv administration of conventionally modified " In-DTPA-antimyosin Fab (AM-Fab) images b, d, f, and h were obtained after administration of AM-Fab modified with 11 In-loaded chelating polymer (PL-AM Fab). The heavy load of chelating polymer-modified AM Fab with reporter radiometal permits one to obtain a clear infarct image in the left lateral oblique position only in 3 h (bright spot on image h). Whereas the use of traditionally labeled antibody requires <24 h to yield images of a comparable quality. From ref. 21.
A similar treatment to that described in section 4.2.1.4 (equations (4-12) to (4-18)) is applied to the readings using a UVX radiometer with the UV-opaque collimator, qi,c in order to relate them to the rate of transmission non-scattered photons, Pfnsit). The resultant equation becomes... [Pg.80]

The quantities in braces and in square brackets in (7.188) are both typically of order unity. Since fn yi+yl) (v "/cy fyy for y>B while it is (t jj /c)[/(/i —f2)V for yradial velocity, the system provides increasing advantage at higher radiation frequencies / (since the effective bandwidth / ) for fixed y and (/j — /2). Small linewidths and close spacing of the lines are also important. For certain choices of parameters, which are determined by the species which it is desired to detect, the SNR at the output of the three-frequency selective system will provide a sufficient confidence level for detection. For situations in which this is not the case, further improvement in the SNR could be obtained by using a multichannel receiver and/or a classical radiometer, as mentioned previously. [Pg.294]

It is to be noted that medical thermometer must measures to an accuracy of 0.1 °C. However non-invasive temperature measurement of inside a biological body is challenges and difficulties, especially for a microwave radiometer which measure thermal noise power, which is several orders of magnitude smaller than that of background noise since the system must operate at room temperature, not at liquid helium temperature like radio telescope. Moreover, it is expected to operate in a normal room (not in a shielded room). Therefore, the present target value of measurement accuracy of our system is 1 °C. [Pg.335]

Only components of detector and amplifier noise within the side bands enter the system. Low frequency detector noise is unimportant. Also eliminated are all problems associated with d.c. or very low frequency amplification. Not eliminated are effects caused by the temperature dependence of detector responsivity or of optical transmission functions. The main disadvantage of chopping, besides the complexity of the chopper mechanism and associated circuitry, is the need for a faster detector in comparison with that for a d.c. radiometer of identical base band. In a chopping radiometer the detector must respond up to the frequency limit set by the upper side band. [Pg.182]

Figure 4.14 Longitudinal cross-section of a pC02 Radiometer electrode which is used as a transducer in a number of biosensors (r- reference electrode e - intermediate electrolyte solution m - hydrophobic membrane c-protective cap). Figure 4.14 Longitudinal cross-section of a pC02 Radiometer electrode which is used as a transducer in a number of biosensors (r- reference electrode e - intermediate electrolyte solution m - hydrophobic membrane c-protective cap).
Fig. 3.83. Potentiometric dead-stop (1) and reversed dead-stop (2) end-point titrations, (a) Metrohm Polarecord 626 PARC Model 174A Polarographic Analyzer, (b) Radiometer ISS 820 Ion Scanning System, (c) Tacussel PRG4 Polarograph. [Pg.226]

Conductivity was measured as described (10) using a conductivity meter (Radiometer, Copenhagen, Denmark) type CDM3 equipped with a CDC 304 immersion electrode with manual temperature compensator type CDA 100. The instrument was calibrated as specified by the manufacturer. The determination of the (NH4)2S04 concentration from the conductivity measurements was done at constant temperature (4°C) using a calibration curve, in the range of 0.016 mM to 120 mM AS in glucose or sucrose (total... [Pg.15]

Shetter, R. E A. H. McDaniel, C. A. Cantrell, S. Madronich, and J. G. Calvert, Actinometer and Eppley Radiometer Measurements of the N02 Photolysis Rate Coefficient during the Mauna Loa Observatory Photochemistry Experiment, J. Geophys. Res., 97, 10349-10359 (1992). [Pg.85]

The thermodynamic stability constant of silver sulfadiazine in water has been measured by Boelema et al. (20). They made use of a microcomputer-controlled titrator and measured simultaneously the pH value with a Radiometer G2040C glass electrode and the silver ion concentration with an Orion research 941600 ion-selective silver sulfide electrode. The measured stability constant and standard deviation were logK=3.62i 0.05 n = 9, temperature 25°C, ionic strength 0.1 (sodium nitrate). The calculated conditional stability constant log K at pH 7.4 was 3.57. [Pg.563]

The radiation has to be chopped, and so the detector receives pulses of radiation from the source and the chopper alternately. Therefore the temperature is measured relative to the temperature of the chopper surface. In practice, radiometers are calibrated so that the temperature can be read directly from a scale. A particular commercial model which operates over the temperature range 0-600 °C uses a LiTa03 single crystal as the pyroelectric detector. [Pg.425]

Figure 1-9. Effect of varying pH on the output potential of a glass electrode at 20 and 40°C. (Courtesy of Radiometer A/S, Copenh en.)... Figure 1-9. Effect of varying pH on the output potential of a glass electrode at 20 and 40°C. (Courtesy of Radiometer A/S, Copenh en.)...
Longhurst, A., Sathyendranath, S., Platt, T., and CaverhiU, C. (1995). An estimate of global primary production in the ocean from satelhte radiometer data.J. Plank. Res. 17, 1245—1271. [Pg.298]


See other pages where A.c. radiometer is mentioned: [Pg.170]    [Pg.179]    [Pg.182]    [Pg.225]    [Pg.170]    [Pg.179]    [Pg.182]    [Pg.225]    [Pg.290]    [Pg.477]    [Pg.339]    [Pg.282]    [Pg.917]    [Pg.419]    [Pg.3236]    [Pg.2002]    [Pg.170]    [Pg.175]    [Pg.175]    [Pg.179]    [Pg.184]    [Pg.818]    [Pg.610]    [Pg.900]    [Pg.163]    [Pg.200]    [Pg.392]    [Pg.152]    [Pg.287]    [Pg.1062]    [Pg.209]    [Pg.55]    [Pg.263]    [Pg.170]    [Pg.126]    [Pg.392]    [Pg.186]   
See also in sourсe #XX -- [ Pg.170 , Pg.179 , Pg.182 , Pg.225 ]




SEARCH



Radiometals

© 2024 chempedia.info