Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Trays, sieve column diameter

Solution Table 14-12 presents measurements by Billet (loc. cit.) for ethyl-benzene-styrene under similar pressure with sieve and valve trays. The column diameter and tray spacing in Billets tests were close to those in Example 9. Since both have single-pass trays, the flow path lengths are similar. The fractional hole area (14 percent in Example 9) is close to that in Table 14-12 (12.3 percent for the tested sieve trays, 14 to 15 percent for standard valve trays). So the values in Table 14-12 should be directly applicable, that is, 70 to 85 percent. So a conservative estimate would be 70 percent. The actual efficiency should be about 5 to 10 percent higher. [Pg.53]

Testing of plates and other devices is carried out by Fractionation Research, Inc. for industrial sponsors. Some of the test data for sieve plates have been published for the cyclohexane//i-heptane and isobu-tane//i-butane systems. Representative data are shown in Fig. 14-43. These are taken from Sakata and Yanagi Jn.stn. Chem. Engis. Symp. See. No. 56, 3.2/21 (1979)] and Yanagi and Sakata [Jnd. Eng. Chem. Proc. Des. Devel, 21, 712 (1982)]. The column diameter was 1.2 m, tray spacing was 600 mm, and weir height was 50 mm. [Pg.1384]

Column diameter for a particular service is a function of the physical properties of the vapor and liquid at the tray conditions, efficiency and capacity characteristics of the contacting mechanism (bubble trays, sieve trays, etc.) as represented by velocity effects including entrainment, and the pressure of the operation. Unfortunately the interrelationship of these is not clearly understood. Therefore, diameters are determined by relations correlated by empirical factors. The factors influencing bubble cap and similar devices, sieve tray and perforated plate columns are somewhat different. [Pg.126]

The method of calculation introduced in this chapter not only allows an exact determination of the column diameter for nonpulsed sieve tray columns, but also allows a good estimation of the diameters of pulsed and stirred extractors. For the latter, however, more exact specific equations exist for the flooding point, see for example [1,4]. [Pg.394]

Sieve trays are widely used in industry with column diameters up to 3.66 m (Ref. A6 p.21.74), this limit was imposed upon the testing procedure. Column diameters of less than 1.5 m would not prove economical under these conditions because of the very large tower height and number of trays required. Although a larger column diameter would substantially reduce the required number of trays,... [Pg.290]

An extractor column is generally a tall, vertical packed tower that has two or more bed sections. Each packed bed section is typically limited to no more than 8 ft tall, making the overall tower height about 40 to 80 ft. Tower diameter depends fully upon liquid rates, but is usually in the range of 2 to 6 ft. Liquid-liquid extractors may also have tray-type column internals, usually composed of sieve-type trays without downcomers. These tray-type columns are similar to duoflow-type vapor-liquid separation, but here serve as contact surface area for two separate liquid phases. The packed-type internals are more common by far and are the type of extractor medium considered the standard. Any deviation from packed-type columns is compared to packing. [Pg.278]

Using low fractional hole areas A fractional hole area reduction to about 5 percent of the bubbling area typically boosts sieve tray turndown to about 3 to 4 1 at the expense of a lower maximum capacity, i.e., of a larger column diameter. This technique is not recommended because traying the column with valve trays is normally a cheaper alternative. [Pg.321]

Example 1 Determination of distillation-column diameter on basis of allowable vapor velocity. A sieve-tray distillation tower is to be operated under the following conditions ... [Pg.659]

The calculation of column diameter for distillation and absorption columns %h,6 is usually based on the flooding velocity, which, in turn, requires values of the flooding capacity factor, Cf- Fair s flooding-capacity plot for sieve trays [1] correlates the flooding capacity factor with a flow parameter Flv for each tray-spacing value, t, as shown in Figure 2. The flow parameter involves the liquid mass flow, LMi, and vapor mass flow, F v (both in lb/ s), as well as the densities of the two streams. [Pg.66]

To size the trays in UniSim Design, we must activate the tray sizing utility (from the Tools menu via Tools/Utilities/Tray Sizing). When sieve trays are selected with the default spacing of 609.6 mm (2 ft) and the other default parameters shown in Figure 4.25, then the results in Figure 4.26 are obtained. The column diameter is found to be 4.42 m. [Pg.193]

FIG. 15-40 Sieve tray flooding data. System toluene (dispersed) + water (continuous). Tray spacing = 30.5 cm. Column diameter = 42.8 cm. [Taken from Seibert, Bravo, and Fair ISEC 02 Proc., 2, pp. 1328-1333 (2002), with permission. Copyright 2002 South African Institute of Mining and Metallurgy.]... [Pg.1764]

