Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Packed columns separation time

Current commercial silica-based columns have two important characteristics (1) they can produce efficiency similar to that of columns packed with 3.5 /tm particles and (2) they typically produce a pressure drop of half that caused by a column packed with 5 /tm particles.35 Monolithic columns have been shown to exhibit flat van Deemter curves, resulting in little loss of efficiency at high flow rates.36 As a result, high-throughput separations on conventional HPLC instruments can be achieved by increasing flow rate up to nine times (up to 9 ml/min) the usual rate in a conventional packed column. Cycle times for HPLC analysis as short as 1 min (injection-to-injection) have been reported by users of monolithic columns. Additional benefits of monolithic columns cited include... [Pg.257]

Packed column separation and analysis is certainly valuable for many routine tests and even very useful and necessary at times. However, it cannot meet the analytical requirements of many of today s investigations. For example, if the sample size is a limiting and critical factor, as in blood testing or other biological samples, or if complex mixtures are of interest, the packed column will in most cases fail to generate satisfactory results. [Pg.724]

A more complex sample containing many more (at least five times as many) components can be separated on a capillary column than in an equivalent packed column analysis time alternatively longer analysis times can be used to obtain the maximum separating capability for complex samples. [Pg.185]

Figure 2.123 Simulation of a packed column separation performed on a 30 m capillary column. Carrier gas has been changed from nitrogen to helium and compared with a 3% OV17 packed column, the 5% phenyl-methylpolysiloxane capillary column separates four previously unresolved peak pairs in a similar run time. Column 30 m TRACE... Figure 2.123 Simulation of a packed column separation performed on a 30 m capillary column. Carrier gas has been changed from nitrogen to helium and compared with a 3% OV17 packed column, the 5% phenyl-methylpolysiloxane capillary column separates four previously unresolved peak pairs in a similar run time. Column 30 m TRACE...
In the previous two chapters, equations were developed to provide the optimum column dimensions and operating conditions to achieve a particular separation in the minimum time for both packed columns and open tubular columns. In practice, the vast majority of LC separations are carried out on packed columns, whereas in GC, the greater part of all analyses are performed with open tubular columns. As a consequence, in this chapter the equations for packed LC columns will first be examined and the factors that have the major impact of each optimized parameter discussed. Subsequently open tubular GC columns will be considered in a similar manner. [Pg.395]

Although the OTHdC has several unique applications in polymer analysis, this technique has several limitations. First, it requires the instrumentation of capillary HPLC, especially the injector and detector, which is not as popular as packed column chromatography at this time. Second, as discussed previously, the separation range of a uniform capillary column is rather narrow. Third, it is difficult to couple capillary columns with different sizes together as SEC columns. [Pg.601]

Optimum flowrates are higher in packed column SFC than in LC. Flowrates as high as 5.0 mL min generally do not dramatically reduce efficiency in SFC [12]. Bier-manns and co-workers reported the separation of (3-blockers at a flowrate of 4.0 mL miiT a rate eight times higher than the flowrate recommended for LC [56]. No deterioration of column performance was observed. [Pg.312]

The product is analyzed by vapor phase chromatography using a 6-ft., f-in. O.D. copper tube, packed with 5% Bentone-34 (Wilkins Instrument Co.) and 0.5% XF-1150 (General Electric Silicone Products) on Diatoport-S (80-100 mesh) (F and M Co.) flow rate of helium 60 ml./min., oven temperature 85°. This column separates m-cymene (retention time 12 minutes) from />-cymene (retention time 10 minutes) but does not resolve the ortho isomer. The purity of the distilled w-cymene is above 98%. [Pg.43]

Totally porous particles of relatively large particle sizes were widely used in low pressure liquid chromatography for many years. These column packings had good sample capacity but only limited efficiency accompanied by long separation times, due to the large size and unfavorable size distribution of the particles and the presence of relatively deep pores within the particles through whick sample molecules diffused in and out of very slowly. [Pg.675]


See other pages where Packed columns separation time is mentioned: [Pg.17]    [Pg.1250]    [Pg.224]    [Pg.1250]    [Pg.121]    [Pg.15]    [Pg.16]    [Pg.18]    [Pg.119]    [Pg.237]    [Pg.19]    [Pg.384]    [Pg.130]    [Pg.407]    [Pg.440]    [Pg.19]    [Pg.28]    [Pg.46]    [Pg.365]    [Pg.41]    [Pg.69]    [Pg.49]    [Pg.80]    [Pg.736]    [Pg.24]    [Pg.37]    [Pg.38]    [Pg.40]    [Pg.45]    [Pg.110]    [Pg.319]    [Pg.324]    [Pg.536]    [Pg.544]    [Pg.556]    [Pg.563]    [Pg.565]    [Pg.638]    [Pg.658]    [Pg.677]    [Pg.696]    [Pg.818]   
See also in sourсe #XX -- [ Pg.587 ]




SEARCH



Packed columns

Packed columns, packing

Separation column packing

Separation time

Separator column

© 2024 chempedia.info