Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Signal drift

In flame photometry, signal drift and lamp flicker require that one or a few unknowns be bracketed by calibrations. Here, independent measurements on the same solutions means repeating the whole calibration and measurement cycle. [Pg.22]

Detection in 1C is strictly connected with the nature of eluents (composition, concentration), analytes and the sensitivity required. The ideal characteristics of a chromatographic detector are essentially the following (1) high sensitivity, (2) low cell dead volume, (3) linear relationship between concentration and signal, (4) stable and low background noise, (5) high speed of response, and (6) no signal drift. [Pg.406]

IDMS is based on measurements of masses and isotope ratios only. Some important advantages, compared with other calibration strategies, such as external calibration and standard additions, are that instrumental instabilities such as signal drift and matrix effects will have no influence in the final concentration in the sample, high accuracy and small measurement uncertainties are enabled, possible loss of substance of the isotope-diluted sample will have no influence on the final result and there is no need to resort to an external instrumental calibration or standard additions to the sample. [Pg.32]

Catasus et al. [67] studied two types of neural networks traditional multilayer perceptron neural networks and generalised regression neural networks (GRNNs) to correct for nonlinear matrix effects and long-term signal drift in ICP-AES. [Pg.272]

Optical fibers are now used extensively to bring the analyzing light to and from the sample for optical spectrometers. What was not known about optical fibers in early implementations was the problems they can bring to spectroscopy. In early applications of fiber optics, they were found to have limited lifetime, cause signal drift and attenuation, and cause process contamination.4 Fiber optics were developed for the on/off application... [Pg.7]

The space-charge definition within the dielectric (PVC) membrane changes with time. Such changes are a direct consequence of membrane exposure to sample solution and are controlled by membrane formulation. The two factors of consequence to pragmatic design (not accounted for by equations 1-5) are the transient response of the membrane and signal drift due to membrane dielectric constant and interfacial potential changes. [Pg.271]

The photomultiplier detects both the thermal emission from the determinant and also any other atomic or molecular emission from either concomitant elements present in the sample or from the flame itself. Figure 8, for example, shows a typical section of a flame emission spectrum. While it is possible for some determinations by FES to work at a single fixed wavelength, as in flame AAS, it is advisable, at least initially, to scan the emission spectrum in the vicinity of the wavelength of interest to confirm the absence of spectral interferences. In any event, regular re-zeroing and aspiration of an appropriate standard to check for signal drift is essential. [Pg.22]

The ionophore must have high lipophilicity so that the leaching rate, and thus the signal drift, is kept to a minimum.5... [Pg.326]

In the case of solutions with high salt contents, the danger for clogging and signal drift is high as compared with other nebulizers (Fig. 41) [109]. The maximum admissible salt content is 20-40 g/L but it varies considerably along with the type of salt involved. It is particularly high for easily hydrolyzable compounds such as aluminum salts and phosphates. [Pg.92]

For CL spectra we found it critical to obtain a "dark" spectrum immediately before automated initiation of the reaction by injection of the final reagent and acquisition of the CL spectrum. Subtraction of the dark spectrum from the CL spectrum compensates for dark signal drift. Also, signal averaging of many spectra (one from each reaction run) is often essential to achieve a reasonable S/N. [Pg.166]

Current problems with photobleaching and signal drift do not severely limit the performance of these sensors. In fact, they are sometimes not intended to provide extremely precise absolute analytical data. Rather, it is the relative change of a parameter that has to be recognized as soon as possible during the course of an operation on a critically ill patient, or a fairly quick indication of whether abnormal clinical values require a more reliable assay in a clinical laboratory. [Pg.241]


See other pages where Signal drift is mentioned: [Pg.101]    [Pg.111]    [Pg.249]    [Pg.286]    [Pg.310]    [Pg.152]    [Pg.319]    [Pg.306]    [Pg.13]    [Pg.244]    [Pg.158]    [Pg.655]    [Pg.271]    [Pg.272]    [Pg.487]    [Pg.305]    [Pg.46]    [Pg.207]    [Pg.571]    [Pg.378]    [Pg.378]    [Pg.90]    [Pg.306]    [Pg.3684]    [Pg.400]    [Pg.276]    [Pg.87]    [Pg.136]    [Pg.291]    [Pg.159]    [Pg.14]    [Pg.430]    [Pg.459]    [Pg.196]    [Pg.324]    [Pg.54]    [Pg.226]   
See also in sourсe #XX -- [ Pg.310 ]

See also in sourсe #XX -- [ Pg.157 , Pg.160 , Pg.414 , Pg.423 , Pg.433 ]




SEARCH



Drift

Drifting

© 2024 chempedia.info