Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Three-channel measurements

In order that the data acquisition system can obtain information about the spatial location and orientation of the probe, a four-channel incremental encoder interface board is installed. Three channels are used to define position in three-dimensional space, while the fourth monitors the skew of the probe (skew is defined as rotation about an axis normal to the probe face). Although six measurements are required to completely define the location and orientation, it is assumed that the probe remains in contact with the inspection surface. [Pg.768]

Figure 2. Schematic of the SLM 8000 fluorometer. Excitation occurs through the excitation monochromator, and light emitted from the sample is observed in as many as four different positions. Photomultiplier tubes (PMTs) A, B, and C can be used to monitor fluorescence or right-angle light scatter through the monochromator (PMT A) or through filters (PUT B and C), and position D measures transmittance. Three channels can be monitored simultaneously with measurements being acquired at intervals of 1 s or less. The data are stored by the computer for subsequent manipulation. Figure 2. Schematic of the SLM 8000 fluorometer. Excitation occurs through the excitation monochromator, and light emitted from the sample is observed in as many as four different positions. Photomultiplier tubes (PMTs) A, B, and C can be used to monitor fluorescence or right-angle light scatter through the monochromator (PMT A) or through filters (PUT B and C), and position D measures transmittance. Three channels can be monitored simultaneously with measurements being acquired at intervals of 1 s or less. The data are stored by the computer for subsequent manipulation.
Clouds and the Earth s Radiant Energy System (CERES) CERES is a broadband three-channel scanning radiometer. One channel measures reflected solar radiation in the range 0.3 pm-5.0 pm. Two other channels (0.3 pm-100 pm and 8 pm-12 pm) measure reflected and emitted radiant energy at the top of the atmosphere. [Pg.296]

Fig. 1.27. Methane steam reformer with cocurrent flow in the reaction zone, (a) Flow scheme, (b) Simulated and measured temperature profiles in the reaction zone of the three-channel reactor, (c) Sketch of the three-channel... Fig. 1.27. Methane steam reformer with cocurrent flow in the reaction zone, (a) Flow scheme, (b) Simulated and measured temperature profiles in the reaction zone of the three-channel reactor, (c) Sketch of the three-channel...
There are three primary measures of performance for a communications channel. They are distance, speed, and cost. Speed and distance are conversely related to one another. That is you can have speed or distance but not both Cost is directly related to both. That is to say if you want either speed or distance you are going to pay for it. [Pg.45]

Single-channel measurements of 39 in planar bilayers revealed three types of independent but constant conductance levels. These comprised (i) high conductance levels (700-1400 pS) which stayed open almost indefinitely (ii) an intermediate level (70-100 pS) with millisecond opening and closing kinetics, and (iii) a low conductance level (10.85 1.58 pS) which showed relatively slow gating. The... [Pg.36]

All three types of monitor may be used in the given system, connected in series. In a system used by the author the output of each of the three monitors was adjusted to range from 0 to 10 mV and all three were fed to a three-channel dot printing potentiometric chart recorder. Three different coloured traces were obtained corresponding to ultraviolet transmission, an approximately linear function of conductivity, and the count rate measured by the radio-activity flow... [Pg.269]

X-ray Photoelectron Spectroscopy. Protein films without plasticizer and uncoated from the glass plates were studied by XPS, using an Escalab MK II spectrometer with an A1 X-ray source (1486.6 eV of photon energy) and a three channel detector. Measurements were taken at an angle of 90°. [Pg.262]

The quantum yield of CN(/4) from HCN has been measured at wavelengths X< ISO nm which corresponds to excitation into the C A state, thou at 150 nm, where 0 [CN(i4)] = 0.11, the absorbed ener still lies below the t e old for channel (iii). Figure 3.4.5 indicates a direct correlation between HCN (C A ) and H( S) + CN(fi S ), (under C point group mmetiy though not under C o , where the A state becomes II) and fluorescence from CN(5) has been detected once the threshold for channel (iii) is crossed However, the combined quantum yields of CN(i4) and CN( ) remain smaller than unity throughout the range 150— 105 nm and presumably all three channels contribute in this region. [Pg.35]

