Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Signal comparisons

The signal comparison function has been impieinented and tested during a steam generator tube inspection. A formal test in the real-life context was successfully done with a simple rule (parameters phase and amplitude) based on the central frequency for the distance signal. [Pg.1026]

Extraction of Electron Transfer Kinetics from Cyclic Voltammetric Signals. Comparison with Other Techniques... [Pg.44]

There are two kinds of comparison relationships comparison by similarity and comparison by restatement. Words that signal comparison are words such as likewise, similarly, and and itself. Phrases that introduce comparisons include just as, as as, for example, as shown, and as illustrated by. An example of a comparison... [Pg.105]

Figure 19.7 Monovalent and trivalent ASAXS signal comparison. This figure illustrates differences in shape between the ASAXS signals for monovalent Rb around DNA and trivalent cobalt hexammine around DNA. The curves have been scaled to match at the lowest q to enable shape comparison. The increase in high-angle scatter associated with the trivalent relative to monovalent ion signal, suggests that the trivalent ions are more tightly localized to the DNA. Figure 19.7 Monovalent and trivalent ASAXS signal comparison. This figure illustrates differences in shape between the ASAXS signals for monovalent Rb around DNA and trivalent cobalt hexammine around DNA. The curves have been scaled to match at the lowest q to enable shape comparison. The increase in high-angle scatter associated with the trivalent relative to monovalent ion signal, suggests that the trivalent ions are more tightly localized to the DNA.
Table 11.2. Signal Comparison for Various Microscope Objectives on Dilor V-T... Table 11.2. Signal Comparison for Various Microscope Objectives on Dilor V-T...
Raber J, Koob GF, Bloom FE. Interleukin-2 induces corticotropin-releasing factor (CRF) release from the amygdala and involves a nitric oxide-mediated signaling comparison with the hypothalamic response. J Pharmacol Exp Ther 1995 272 815-824. [Pg.418]

Figure 12-5 illustrates a second example using two transmitters. However, only one transmitter is part of the safety instrumented function equipment set. For this application signal comparison to achieve high diagnostics is done in the Safety PLC. It is also recommended that the BPCS transmitter be hardwired to the Safety PLC. This hardwiring could be via a signal splitter as shown. [Pg.183]

The analog signal comparison can be performed between two or more transmitters that are part of the SIF or between a SIF transmitter and a DCS transmitter (measuring the same process variable). In order to take credit for the comparison it needs to be performed in the SIS logic solver. The software in the Safety Instrumented System is safety critical rated. This is not the case for the DCS software. [Pg.183]

Two different terms can be found in the literature when dealing with qualitative analysis although they present slightly different connotations. The word detection is normally used to refer to a chemical measurement process for qualitative purposes, opposed to determination, which is reserved for quantitative analysis. Identification is usually employed for qualitative analysis aimed at recognizing the analyte (or its reaction product) from some chemical or physicochemical properties. As it entails the use of a standard for signal comparison, it is more appropriate as an alternative to qualitative analysis from a metrological point of view. [Pg.3973]

All band areas are set to a single value for the purpose of signal comparison and/or correlation techniques (Fig. 4). [Pg.499]

The coupling constant, 8 cycles sec. l, is typical of non-equivalent aromatic protons positioned ortho to each other. The polymer from reaction (6) showed a more con5)licated spectrum, the AB quartet being distorted and several additional peaks appeared on the high field sides of the main resonance signals. Comparison with the spectra of model con KJunds and eventually with the spectra of co-polymers made by the polyether synthesis (8) [18] to contain known quantities of the ortho, para-repeat unit, IV,... [Pg.176]

TABLE 2.6. MDSC Phase-Corrected Signal Comparison for TAI Modules... [Pg.177]

Nuclear magnetic resonance (NMR) spectroscopy is the most informative analytical technique and is widely applied in combinatorial chemistry. However, an automated interpretation of the NMR spectral results is difficult (3,4). Usually the interpretation can be supported by use of spectrum calculation (5-18) and structure generator programs (8,12,18-21). Automated structure validation methods rely on NMR signal comparison using substructure/ subspectra correlated databases or shift prediction methods (8,15,22,23). We have recently introduced a novel NMR method called AutoDROP (Automated Definition and Recognition of Patterns) to rapidly analyze compounds libraries (24-29). The method is based on experimental data obtained from the measured ID or 2D iH,i C correlated (HSQC) spectra. [Pg.123]


See other pages where Signal comparisons is mentioned: [Pg.130]    [Pg.124]    [Pg.260]    [Pg.32]    [Pg.301]    [Pg.189]    [Pg.260]    [Pg.37]    [Pg.41]    [Pg.183]    [Pg.242]    [Pg.170]    [Pg.180]    [Pg.796]    [Pg.235]    [Pg.40]   
See also in sourсe #XX -- [ Pg.597 , Pg.598 ]




SEARCH



Comparison of noise and signal

Extraction of Electron Transfer Kinetics from Cyclic Voltammetric Signals. Comparison with Other Techniques

Signal-to-Noise Ratio Comparisons

© 2024 chempedia.info