A droplet containing a mixture of acetone(l)-benzene(2)-methanol(3) has a diameter of 8 mm and attains a velocity of 0.1 m/s in a sieve tray extraction column when it is dispersed in a continuous hydrocarbon phase. Use the penetration model to estimate the matrix of low flux mass transfer coefficients [A ] inside the droplet. [Pg.493]

The column has 3 m diameter sieve trays with 0.5 cm diameter holes and 10% hole area. The tray spacing is 45 cm. Assuming a foaming factor of 0.85, calculate the vapor flood velocity at the top tray. Check if the column diameter is acceptable. The fraction of flood velocity should be within a 60-85% range. [Pg.525]

Choose materials of construction based on corrosion considerations. Column diameters are determined by specifying linear velocities for the two phases. Column heights are determined by estimating the actual number of stages based on the theoretical stage requirements and average stage efficiency. Internals in pulse columns are very similar to those in distillation towers, especially for sieve trays. Therefore, distillation correlations can be used to estimate FOB purchased and installed costs for continuous differential contactors, if they are assumed to be pulse columns. [Pg.729]

Tray design encompasses the determination of the column diameter and the tray spacing as well as a number of mechanical considerations. The scope of the material in this chapter is limited primarily to the fundamentals involved in the design of single-pass sieve trays. The fundamentals involved in the design of valve trays are essentially the same as those involved in sieve trays. No attempt is made to treat bubble-cap trays, since valve and sieve trays have been used extensively in new installations since the early 1950s. Up until that time, bubble-cap trays were used almost exclusively. The design of bubble-cap trays has been treated by a number of authors see for example Van Winkle.17... [Pg.415]

Example 18.6. A sieve-plate column operating at atmospheric pressure is to produce nearly pure methanol from an aqueous feed containing 40 mole percent methanol. The distillate product rate is 5800 kg/h. (a) For a reflux ratio of 3.5 and a plate spacing of 18 in., calculate the allowable vapor velocity and the column diameter. b) Calculate the pressure drop per plate if each sieve tray is in, thick with j-in, holes on a -in. triangular spacing and a weir height of 2 in. (c) What is the froth height in the downcomer ... [Pg.566]

We consider sieve trays with a diameter d = 0.13 m. Reflux drum and column sump are sized by assuming residence times of 5 min and 10 min, respectively. [Pg.495]

Plate Columns. The much preferred plate is the sieve tray. Columns have been built successfully in diameters larger than 4.5 m. Holes from 0.64 to 0.32 cm in diameter and 1.25 to 1.91 cm apart are commonly used. Tray spacings are much closer than in distillation—10 to 15 cm in most applications involving low-interfacial-tension liquids. Plates are usually built without outlet weirs on the downspouts. A variation of the simple sieve column is the Koch Kascade Tower , where perforated plates are set in vertical arrays of moderately complex designs. [Pg.434]

When vapor and liquid flow rates change appreciably from tray to tray, column diameter, tray spacing, or hole area can be varied to reduce column cost and insure stable operation at high efficiency. Variation of tray spacing is particularly applicable to columns with sieve trays because of their low turndown ratio. [Pg.644]

The entire assembly is enclosed in a highly insulated cold box to conserve energy. The columns must be made as compact as possible to minimize capital investment as well as reduce heat leak. Up to 150 distillation trays are used in a double column system and tray spacings are kept small at 10 to 20 cm. The trays are typically multipass sieve trays with small diameter perforations. Because the distillation is an extremely clean service, perforations are typically as small as 1 mm. Many tray geometries are used, including multipass cross-flow, split cross-flow (parallel) and circular flow (race track) trays. Each has certain attributes which are used to optimize the column design for different design conditions. [Pg.11]

Which tray gives the largest column diameter (in meters) with sieve trays when one uses the... [Pg.265]


See other pages where Trays, sieve column diameter is mentioned: [Pg.18]    [Pg.266]    [Pg.1292]    [Pg.30]    [Pg.193]    [Pg.498]    [Pg.245]    [Pg.164]    [Pg.291]    [Pg.333]    [Pg.184]    [Pg.1115]    [Pg.193]    [Pg.508]    [Pg.125]    [Pg.276]    [Pg.1296]    [Pg.584]    [Pg.394]   
See also in sourсe #XX -- [ Pg.370 , Pg.371 , Pg.372 , Pg.373 , Pg.374 , Pg.375 , Pg.376 , Pg.377 ]




SEARCH



Column diameter

Columns sieve trays

Sieve trays

Tower/column diameter sieve trays

Tray Diameter

Tray columns

Trays, sieve column diameter calculation

© 2024 chempedia.info