Standardization peak tailing

Because of some peak tailing, the number of theoretical plates was based on peak width at one-half peak height Ns5.S4 he pooled standard deviation (all temperatures) of retention time measurements (dfs34) was t 0.007 minutes. [Pg.210]

System suitability tests for chromatographic impurities procedures, such as precision, resolution factor, calibration standard, and tailing factor, should be considered as appropriate. In the presence of multiple peaks, a resolution factor between the two closest peaks should be proposed. For reliable quantitation, baseline resolution of the impurities will provide accurate measurement of the... [Pg.17]

All system suitability criteria for injection reproducibility, standard confirmation, resolution, sensitivity determination, and peak tailing must be met on each day. [Pg.174]

In another method, Pandey and Rathanand [35] used a combination of acetonitrile and water to analyze ezetimibe as a single compound. In contrast, Sistla et al. [36] reported that combination of water and acetonitrile produced peak tailing and poor resolution of ezetimibe with amoxicillin trihydrate as the internal standard. To overcome this, an ion pair reagent was added to the system, successfully reducing the peak tailing and increasing resolution between ezetimibe and amoxicillin trihydrate. [Pg.121]

One of the key advances in column technologies is the development of high-purity silica.1,9 In recent years, it has become a de facto industry standard for almost all new column offerings. This development stems from the realization that batch-to-batch reproducibility and peak tailing of basic solutes are mostly caused by acidic residual silanols. Figure 3.9 shows different types of silanols and their relative acidity. The worst culprits turned out to be the very acidic silanols adjacent to and activated by metallic oxides. Many older silica materials have high metallic contents (e.g., Spherisorb) and are extremely acidic. They often require the use of amine additives in the mobile phase (e.g., tri-ethylamine) to prevent adsorptive interaction with basic analytes. The inherent variations of these active (acidic) silanols are also responsible for the lack of batch-to-batch consistency of these acidic silica materials. [Pg.58]

Band broadening prevention strategies for SCCE are similar to those used in UHVCE and FCCE. Known formulas can be used for buffer components if little peak tailing is seen under standard CE separations. Electrostatic repulsion from channel walls that share the same electrical charge as the analytes can reduce broadening greatly. [Pg.754]

Pu5/92138 issued by AERE, Harwell (Hamilton et al. 1989). The isotope fractionation at each step is corrected by reference to the certified isotope ratios. The idea was further developed by Dubois et al. into a dynamic multidetection measurement mode (Dubois et al. 1989), which practically eliminates mass fractionation effects and possible mismatches of cup efficiencies with a 2-isotope internal standard and MIC/TIMS. According to the results presented, precisions and accuracies of 0.01 % are achievable with this procedure. An ultimate refinement has been introduced by performing total evaporation measurements with peak tailing correction in dynamic multicollection mode, using a MIC/TIMS with magnetic sector equipped with a dispersion quadrupole (Goldberg et al. 2002 Richter and Goldberg 2003). [Pg.2965]

A = 254 ran) using 40/60, 50/50, and 60/40 acetonitrile/water (6.25 mM sodium phosphate buffer at pH 3) mobile phases [1489], The k values for each analyte and each mobile phase were tabulated. Standards of 25 or 50 pg/mL (50 pL injected) were used. With base silica as the support, the effect of decreasing the concentration of analyte may become extremely important. The diagnostic chromatographic effects are increased peak tailing and longer overall retention times. [Pg.512]


See other pages where Standardization peak tailing is mentioned: [Pg.318]    [Pg.36]    [Pg.84]    [Pg.247]    [Pg.41]    [Pg.82]    [Pg.582]    [Pg.587]    [Pg.93]    [Pg.137]    [Pg.391]    [Pg.304]    [Pg.225]    [Pg.34]    [Pg.42]    [Pg.44]    [Pg.19]    [Pg.16]    [Pg.841]    [Pg.878]    [Pg.396]    [Pg.1355]    [Pg.196]    [Pg.447]    [Pg.1836]    [Pg.291]    [Pg.472]    [Pg.1828]    [Pg.150]    [Pg.1537]    [Pg.362]    [Pg.265]    [Pg.1334]    [Pg.1170]    [Pg.830]    [Pg.4]    [Pg.111]    [Pg.1423]    [Pg.1397]    [Pg.1420]    [Pg.291]    [Pg.86]    [Pg.111]   