Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Signal-to-noise ratio s

The base of this estimation is the signal to noise ratio. The lowest signal to noise ratio S/N which is necessary as a minimum to discern a signal from noise is S/N = 2 1 (4). Referring to the limiting values for the granularity Oj, of the film system classes the smallest density difference AD of an defect which would just be visible should be at least two times greater than On. [Pg.551]

The intensity of signal from the analyte compared to that from the noise is termed the signal-to-noise ratio (S/N). This is used by the analyst to determine, in the first instance, whether a detector signal can be said to be real , and therefore whether an analyte is present, and then to calculate the accuracy and precision with which that analyte can be qnantified. [Pg.41]

From Figure 3 it is apparent that when both benzene species are present the signal to noise ratio (S/N) of the spectrum is greatly reduced. The spectra in Figure 3 are not normalized but displayed with the same vertical height. Thus the "observed" reduction in S/N... [Pg.489]

The minimum detectable level is estimated with the dinifroaniline signal-to-noise ratios (S/N). With fortification levels between 0.2 and 0.5mgkg the recovery of trifluralin from plant matrices is 70-99% with the LOD/LOQ being 0.005 mg kg according to the analytical method of the Ministry of the Environment, Japan. In multiresidue analysis by GC/NPD, the percent recoveries of pendimethalin from each crop with a fortification level of 0.25 mg kg were brown rice 70, potato 70, cabbage 80, letmce 89, carrot 84, cucumber 64, shiitake 74, apple 76, strawberry 99, and banana 99%. The LOD for each sample was 0.01 mg kg for pendimethalin. In residue analysis by GC/ECD, recoveries of the majority of dinifroaniline herbicides from fortified samples of carrot, melon, and tomato at fortification levels of 0.04—0.10 mg kg ranged from 79 to 92%. The LODs were benfluralin 0.001, pendimethalin 0.002 and trifluralin 0.001 mg kg for the GC/ECD method. ... [Pg.394]

The absolute strength of the MRI signal depends on many factors including the number of nuclear spins in an imaging voxel. Thus, there is a tradeoff between image resolution and signal-to-noise ratio, S/N. For the commonly used granular... [Pg.495]

As shown in Sect. 7.1, signal-to-noise ratio S/N can be used to characterize the precision of analytical methods. Noise is a measure of the uncertainty of dynamic blank measurements (of the background ). [Pg.232]

But the main advantage of the SNR concept in modern analytical chemistry is the fact that the signal function is recorded continuously and, therefore, a large number of both background and signal values is available. As shown in Fig. 7.9, the principles of the evaluation of discrete and continuous measurement values are somewhat different. The basic measure for the estimation of the limit of detection is the confidence interval of the blank. It can be calculated from Eq. (7.52). For n = 10 measurements of both blank and signal values and a risk of error of a = 0.05 one obtains a critical signal-to-noise ratio (S/N)c = fo.95,9 = 1.83 and a = 0.01 (S/N)c = t0.99,9 = 2.82. The common value (S/N)c = 3 corresponds to a risk of error a = 0.05... 0.02 in case of a small number of measurements (n = 2... 5). When n > 6, a... [Pg.232]

Larger frequency shifts lead to more sensitive structural discriminations. In the IR, the (i-shcct amide I is distinctively lower in frequency than other structural types, particularly for aggregates, a form often seen in unfolding experiments. However, owing to the high signal-to-noise ratio (S/N) of Fourier transform IR (FTIR), one can detect components having smaller frequency separations. This effective resolution... [Pg.136]

To simultaneously maximize signal-to-noise ratio (S/N) and minimize distortion, it is best to adjust the microwave power, as measured by the power... [Pg.12]

As we have seen in Section 9.5.3, in the case of resistance thermometry, the signal produced by a low-temperature thermometer is very low (microvolt range). Low-pass filters are not sufficient to narrow the detection bandwidth in order to get a suitable signal to noise ratio (S/N). Bandpass filters are needed. The most commonly used method is the synchronous demodulation, usually simply called lock-in technique, as shown in the block diagram of Fig. 10.7. [Pg.247]

In the CL detection method, the excitation of a molecule is achieved via a chemical reaction that is generally an oxidation process. That is, an exciting light source is not required thus, the CL is not accompanied by any scattering light and source instability. This permits a large signal-to-noise ratio (S/N), which finally provides an increase in sensitivity. [Pg.394]

LOD is defined as the lowest concentration of an analyte that produces a signal above the background signal. LOQ is defined as the minimum amount of analyte that can be reported through quantitation. For these evaluations, a 3 x signal-to-noise ratio (S/N) value was employed for the LOD and a 10 x S/N was used to evaluate LOQ. The %RSD for the LOD had to be less than 20% and for LOQ had to be less than 10%. Table 6.2 lists the parameters for the LOD and LOQ for methyl paraben and rhodamine 110 chloride under the conditions employed. It is important to note that the LOD and LOQ values were dependent upon the physicochemical properties of the analytes (molar absorptivity, quantum yield, etc.), methods employed (wavelengths employed for detection, mobile phases, etc.), and instrumental parameters. For example, the molar absorptivity of methyl paraben at 254 nm was determined to be approximately 9000 mol/L/cm and a similar result could be expected for analytes with similar molar absorptivity values when the exact methods and instrumental parameters were used. In the case of fluorescence detection, for most applications in which the analytes of interest have been tagged with tetramethylrhodamine (TAMRA), the LOD is usually about 1 nM. [Pg.174]

Fourier transform spectroscopy technology is widely used in infrared spectroscopy. A spectrum that formerly required 15 min to obtain on a continuous wave instrument can be obtained in a few seconds on an FT-IR. This greatly increases research and analytical productivity. In addition to increased productivity, the FT-IR instrument can use a concept called Fleggetts Advantage where the entire spectrum is determined in the same time it takes a continuous wave (CW) device to measure a small fraction of the spectrum. Therefore many spectra can be obtained in the same time as one CW spectrum. If these spectra are summed, the signal-to-noise ratio, S/N can be greatly increased. Finally, because of the inherent computer-based nature of the FT-IR system, databases of infrared spectra are easily searched for matching or similar compounds. [Pg.150]

FIGURE 7 Comparison of two peaks with signal-to-noise ratio (S/N) of 50 and 10. The random distribution of noise in the noisy peak controls the variation of the peak area measurement. [Pg.269]


See other pages where Signal-to-noise ratio s is mentioned: [Pg.472]    [Pg.2955]    [Pg.107]    [Pg.391]    [Pg.321]    [Pg.112]    [Pg.86]    [Pg.426]    [Pg.831]    [Pg.78]    [Pg.293]    [Pg.359]    [Pg.379]    [Pg.139]    [Pg.95]    [Pg.376]    [Pg.952]    [Pg.301]    [Pg.228]    [Pg.393]    [Pg.277]    [Pg.67]    [Pg.74]    [Pg.373]    [Pg.578]    [Pg.16]    [Pg.149]    [Pg.86]    [Pg.423]    [Pg.409]    [Pg.125]    [Pg.286]    [Pg.228]   
See also in sourсe #XX -- [ Pg.41 , Pg.54 , Pg.64 , Pg.67 , Pg.104 , Pg.128 , Pg.129 , Pg.143 , Pg.161 , Pg.242 , Pg.247 , Pg.284 , Pg.303 , Pg.336 , Pg.378 , Pg.391 , Pg.464 ]




SEARCH



Signal noise

Signal-to-noise

Signal-to-noise ratio

Signal/noise ratio

© 2024 chempedia.info