Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Solution resistance reference electrode

Figure Bl.28.8. Equivalent circuit for a tliree-electrode electrochemical cell. WE, CE and RE represent the working, counter and reference electrodes is the solution resistance, the uncompensated resistance, R the charge-transfer resistance, R the resistance of the reference electrode, the double-layer capacitance and the parasitic loss to tire ground. Figure Bl.28.8. Equivalent circuit for a tliree-electrode electrochemical cell. WE, CE and RE represent the working, counter and reference electrodes is the solution resistance, the uncompensated resistance, R the charge-transfer resistance, R the resistance of the reference electrode, the double-layer capacitance and the parasitic loss to tire ground.
A typical Evans diagrams for the corrosion of a single metal is illustrated in Fig. 1.26a (compare with Fig. 1.23 for two separable electrodes), and it can be seen that the E -I and E -I curves are drawn as straight lines that intersect at a point that defines and (it is assumed that the resistance for the solution is negligible). E can of course be determined by means of a reference electrode, but since the anodic and cathodic sites are inseparable direct determination of /co by means of an ammeter is not... [Pg.93]

Accurate control of potential, stability, frequency response and uniform current distribution required the following low resistance of the cell and reference electrode small stray capacitances small working electrode area small solution resistance between specimen and point at which potential is measured and a symmetrical electrode arrangement. Their design appears to have eliminated the need for the usual Luggin capillary probe. [Pg.1125]

The reference electrode must be sited as close as possible to the D.M.E. so that the resistance of the solution between the two electrodes is reduced to a minimum, and the potentiostat then maintains the e.m.f. of the D.M.E.-reference electrode combination at the correct value. This arrangement has the further advantage that no sensible current is passed through the reference electrode... [Pg.610]

It should be pointed out that not all of the iR drop is removed by the potentiostatic control. Some fraction, called iRu (where Ru is the uncompensated solution resistance between the reference and working electrodes) will still be included in the measured potential. This component may be significantly large when resistive nonaqueous media are used, and thus may lead to severe distortion of the... [Pg.105]

On the basis of this argument, the mechanism for the current oscillation and the multilayer formation can be explained as follows. First note that U is kept constant externally with a potentiostat in the present case. In the high-current stage of the current oscillation, the tme electrode potential (or Helmholtz double layer potential), E, is much more positive than U because E is given hy E=U —JAR, where A is the electrode area, R is the resistance of the solution between the electrode surface and the reference electrode, andj is taken as negative for the reduction current. This implies that, even if U is kept constant in the region of the NDR, is much more... [Pg.244]

The cyclic voltammograms of these systems display quasi-reversible behavior, with AEv/v being increased because of slow electrochemical kinetics. Standard electrochemical rate constants, ( s,h)obs> were obtained from the cyclic voltammograms by matching them with digital simulations. This approach enabled the effects of IR drop (the spatial dependence of potential due to current flow through a resistive solution) to be included in the digital simulation by use of measured solution resistances. These experiments were performed with a non-isothermal cell, in which the reference electrode is maintained at a constant temperature... [Pg.384]

It must, however, be kept in mind that one cannot eliminate the fraction of the non-compensated solution resistance Rnc, which generates the ohmic drop iRnc. Unfortunately, the positioning of the reference electrode even closer to the working electrode (<2d) would cause current oscillations. [Pg.21]

Figure 6J3 Schematic representation of a Luggin capillary used for minimizing IR drop. Calculations of uncompensated solution resistance require a knowledge of the distance d between the reference tip of the capillary and the working electrode (depicted here as a dropping-mercury electrode (DME)). Figure 6J3 Schematic representation of a Luggin capillary used for minimizing IR drop. Calculations of uncompensated solution resistance require a knowledge of the distance d between the reference tip of the capillary and the working electrode (depicted here as a dropping-mercury electrode (DME)).
An alternative, and usually superior, method involves using a Luggin capillary (see Figure 6.33). At heart, a Luggin capillary is a reference electrode connected to a tube filled with KCl solution at high concentration, so the resistance of the internal solution is tiny. In addition, the capillary is slender in order to minimize disruption of the flux at the electrode surface, and its tip is narrow to ensure that none of the internal KCl solution seeps into the analyte solution. [Pg.193]

