System suitability parameters tailing factor

System suitability parameters with their respective acceptance criteria should be a requirement for any method. This will provide an added level of confidence that the correct mobile phase, temperature, flow rate, and column were used and will ensure the system performance (pump and detector). This usually includes (at a minimum) a requirement for injection precision, sensitivity, standard accuracy (if for an assay method), and retention time of the target analyte. Sometimes, a resolution requirement is added for a critical pair, along with criteria for efficiency and tailing factor (especially if a known impurity elutes on the tail of the target analyte). This is added to ensure that the column performance is adequate to achieve the desired separation. [Pg.389]

In addition to validation of the automation, full validation of the chromatographic procedure, as described in Chapter 12, should be conducted for late-phase methods. This should include specification of system suitability parameters to ensure that the performance obtained during method development and validation is maintained during routine use. The system suitability parameters may include specification of acceptable injection repeatability, criteria for resolution between critical pairs, maximum allowable tailing factors, and a means of verifying that the requisite sensitivity is obtained. As recommended by Vander Heyden et al., system suitability limits are best set following robustness tests. [Pg.369]

Figure 5.12. A summary report for system suitability testing documenting result i.d., precision, and other system suitability parameters (S/N ratio, tailing factors, plate count, and resolution). Note that means and precision of each table column can be automatically calculated. Figure 5.12. A summary report for system suitability testing documenting result i.d., precision, and other system suitability parameters (S/N ratio, tailing factors, plate count, and resolution). Note that means and precision of each table column can be automatically calculated.
System suitability should be based on criteria and parameters collected as a group that will be able to define the performance of the system. Some of the common parameters used include precision of repetitive injections (usually five or six), resolution (R), tailing factor (T), number of theoretical plates (N), and capacity factor ( ). [Pg.15]

For system suitability smdies, five replicate injections of mixed standard solutions were injected and parameters such as relative standard deviation of peak area, column efficiency, resolution, and tailing factors of the peaks were calculated. Results are shown in Table 3. [Pg.992]


See other pages where System suitability parameters tailing factor is mentioned: [Pg.389]    [Pg.487]    [Pg.528]    [Pg.244]    [Pg.275]    [Pg.760]    [Pg.671]    [Pg.163]    [Pg.235]    [Pg.173]    [Pg.196]    [Pg.439]    [Pg.191]    [Pg.579]   
See also in sourсe #XX -- [ Pg.439 ]