Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Linear concentration range

Instrument makers nonetheless provide this less-than-useful information, but hardly anybody recognizes as the outflow of the wide calibration range, the linear concentration-to-signal transfer function, and the excellent repeatability. [Pg.6]

Instrumental response ratios (tebuconazole/i NsJtebuconazole) versus concentration of tebuconazole present should be proven in solvent and each matrix analyzed up to the highest undiluted final sample extract concentration expected. Once proven linear, final sample extract residues found to lie above the range of linearity are to be diluted prior to addition of IS solution and re-analyzed. [Pg.1239]

The peak symmetry, resolution, and detector response are directly dependent on the concentration of the sample. As the concentration of a sample increases, the retention time, separation, and peak symmetry generally decrease. These phenomena are due to isotherm nonlinearity. The detector response may also be nonlinear above or below certain concentrations. In some cases, small amounts of a dilute component are irreversibly adsorbed to the column, leading to reduced recovery. Above some concentration, the response of any detector will cease to be linear. The UV-VIS is one of the most linear detectors, generally exhibiting at least three decades of linearity, while RI, electrochemical, and fluorimetric detectors have a markedly narrower range of linearity. [Pg.153]

In the range of linear adsorption behaviour, whatever the number of site types (see Section 2.3.1 for the merging of parameters of two sites), the surface concentration F is related to via an effective linear coefficient, Ah, while the first-order internalisation processes can also be described by an effective first-order constant, k. Thus, equation (39) can be recast, for instance, in terms of r as ... [Pg.172]

The data in Figures 2.17 and 2.18 are displayed in terms of the dimensionless centre-to-centre separation of particles, i.e. r/2a — (2a + h)/2a. This has been done to illustrate another important point the range of linear elastic response. In a concentrated system, which is showing solid-like or elastic responses, the structure has to be able... [Pg.57]

Clinically relevant drug concentrations Linear absorption characteristics at clinically relevant concentrations should be demonstrated. Lack of effect of concentration on Peff determined in a range of physiologically relevant concentrations can be used to demonstrate lack of a saturable absorption mechanism (i.e. carrier mediated). Concentrations used in experiment should represent clinically relevant concentrations [132]... [Pg.63]

Protein A chromatography yields inununoglobulins in very concentrated form. Therefore, the absorbance of the solution should not be measured directly. Dilute the sample into PBS, and take areading using PBS as a blank. The optical density of the diluted sample should not be above 1-1.5 to fall within the range of linearity. [Pg.32]

After the administration of a drug, its concentration in plasma rises, reaches a peak, and then declines gradually to the starting level, due to the processes of distribution and elimination (p. 46). Plasma concentration at a given point in time depends on the dose administered. Many drugs exhibit a linear relationship between plasma concentration and dose within the therapeutic range (dose-linear kinetics (A) note different scales on ordinate). However, the same does not apply to drugs whose elimination processes are already sufficiently activated at therapeutic plasma levels so as to preclude further proportional increases in the rate of elimination when the concentration is increased further. Under these conditions, a smaller proportion of the dose administered is eliminated per unit of time. [Pg.68]

Upon use of structurally modified variants as internal standards for the particular analytes, the relative quantificahon of oligonucleotides, peptides, and small proteins was demonstrated [44]. The potential of the ILM to allow quantitative analyses of peptides without the use of internal standards was presented recently [43]. Linear correlahons between peptide amount and signal intensities could be found upon applicahon of increased matrix-to-analyte ratios between 25,000 and 250,000 (mokmol). The dynamic range of linearity thus spanned one order of magnitude. Unfortunately, the importance of the M/A ratio prevents the use of this method in samples with unknown orders of concentration, for example, in a proteomics environment. On the other hand, the method is applicable for the screening of enzyme-catalyzed reactions because the starting concentrahons of the peptides are generally known in such assays. [Pg.391]

For the determination of potency assay of a drug substance or a drug product, the usual range of linearity should be +20% of the target or nominal concentration. For the determination of content uniformity, it should be 30% of the target or nominal concentration. Figure 3 illustrates the linearity of a sample set of data. [Pg.735]

Range. Ideally, linearity should be established from 50% of the ICH reporting limit to the nominal concentration of drug substance in the sample solution (for area percent method). If the linearity does not support such a wide range of concentration, determine the linearity from 50% of the ICH reporting level to 150% of the proposed shelf life specifications of the related substance (for the high-low and external standard methods) as a minimum. This will ensure a linear response for related substances at all concentration levels to be detected during stability. [Pg.39]

Linear range concentration range over which calibration curve is linear (Figure 4-12)... [Pg.84]

Range concentration range over which linearity, accuracy, and precision meet specifications for analytical method. ... [Pg.84]

Recently, dopamine and catechol were detected using voltammetric MIP chemosensors [138]. Here, again, the chemosensors could determine dopamine and catechol in the presence of excess of interfering compounds, such as phenol, resorcinol, hydroquinone, serotonine, ascorbic acid, etc. For catechol, LOD was 228 nM and the linear concentration range was linear from 228 nM to 144 pM. [Pg.251]

Several electrolytic-conductivity detectors are produced (Table 3.5). The Laboratory Data Control Model 701 Conducto Monitor (Fig.3.59) may be operated in either a differential mode or an absolute mode. It provides direct readout in units of specific conductance and differences as small as 0.01% in the differential mode between the carrier and the carrier plus solute can be measured. The dynamic range of linearity is 0.01-100,000 pSl 1 /cm. The detector can function in solvents ranging from distilled water to concentrated salt solutions without the necessity of changing the cell. The volume of the cell is 2.5 pi, and the nominal cell constant is 20 cm-1. This type of detector is of use mainly in high-speed ion-exchange chromatography for the detection of ionic species. [Pg.99]

Since at the beginning the microcystin content in the real samples is unknown, pure and 100-fold diluted solutions are analysed in order to obtain, at least for one of the two dilutions, inhibition values into the linear range. Concentration values (pg L-1) are then converted to toxin contents (pgg-1 dry weight). [Pg.1106]


See other pages where Linear concentration range is mentioned: [Pg.161]    [Pg.255]    [Pg.306]    [Pg.48]    [Pg.14]    [Pg.350]    [Pg.282]    [Pg.126]    [Pg.279]    [Pg.147]    [Pg.366]    [Pg.60]    [Pg.480]    [Pg.348]    [Pg.134]    [Pg.386]    [Pg.244]    [Pg.39]    [Pg.14]    [Pg.122]    [Pg.65]    [Pg.14]    [Pg.99]    [Pg.111]    [Pg.260]    [Pg.695]    [Pg.226]    [Pg.598]    [Pg.975]    [Pg.226]    [Pg.18]    [Pg.33]    [Pg.61]    [Pg.157]   
See also in sourсe #XX -- [ Pg.121 ]




SEARCH



Concentration range

Linear range

Linearity range

© 2024 chempedia.info