The differential technique described under (a) has an advantage in removal of the liquid-junction potential and of mechanical faults often encountered in work with reference electrodes of the second kind. The procedure described under (b) suppresses the potential fluctuations, but difficulties can arise from the very high resistance of a cell containing two ISEs. A differential amplifier was designed for this prupose [15]. The two ISEs used can also exhibit different slopes electrode membranes were therefore prepared by cutting a single crystal into two halves, where each half contains a chaimel for passage of the solution and functions as an ISE [163]. [Pg.115]

In the bypass position, the carrier solution flows through the bypass loop and across the ISFET. The sample is injected into the sampling valve and is introduced into the carrier solution. The bypass loop has a high hydrodynamic resistance and thus the solution proceeds to the detector. The reference electrode is always immersed only in the carrier solution and is electrically connected with the ISFET through the solutioa The apparatus is regularly calibrated by K, Ca and pH standard solutions. [Pg.129]

If the three-electrode instrument is equipped with an iR-drop compensator, most of the iT-drop caused by the solution resistance can be eliminated. However, in order to minimize the effect of the iT-drop, a Fuggin capillary can be attached to the reference electrode with its tip placed close to the indicator electrode. Moreover, for a solution of extremely high resistance, it is effective to use a quasi-reference electrode of a platinum wire (Fig. 8.1(a)) or a dual-reference electrode (Fig. 8.1(b)), instead of the conventional reference electrode [12]. [Pg.225]

The second factor is associated with the fact that all electrolyte solutions exhibit finite resistance to the flow of current. Thus, the potential that is measured (Em eas) between the working and reference electrodes consists of two contributors, the real thermodynamic potential Cereal) and that arising from uncompensated solution resistance (IRU)... [Pg.157]

Figure 7.1 (A) Typical controlled-potential circuit and cell OA1, the control amplifier OA2, the voltage follower (Vr = Er) OA3, the current-to-voltage converter. (B) Equivalent circuit of cell Rc, solution resistance between auxiliary and working electrodes Ru, solution resistance between reference and working electrodes, Rs = Rc + Ru and Cdl, capacitance of interface between solution and working electrode. (C) Equivalent circuit with the addition of faradaic impedance Zf due to charge transfer. Potentials are relative to circuit common, and working electrode is effectively held at circuit common (Ew = 0) by OA3. Figure 7.1 (A) Typical controlled-potential circuit and cell OA1, the control amplifier OA2, the voltage follower (Vr = Er) OA3, the current-to-voltage converter. (B) Equivalent circuit of cell Rc, solution resistance between auxiliary and working electrodes Ru, solution resistance between reference and working electrodes, Rs = Rc + Ru and Cdl, capacitance of interface between solution and working electrode. (C) Equivalent circuit with the addition of faradaic impedance Zf due to charge transfer. Potentials are relative to circuit common, and working electrode is effectively held at circuit common (Ew = 0) by OA3.
One of the first observations one is likely to make when carrying out low-temperature voltammetric measurements is that the effects of solution iR drop that may have been scarcely noticeable at room temperature are suddenly alarmingly pronounced. The iR drop, of course, is governed by the cell current and the effective resistance between the working electrode and the Luggin capillary of the reference electrode. For example, the solution resistance for an embedded circular disk electrode of radius r with a distant reference electrode is given by Equation 16.13, where p is the resistivity of the solution. [Pg.506]


See other pages where Solution resistance reference electrode is mentioned: [Pg.930]    [Pg.14]    [Pg.1037]    [Pg.124]    [Pg.1939]    [Pg.467]    [Pg.51]    [Pg.1008]    [Pg.1244]    [Pg.105]    [Pg.129]    [Pg.613]    [Pg.78]    [Pg.303]    [Pg.224]    [Pg.178]    [Pg.216]    [Pg.171]    [Pg.37]    [Pg.13]    [Pg.100]    [Pg.416]    [Pg.699]    [Pg.162]    [Pg.264]    [Pg.179]    [Pg.23]    [Pg.24]    [Pg.143]    [Pg.143]    [Pg.172]    [Pg.195]    [Pg.269]    [Pg.387]    [Pg.497]   
See also in sourсe #XX -- [ Pg.244 ]




SEARCH



Electrode resistance

Electrode solution

Reference electrode solution

Reference electrodes

Reference resistance

Resistive electrodes

Solution resistance

© 2024 chempedia.info