Tandem soft ionisation techniques

Tandem mass spectrometry or ms/ms was first introduced in the 1970s and gained rapid acceptance in the analytical community. The technique has been used for stmcture elucidation of unknowns (26) and has the abiUty to provide sensitive and selective analysis of complex mixtures with minimal sample clean-up (27). Developments in the mid-1980s advancing the popularity of ms/ms included the availabiUty of powerhil data systems capable of controlling the ms/ms experiment and the viabiUty of soft ionisation techniques which essentially yield only molecular ion species. [Pg.405]

Soft ionisation modes, such as API, which leave the (pseudo)molecular ion intact without much fragmentation, offer more sensitivity, and are ideal for quantitative work at low levels (e.g. breakdown products). With the use of soft ionisation techniques in LC-MS, tandem MS... [Pg.507]

Selection of a suitable ionisation method is important in the success of mixture analysis by MS/MS, as clearly shown by Chen and Her [23]. Ideally, only molecular ions should be produced for each of the compounds in the mixture. For this reason, the softest ionisation technique is often the best choice in the analysis of mixtures with MS/MS. In addition to softness , selectivity is an important factor in the selection of the ionisation technique. In polymer/additive analysis it is better to choose an ionisation technique which responds preferentially to the analytes over the matrix, because the polymer extract often consists of additives as well as a low-MW polymer matrix (oligomers). Few other reports deal with direct tandem MS analysis of extracts of polymer samples [229,231,232], DCI-MS/MS (B/E linked scan with CID) was used for direct analysis of polymer extracts and solids [69]. In comparison with FAB-MS, much less fragmentation was observed with DCI using NH3 as a reagent gas. The softness and lack of matrix effect make ammonia DCI a better ionisation technique than FAB for the analysis of additives directly from the extracts. Most likely due to higher collision energy, product ion mass spectra acquired with a double-focusing mass spectrometer provided more structural information than the spectra obtained with a triple quadrupole mass spectrometer. [Pg.403]

Direct solid-state polymer/additive mass analysis has involved various ionisation modes El (Section 6.2.1), Cl (Section 6.2.2), DCI (Section 6.2.2.1), FAB (Section 6.2.4), FI (Section 6.2.5), FD (Section 6.2.6) and LD. Survey mass spectra obtained with soft ionisation methods (FI-MS, CI-MS) provide diagnostic overviews of chemical composition. The supplemental tandem (MS/MS) and atomic composition (AC-MS) techniques are used to make specific identifications of various organic ingredients. Direct analysis of polymer systems for more than a few thousand daltons has only just begun. Ionisation methods employed are FD, ESI and MALDI. Solid-probe ToF-MS (or DI-HRMS) is a breakthrough [188]. [Pg.412]

Mixtures can be analysed by using soft ionisation methods and hyphenated techniques such as gas chromatography MS, liquid chromatography MS and tandem MS (GC/MS, LC/MS, and MS/MS, respectively). [Pg.304]


See other pages where Tandem soft ionisation techniques is mentioned: [Pg.402]    [Pg.411]    [Pg.417]    [Pg.235]    [Pg.405]    [Pg.406]    [Pg.540]    [Pg.20]    [Pg.216]    [Pg.237]    [Pg.378]   
See also in sourсe #XX -- [ Pg.308 ]