Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Signal acquisition

Depending on the isotopes monitored and the nature of the sample, several types of mass bias correction models can be used when coupling chromatography to MC-ICP-MS. Some studies report the use of internal mass bias correction with an exponential law when an invariant isotope ratio is available (as is the case, for example, for Sr or Nd) [27, 29-31]. It is the simplest model, as merely another isotope ratio of the element of interest is used for the correction. [Pg.504]

It is also possible to use external mass bias correction. In this case, another element, which is absent in the sample and which is located in the same mass range as the analyte, is simultaneously and continuously introduced [33]. In this case the exponential law is preferentially used for correction of the measured ratios [20]. [Pg.504]

Finally, another approach to correct for the mass bias is the standard-sample-standard bracketing technique [13, 15, 28, 34], in which a certified isotopic standard is measured before and after each sample. Generally, for isotopic analysis in species, this type of correction should be applied species by species, which means that the standard must contain the element of interest in the same species as the sample. [Pg.504]

To improve the precision and accuracy of results, both external mass bias correction and standard-sample-standard bracketing techniques can be used simultaneously. This has been done, for example, for Hg isotopic studies in elemental species [13, 15]. [Pg.504]


REAL SIGNALS ACQUISITION or AE SIGNALS EROM THE FIELD (MAXIMUMll TRANSDUCERS)... [Pg.72]

Such requirements are meant to change the usual configurations and technologies usually associated with photothermal set-up. These changes mainly affect the IR detection devices, the optical components associated with the excitation and detection flux and the signal acquisition hardware and software. Figure 9 presents a sketch of the so-called pre-industrial demonstrator built from those different improvements. [Pg.397]

Set the number of decimal places displayed to simulate the action of digital signal acquisition 1 would mean the result of a measurement is clipped to the first decimal. [Pg.379]

Approximately 1 g polymer and 0aQ6 M Cr(acac). were dissolved in CDCl. to prepare solutions for ySi and JC NMR spectroscopy. NMR spectra were run on a Varian XL-200 FT-NMR instrument. To aid in obtaining quantitative data, the solution was doped with 0.06 M chromium acetylacetonate [Cr(acac) )] to remove possible signal artifacts resulting from long spin-lattice relaxation times (T s) and tt> nucleay Overhauser effect, well-known features associated with 3Si and JC NMR spectroscopy. This permits quantitative signal acquisition. From the literature (16) and additional work done in this laboratory, it was expected that Cr(acac) would be an inert species. A solution of HMDZ (2.04 g, 12.67 mmole),... [Pg.157]

At present, increasing accuracy and precision are achieved in separation performance and signal acquisition for complex 2D separations. However, the complexity of the plethora of data obtained requires proper signal processing procedures for a complete extraction of all of the analytical information. [Pg.88]

The principle of operation of RF transmission and signal acquisition has been described in the Refs. 1,2. Flere we briefly overview the architecture of the spectrometer, to the extent that is sufficient for the discussion of the system customization described below. [Pg.359]

Most clinical examinations apply robust spin-echo or fast spin-echo sequences. These types of sequences provide tissue contrast changes by variation of the chosen repetition time TR (time interval between succeeding RF excitations) and echo time TE (time delay between RF excitation and signal acquisition). [Pg.11]

Figure 3.9 Crab sampling in different practical manifestations (aggregate material), to which must be added all PAT sensor alternatives in Figures 3.12-3.14 dealing with fluids and slurries. All principal characteristics are identical for physical sampling and signal acquisition there is only sampling from some part of the stream cross section. This creates IDE and therefore cannot result in representative sampling. Figure 3.9 Crab sampling in different practical manifestations (aggregate material), to which must be added all PAT sensor alternatives in Figures 3.12-3.14 dealing with fluids and slurries. All principal characteristics are identical for physical sampling and signal acquisition there is only sampling from some part of the stream cross section. This creates IDE and therefore cannot result in representative sampling.
Figure 3.12 Generic illustration of incorrect valve designs for the case of horizontal flow. In these examples only part of the cross section of the stream is sampled all of the time intermittent sampling/signal acquisition is in no way better, see Figure 3.11. The sampling valves shown there offer no counteraction for gravitational segregation and/or flow differentiation effects. All solutions shown are structurally incorrect. A specific valve opening design alone does not eliminate IDE. Figure 3.12 Generic illustration of incorrect valve designs for the case of horizontal flow. In these examples only part of the cross section of the stream is sampled all of the time intermittent sampling/signal acquisition is in no way better, see Figure 3.11. The sampling valves shown there offer no counteraction for gravitational segregation and/or flow differentiation effects. All solutions shown are structurally incorrect. A specific valve opening design alone does not eliminate IDE.
Fluorescence from pharmaceutical capsule shells and tablet coatings has hindered measurement of their composition by Raman spectroscopy. By switching from the conventional backscattering mode to a transmission mode, Matousek et al. demonstrated that fluorescence could be eliminated in many instances [8]. Backscattering- and transmission-mode Raman spectra of several samples are shown in Figure 7.5. Each spectrum was acquired in 10s with 80mW 830-mn laser power. Matousek et al. also speculate that signal acquisition times could be relatively easily shortened to well below 0.1 s when the transmission mode is combined with optimized optics [8]. [Pg.210]

In order to understand this, consider that in an FFC experiment the amplitudes of the acquired signals are approximately proportional to the signal acquisition field Ba- For example, in the case of the basic prepolarized sequence (described in Section VIII.C.), one can show (65) that... [Pg.433]

As shown in Fig. 22, the resulting procedure, referred to as a multi-block experiment, produces a two-dimensional data set, such as an array of FIDs (its exact nature depends upon the signal acquisition method). The data of each x-block are then reduced to a single quantity, S(t) which should be proportional either to the total sample magnetization Ma(x) or to one of its components. Since the vertical scale of the relaxation curve is irrelevant, we can identify S(t) with Ma(x) at the exact time of detection (usually just after the first excitation pulse). [Pg.442]

In general, the signal acquisition process provides us with more data than needed. To extract relaxation parameters, we need a single value M(x) for each X setting, but we generally acquire a whole array of values. Clearly, some kind of data-reduction process must be implemented before the acquired signals can be used in the way we intend it to be. Like the NMR excitation techniques, the data reduction process can be exploited to enhance or suppress particular signal components. [Pg.454]

Modern instruments usually offer an on-board choice between a quadrature phase detector and some kind of diode or square detector. The latter, however is mostly used just for instrument setup (probe tuning, etc.) while signal acquisition is done almost exclusively by the phase detector. [Pg.455]


See other pages where Signal acquisition is mentioned: [Pg.68]    [Pg.55]    [Pg.316]    [Pg.462]    [Pg.224]    [Pg.1078]    [Pg.514]    [Pg.11]    [Pg.16]    [Pg.144]    [Pg.392]    [Pg.430]    [Pg.110]    [Pg.355]    [Pg.359]    [Pg.362]    [Pg.368]    [Pg.390]    [Pg.81]    [Pg.141]    [Pg.114]    [Pg.110]    [Pg.55]    [Pg.121]    [Pg.48]    [Pg.59]    [Pg.70]    [Pg.440]    [Pg.440]    [Pg.208]    [Pg.389]    [Pg.200]   
See also in sourсe #XX -- [ Pg.206 ]




SEARCH



© 2024 chempedia.info