Human liver tissue analyses to calculate absorbed dose in Gy have been done. Fox [28] patients that received resin microspheres during laparotomy provided an opportunity to measure portions of the liver and tumors directly from small biopsies with a specialized three-channel liquid scintillation P-radiation detection probe, which also detected gamma rays. After infusion of 49-118 million microspheres (diameter 17 or 32 pm) [24-26], Burton estimated that the normal liver received 9.0-75 Gy, and tumor 34-1474 Gy, as calculated by tumor normal ratios of 0.4 1 to 45 1 [26]. Fox et al. [28] studied the left lobe of a patient with metastatic colon cancer which was resected after having previously received resin microspheres. Predicted isodose curves by microspheres which could be seen... [Pg.52]

The notion of depth of focus is important in microfluidic and nanofluidic studies, because it influences the spatial resolution of the measurement. Let us consider a microfalaicated channel with a channel depth of d, in which fluorochrome molecules are being contained and measured. If Z > d, then all the fluorochrome molecules are always in focus, and they all are imaged clearly in the optical detector. Consequently, in microfluidic channels, where the fluorochrome molecules are typically much smaller than the channel depth d, the position of the fluorochrome molecules along the channel depth cannot be resolved anymore. Hence, multiple molecules that are positioned near each other cannot be distinguished from each other, and three-dimensional measurements (for instance, to characterize the transport of fluid inside such channels) cannot be performed. In nanofluidic channels, however, where the fluorochrome sizes are more similar to the channel depth d, the measurements can be described as a quasi-2D problem, and it is not required anymore to resolve along the channel depth. [Pg.1209]

Three channel C- F HSQC and HMBC experiments with fluorine detection have been used by Ampt et to measure /cf. fHC. Vhf and /ff couplings for three fluorinated compounds, the fluorinated steroid, 1,1,2,2-tetrafluoro-2-iodopropane and 1,2-dichloro-3,3-difluoro-1 -propene. The signs of couplings have been deduced from the splitting displacements. [Pg.211]

K. P. Proll, J.M. Nivet, C. Voland Enhancement of the dynamic range of the detected intensity in an optical measurement system by a three channel technique. Appl. Opt. 41, 130 (2002)... [Pg.903]

The ZSM-5 zeolite is an aluminosilicate with three-dimensional channels measuring 5.6 X 5.3 A size. High-resolution isotherm of Ar on ZSM-5 at 87 K has been measured by Venero and Chiou (1988). The pores are approximately cylindrical. The PSD of ZSM-5 predicted by the four models is shown in Figure 4.8. It is seen that for the parameters chosen (Table 4.2), no prominent peak was... [Pg.71]

The behavior of the device with potentials applied to all three channels was evaluated according to the scheme indicated in the inset of Figure 2. The current was measured from the potential drop across a 10 kQ resistor between ground and reservoir C, with Vi and V2 applied to reservoirs A and B, respectively. In this configuration the resistances for eq land 2 were Ri = Ra, R2 Rb, and R3 = Re The current was indeed... [Pg.108]


See other pages where Three-channel measurements is mentioned: [Pg.311]    [Pg.311]    [Pg.29]    [Pg.228]    [Pg.546]    [Pg.213]    [Pg.30]    [Pg.33]    [Pg.195]    [Pg.2550]    [Pg.99]    [Pg.294]    [Pg.415]    [Pg.418]    [Pg.138]    [Pg.72]    [Pg.319]    [Pg.172]    [Pg.314]    [Pg.160]    [Pg.76]    [Pg.1156]    [Pg.701]    [Pg.52]    [Pg.628]    [Pg.38]    [Pg.3]    [Pg.1829]    [Pg.146]    [Pg.153]    [Pg.250]    [Pg.457]    [Pg.436]    [Pg.1500]   


SEARCH



Channel three

Channeling measurements

Three Measurement

© 2024 chempedia.info